ENDOPATH XCEL D12LTRecalls & reported problems
Ethicon Endo-Surgery, LLC · Endopath XCEL Dilating Tip Trocars with Stability Sleeves · Class 2
Quick summary
This is an Ethicon Endopath Xcel dilating tip trocar, a surgical sleeve used to make small openings in the body for laparoscopic surgery. The 35 reports mostly describe the trocar cracking, breaking apart, leaking gas, or arriving in packaging that had lost its sterile seal, along with blades that would not extend or retract properly. One injury and no deaths were reported, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 36
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 10705036001577), received between Jan 8, 2018 and Sep 29, 2026 — classified as 1 injury, 35 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Problem10
- Material Integrity Problem9
- Crack7
- Delivered as Unsterile Product6
- Gas/Air Leak5
- Manufacturing, Packaging or Shipping Problem3
- Product Quality Problem3
- Adverse Event Without Identified Device or Use Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions29
- No Consequences Or Impact To Patient6
- Hematoma1
- Hemorrhage/Blood Loss/Bleeding1
- Pneumothorax1
Recent adverse event reports
The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Manufacturing, Packaging or Shipping Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935647 ↗MalfunctionFeb 13, 2026Gas/Air Leak; Material Integrity Problem
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THE TROCAR PRESENTED A DEFECT IN THE MEMBRANE, CAUSING LOSS OF PNEUMOPERITONEUM. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions; Pneumothorax
MAUDE report 24335461 ↗MalfunctionFeb 5, 2026Crack
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THE BLADE IS NOT FIXED. SO REPLACE THE PRODUCT. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24271777 ↗MalfunctionJan 12, 2026Material Integrity Problem
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THE PLASTIC DEFECT OF THE TROCAR. NO FURTHER DETAILS WERE PROVIDED. NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24044937 ↗MalfunctionDec 18, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THERE WAS A GAS LEAK THAT HAPPENED WHEN USING THE TROCAR, HCP USED A NEW TROCAR TO FINISH THE OPERATION. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23838632 ↗MalfunctionDec 8, 2025Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THE SPRING OF THE BLADE INSIDE THE TROCAR CANNOT EXTEND AND RETRACT PROPERLY. NO FURTHER DETAILS WERE PROVIDED. NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23745544 ↗MalfunctionOct 15, 2025Material Integrity Problem; Adverse Event Without Identified Device or Use Problem; Material Integrity Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, ACCORDING TO THE REPORT, DURING USE THE TROCAR EXPLODED. IT WAS NECESSARY TO CONTINUE WITH THE RESET. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23278615 ↗MalfunctionSep 17, 2025Gas/Air Leak; Material Integrity Problem; Gas/Air Leak; Material Integrity Problem
IT WAS REPORTED THAT DURING A LAPAROSCOPIC RETOSIGMOIDECTOMY WAS STARTED, THE TROCAR WAS USED, WHICH BEGAN TO LEAK THE GAS AND WHEN REMOVING IT FROM THE CAVITY, IT WAS OBSERVED THAT THE SAME CAME WITHOUT THE INTERNAL RUBBER THAT SEALS. ANOTHER ITEM WAS REQUESTED AND EVERYTHING OCCURRED IN THE USUAL WAY. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23071973 ↗MalfunctionAug 11, 2025Delivered as Unsterile Product; Product Quality Problem; Delivered as Unsterile Product; Product Quality Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE WHEN OPENING THE STERILE PACKAGING OF THE PRODUCT, LARGE AMOUNT OF WHITE DUSTY MATERIAL (APPEARING TO BE COMING OUT OF THE PLASTIC SEAL OF THE PACKAGING) SPRAYED OUT POTENTIALLY CAUSING CONTAMINATION OF STERILE FIELD AND THE THEATRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22770755 ↗MalfunctionAug 11, 2025Delivered as Unsterile Product; Product Quality Problem; Delivered as Unsterile Product; Product Quality Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE WHEN OPENING THE STERILE PACKAGING OF THE PRODUCT, LARGE AMOUNT OF WHITE DUSTY MATERIAL (APPEARING TO BE COMING OUT OF THE PLASTIC SEAL OF THE PACKAGING) SPRAYED OUT POTENTIALLY CAUSING CONTAMINATION OF STERILE FIELD AND THE THEATRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22770739 ↗MalfunctionAug 11, 2025Delivered as