Skip to content

ENDOPATH BASX TB12LTRecalls & reported problems

Ethicon Endo-Surgery, LLC · Endopath BASX Bladeless Trocars with Stability Sleeves · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon Endopath BASX bladeless trocar is a surgical port used in laparoscopic surgery. Most of the 35 reports describe cracked or broken parts, including small fragments breaking off during surgery that in one case had to be searched for in the abdomen, along with a few units arriving damaged or unsterile in the packaging. The reports mention 2 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
35
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 10705036003007), received between Oct 24, 2017 and Jan 15, 2025 — classified as 2 injuries, 33 malfunctions.

The most frequently reported specific device problem is “Crack” (15 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack15
  • Material Integrity Problem7
  • Component or Accessory Incompatibility5
  • Delivered as Unsterile Product4
  • Break3
  • Packaging Problem2
  • Component Falling1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions26
  • No Consequences Or Impact To Patient7
  • Device Embedded In Tissue or Plaque1
  • Foreign Body In Patient1

Recent adverse event reports

The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 15, 2025Crack; Crack

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, WHEN THE TROCARS WERE REMOVED, IT WAS DISCOVERED THAT A PIECE HAD BROKEN OFF THE INNER EDGE OF ONE TROCAR. THE BROKEN PIECES WERE THEN SEARCHED FOR IN THE ABDOMEN. AFTER A LENGTHY SEARCH, TWO SMALL PIECES WERE FOUND AND REMOVED. THE OPERATING TIME WAS GREATLY EXTENDED AND A SMALL PIECE OF THE TROCAR COULD NOT BE FOUND.

    Patient problems: Foreign Body In Patient

    MAUDE report 21164181 ↗
  • MalfunctionJun 20, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, IT WAS NOTED THE DEVICE WAS DAMAGED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19574790 ↗
  • MalfunctionMar 7, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    THE DEVICE WAS RECEIVED WITH NO EVENT DETAILS. ATTEMPTS WERE MADE TO OBTAIN DETAILS AND NO INFORMATION WAS PROVIDED. NO PATIENT CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18857222 ↗
  • MalfunctionJan 17, 2024Crack; Crack

    IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, IT WAS NOTED THE DEVICE WAS DAMAGED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18528219 ↗
  • MalfunctionAug 18, 2023Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, WHEN OPENING THE PACKAGING IT WAS NOTED THE DEVICE WAS DAMAGED . ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17573581 ↗
  • MalfunctionJun 29, 2023Crack; Crack

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, WHEN THE TROCAR WAS REMOVED, IT WAS FOUND THAT A SMALL PIECE AT THE FRONT END OF THE TROCAR WAS DAMAGED. THE DAMAGED PART WAS STILL CONNECTED WITH THE TROCAR AND WAS NOT LEFT IN THE PATIENT'S BODY. CHANGED TO A NEW DEVICE TO COMPLETE SURGERY. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17229082 ↗
  • MalfunctionJun 14, 2023Crack; Crack

    IT WAS REPORTED THAT DURING AN ENDOSCOPIC STOMACH SURGERY, AFTER INSERT INTO THE TROCAR, NOTED ONE DEVICE WAS BROKEN ,AND ANOTHER DEVICE HAD CRACKS . ANOTHER TWO DEVICE WERE USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17130118 ↗
  • MalfunctionJun 14, 2023Crack; Crack

    IT WAS REPORTED THAT DURING AN ENDOSCOPIC STOMACH SURGERY, AFTER INSERT INTO THE TROCAR, NOTED ONE DEVICE WAS BROKEN ,AND ANOTHER DEVICE HAD CRACKS . ANOTHER TWO DEVICE WERE USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17130091 ↗
  • MalfunctionMay 24, 2023Crack; Crack

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TIP OF TROCAR WAS NOTED BROKEN WHILE THE NURSE WAS CHECKING THE OPERATION ITEMS. AFTER CONFIRMING NO PIECES IN PATIENT, THE SURGEON COMPLETED THE SURGERY. HOWEVER, THE SURGERY WAS EXTENDED FOR ABOUT 2 HOURS DUE TO LOOKING FOR THE BROKEN PIECES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16992935 ↗
  • MalfunctionApr 5, 2023Crack; Crack

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TIP OF TROCAR RUPTURED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16681044 ↗
  • MalfunctionApr 5, 2023Crack; Crack

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TIP OF TROCAR RUPTURED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16681028 ↗
  • MalfunctionApr 5, 2023Crack; Crack

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TIP OF TROCAR RUPTURED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16681021 ↗
  • MalfunctionApr 5, 2023Crack; Crack

