Harmonic ACE HARH36Recalls & reported problems
Ethicon Endo-Surgery, LLC · Harmonic ACE +7 Laparoscopic Shears Advanced Hemostasis 5mm Diameter 36cm Shaft length For Use With Gray Hand Piece Only · Class U
All models of Harmonic ACE (9)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Harmonic ACE laparoscopic shears is an ultrasonic surgical instrument. Reports mostly describe parts separating or breaking off during use, the instrument failing to read input signals, components detaching, and delayed activation. There are 7,678 reports, including 4 deaths and 54 injuries. No recalls name this device. Reports fell to 718 in the last 12 months from 1,096 in the 12 months before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 7,598
- Deaths reported
- 4
- Injuries reported
- 54
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7,598 adverse event reports that list this exact device (DI 10705036014621), received between Nov 20, 2017 and Sep 30, 2026 — classified as 4 deaths, 54 injuries, 7,540 malfunctions.
The most frequently reported specific device problem is “Material Separation” (2,383 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation2,383
- Failure to Read Input Signal1,411
- Detachment of Device or Device Component1,366
- Difficult or Delayed Activation490
- Material Integrity Problem379
- Break357
- Activation, Positioning or Separation Problem268
- Inappropriate or Unexpected Reset221
Patient problems
- No Clinical Signs, Symptoms or Conditions6,035
- No Consequences Or Impact To Patient1,521
- No Code Available23
- Hemorrhage/Blood Loss/Bleeding11
- Blood Loss10
- Foreign Body In Patient8
- Burn(s)7
- Insufficient Information7
Recent adverse event reports
The 25 most recent of 7,598 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Failure to Read Input Signal; Difficult or Delayed Activation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947839 ↗MalfunctionSep 29, 2026Material Separation
It was reported that during a Liver Resection procedure the titanium tip broke off. The broken piece was retrieved by forceps and was discarded. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935635 ↗MalfunctionSep 29, 2026Smoking; Material Integrity Problem
It was reported that during an unknown procedure there was lots of smoke found. Then the tissue pad melted. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931148 ↗MalfunctionSep 28, 2026Activation, Positioning or Separation Problem
It was reported that during a Lobectomy procedure the scalpel was hard to cut. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923256 ↗MalfunctionSep 28, 2026Material Separation
It was reported that during a Hepatectomy procedure the titanium tip broke off. The broken piece was retrieved by forceps and was discarded. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923218 ↗MalfunctionSep 28, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923038 ↗MalfunctionSep 28, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50922336 ↗MalfunctionSep 28, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919356 ↗MalfunctionSep 28, 2026Difficult or Delayed Activation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917315 ↗MalfunctionSep 24, 2026Failure to Read Input Signal
It was reported that during an unknown procedure the ¿Replace Instrument¿ alert screen was displayed. Changed to another device to complete surgery. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50889307 ↗MalfunctionSep 24, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50886514 ↗MalfunctionSep 24, 2026Activation, Positioning or Separation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50885040 ↗MalfunctionSep 24, 2026Activation, Positioning or Separation Problem
It was reported that during an unknown procedure the scalpel was hard to cut. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50884961 ↗MalfunctionSep 24, 2026Failure to Read Input Signal
It was reported that during an unknown procedure the unknown alert screen displayed by the generator. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50884582 ↗MalfunctionSep 24, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50884258 ↗MalfunctionSep 24, 2026Material Separation
It was reported that during an unknown procedure the titanium tip broke off. The broken piece was retrieved by forceps and was discarded. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50883992 ↗MalfunctionSep 24, 2026Material Separation
It was reported that during an unknown procedure the titanium tip broke off. The broken piece was retrieved by forceps and was discarded. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50883842 ↗MalfunctionSep 23, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50873374 ↗MalfunctionSep 23, 2026Failure to Read Input Signal; Difficult or Delayed Activation
It was reported that during an unknown procedure the ¿Replace Instrument¿ alert screen is displayed. And then the device could not be activated. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50873083 ↗MalfunctionSep 23, 2026Failure to Read Input Signal
It was reported that during an unknown procedure the ¿Replace Instrument¿ alert screen was displayed. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50872947 ↗MalfunctionSep 23, 2026Failure to Read Input Signal
It was reported that during an unknown procedure the ¿Relax Pressure on Blade¿ alert screen was displayed by the generator. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50872479 ↗MalfunctionSep 23, 2026Material Separation
It was reported that during an unknown procedure the titanium tip broke off. The broken piece was retrieved by forceps and was discarded. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50870828 ↗MalfunctionSep 23, 2026Material Integrity Problem
It was reported that during an unknown procedure the white tissue pad was partially detached. The fallen piece was retrieved by forceps and was disposed of. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50869723 ↗MalfunctionSep 23, 2026Failure to Read Input Signal
It was reported that during an unknown procedure the instruments cannot be identified. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50869100 ↗MalfunctionSep 23, 2026Failure to Read Input Signal
It was reported that during an unknown procedure the ¿Relax Pressure on Blade¿ alert screen and ¿Replace Instrument¿ alert screen was displayed by the generator. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50868472 ↗
FDA clearance and classification
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Harmonic ACE HARH36
Has the Harmonic ACE HARH36 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705036014621), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Harmonic ACE HARH36?
The FDA's MAUDE database lists 7,598 adverse event reports for this device, received between Nov 20, 2017 and Sep 30, 2026. In those reports, the most common specific device problems are “Material Separation” (2,383), “Failure to Read Input Signal” (1,411) and “Detachment of Device or Device Component” (1,366); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (11), “Blood Loss” (10) and “Foreign Body In Patient” (8). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Harmonic ACE HARH36?
4 reports are classified as deaths, 54 as injuries and 7,540 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Harmonic ACE HARH36, and is it still sold?
The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand Harmonic ACE, model HARH36). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Instrument, ultrasonic surgical devices
- Harmonic HARHD36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical1 recall2 deaths5,036 reports
- Harmonic ACE HAR36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical2,728 reports
- Harmonic HAR1136Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical2 deaths1,532 reports
- THUNDERBEAT 5 mm,35 cm, Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical3 recalls1,372 reports
- THUNDERBEAT 5 mm,35 cm,Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical3 recalls2 deaths1,128 reports
- THUNDERBEAT 5 mm,35 cm,Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical2 recalls1 death967 reports
- Harmonic ACE HARH23Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical934 reports
- Harmonic Ace ACE36EEthicon Endo-Surgery, LLC · Instrument, ultrasonic surgical860 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page