Unsterile Product; Product Quality Problem; Delivered as Unsterile Product; Product Quality Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE WHEN OPENING THE STERILE PACKAGING OF THE PRODUCT, LARGE AMOUNT OF WHITE DUSTY MATERIAL (APPEARING TO BE COMING OUT OF THE PLASTIC SEAL OF THE PACKAGING) SPRAYED OUT POTENTIALLY CAUSING CONTAMINATION OF STERILE FIELD AND THE THEATRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22770698 ↗MalfunctionMay 14, 2025Gas/Air Leak; Material Integrity Problem; Gas/Air Leak; Material Integrity Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THERE WAS A FAULTY SEAL, KEPT LEAKING THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22017786 ↗MalfunctionDec 3, 2024Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE SHAFT HAS LOOSE RUBBER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20835665 ↗MalfunctionOct 30, 2024Crack; Crack
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE SURGICAL TROCAR HAD A DEFECT IN THE NOZZLE, WHICH WAS BROKEN AND LEAKING. REPLACED THE DEVICE WITH ANOTHER, ALLOWING THE PROCEDURE TO BE COMPLETED WITHOUT COMPLICATIONS OR HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20570939 ↗MalfunctionOct 22, 2024Crack; Gas/Air Leak; Crack; Gas/Air Leak
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, BROKEN DIAPHRAGM AND GAS ESCAPING. REPLACED. NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20508663 ↗MalfunctionOct 16, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THERE WAS A HOLE IN THE PRIMARY PACKAGING OF THE PRODUCT. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20462447 ↗MalfunctionJul 25, 2024Crack; Crack
IT WAS REPORTED THAT DURING THE MIE PROCEDURE, DURING THE INSERTION OF THE TROCAR IN THE ABDOMINAL WALL THE TROCAR BROKE IN DIFFERENT LITTLE PIECES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19827680 ↗MalfunctionApr 15, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THERE WAS A SMALL HOLE ON THE PACKAGE. NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19104316 ↗MalfunctionApr 11, 2024Gas/Air Leak; Material Integrity Problem; Gas/Air Leak; Material Integrity Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, AFTER 10 MINUTES THE TROCAR BEGINS TO FAIL WITHOUT INSUFFLATION, IT IS VERIFIED AND IT IS OBSERVED THAT THE DUCKBILL SEAL IS DETACHED SO IT IS EXTRACTED. DEVICE REPLACED AND PROCEDURE CONTINUED WITH 15 MINUTE DELAY. NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19087859 ↗InjuryApr 9, 2024Mechanical Problem; Adverse Event Without Identified Device or Use Problem; Mechanical Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A LAPAROSCOPIC MYOMECTOMY, WHEN INDUCING THE TROCAR IN THE UMBILICAL SITE, BLOOD SPURTING FROM RIGHT COMMON ILEOARTERY WHEN VIEWING IT USING A CAMERA PORT. PROCEDURE WAS CONVERTED TO OPEN LAPAROTOMY TO SECURE BLEEDING. ON MARCH 27, PATIENT WAS FACING BOWEL INJURY AND HAD TO UNDERGO ANOTHER LAPAROTOMY. PATIENT IS STILL IN HOSPITAL CURRENTLY.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 19069579 ↗MalfunctionFeb 12, 2024Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TROCAR WOULD ACTIVATE AND WOULD CLICK ON ENTRY, BUT THE BLADE WOULD NOT RETRACT. NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18691916 ↗MalfunctionFeb 12, 2024Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TROCAR WOULD ACTIVATE AND WOULD CLICK ON ENTRY, BUT THE BLADE WOULD NOT RETRACT. NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18691905 ↗MalfunctionFeb 12, 2024Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TROCAR WOULD ACTIVATE AND WOULD CLICK ON ENTRY, BUT THE BLADE WOULD NOT RETRACT. NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18691892 ↗MalfunctionFeb 12, 2024Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TROCAR WOULD ACTIVATE AND WOULD CLICK ON ENTRY, BUT THE BLADE WOULD NOT RETRACT. NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18691880 ↗MalfunctionJan 18, 2024Crack; Crack
IT WAS REPORTED THAT DURING A LOBECTOMY SURGERY (VATS APPROACH), WHEN TROCAR WAS REMOVED FROM PATIENT, IT WAS NOTICED TO BE BROKEN. SCANS IN THE SURGICAL FIELD WERE COMPLETED TO LOCATE AND REMOVE FRAGMENTS FROM PATIENTS BODY. THERE IS A POSSIBILITY THAT THEY WERE NOT ABLE TO GET ALL THE FRAGMENTS OUT. PROCEDURE COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18535706 ↗
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.