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TIP OF TROCAR RUPTURED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16681010 ↗
  • MalfunctionApr 5, 2023Crack; Crack

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TIP OF TROCAR RUPTURED DURING THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16680883 ↗
  • MalfunctionApr 5, 2023Crack; Crack

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TIP OF TROCAR RUPTURED DURING THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16680840 ↗
  • MalfunctionApr 5, 2023Crack; Crack

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TIP OF TROCAR RUPTURED DURING THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16680826 ↗
  • MalfunctionJan 18, 2023Component or Accessory Incompatibility; Material Integrity Problem; Component or Accessory Incompatibility; Material Integrity Problem

    IT WAS REPORTED THAT DURING A GASTRIC BYPASS, THE TROCAR SEAL WAS LEAKING. THE LEAKAGE IS CAUSED BY THE SEAL BEING DAMAGED FROM THE ENDO-GIA STAPLER BEING REMOVED THROUGH THE TROCAR. PROCEDURE PROLONGED 5 MINUTES. NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16194485 ↗
  • MalfunctionJan 18, 2023Component or Accessory Incompatibility; Material Integrity Problem; Component or Accessory Incompatibility; Material Integrity Problem

    IT WAS REPORTED THAT DURING A GASTRIC BYPASS, THE TROCAR SEAL WAS LEAKING. THE LEAKAGE IS CAUSED BY THE SEAL BEING DAMAGED FROM THE ENDO-GIA STAPLER BEING REMOVED THROUGH THE TROCAR. PROCEDURE PROLONGED 5 MINUTES. NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16194446 ↗
  • MalfunctionJan 18, 2023Component or Accessory Incompatibility; Material Integrity Problem; Component or Accessory Incompatibility; Material Integrity Problem

    IT WAS REPORTED THAT DURING A GASTRIC BYPASS, THE TROCAR SEAL WAS LEAKING. THE LEAKAGE IS CAUSED BY THE SEAL BEING DAMAGED FROM THE ENDO-GIA STAPLER BEING REMOVED THROUGH THE TROCAR. PROCEDURE PROLONGED 5 MINUTES. NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16194397 ↗
  • MalfunctionJan 18, 2023Component or Accessory Incompatibility; Material Integrity Problem; Component or Accessory Incompatibility; Material Integrity Problem

    IT WAS REPORTED THAT DURING A GASTRIC BYPASS, THE TROCAR SEAL WAS LEAKING. THE LEAKAGE IS CAUSED BY THE SEAL BEING DAMAGED FROM THE ENDO-GIA STAPLER BEING REMOVED THROUGH THE TROCAR. PROCEDURE PROLONGED 5 MINUTES. NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16194311 ↗
  • MalfunctionJan 18, 2023Component or Accessory Incompatibility; Material Integrity Problem; Component or Accessory Incompatibility; Material Integrity Problem

    IT WAS REPORTED THAT DURING A GASTRIC BYPASS, THE TROCAR SEAL WAS LEAKING. THE LEAKAGE IS CAUSED BY THE SEAL BEING DAMAGED FROM THE ENDO-GIA STAPLER BEING REMOVED THROUGH THE TROCAR. PROCEDURE PROLONGED 5 MINUTES. NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16194288 ↗
  • MalfunctionApr 14, 2022Crack; Crack

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, AFTER INSERTING THE TROCAR, NOTED THE TROCAR WAS DAMAGED AS PHOTO SHOWS. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14113195 ↗
  • MalfunctionApr 14, 2022Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, BEFORE USED ON THE PATIENT, THE PACKING WAS DAMAGED AS THE PHOTO SHOWS. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14113026 ↗
  • MalfunctionApr 14, 2022Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, BEFORE USED ON THE PATIENT, THE PACKING WAS DAMAGED AS THE PHOTO SHOWS. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14112960 ↗
  • MalfunctionJan 28, 2022Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT DURING THE SURGERY FOR TREATMENT OF ADENOCARCINOMA OF COLON, "WHEN THE DEVICE WAS PLACED INTO THE ABDOMINAL CAVITY WITH FORCEPS FOR THE PATIENT, THE SEALING OF THE DEVICE FELL INTO THE ABDOMINAL CAVITY IMMEDIATELY." THE DOCTOR IMMEDIATELY SUSPENDED THE SURGERY, FIRST REMOVED THE SEALING, WHICH DELAYED THE PROGRESS OF THE SURGERY, THEN CHANGED TO ANOTHER DEVICE TO COMPLETE THE SURGERY. PATIENT CONSEQUENCE WAS NOT REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13385393 ↗

Other Laparoscope, general & plastic surgery devices

All Laparoscope, general & plastic surgery devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page