Skip to content

Harmonic HARHD36Recalls & reported problems

Ethicon Endo-Surgery, LLC · HARMONIC HD 1000i Shears, 36cm Shaft Length · Class U

All models of Harmonic (13)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Ethicon Harmonic is an ultrasonic surgical instrument. Reports most often describe the device failing to read its input signal, parts separating and incorrect warnings. Some reports say it could not be activated during surgery. The device appears in 5,056 reports, including 2 deaths and 29 injuries. It has 1 open recall from July 2021 for a cracked part that can get stuck behind the energy button and may lead to burns.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
5,036
Deaths reported
2
Injuries reported
29
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5,036 adverse event reports that list this exact device (DI 10705036015055), received between Nov 20, 2017 and Sep 30, 2026 — classified as 2 deaths, 29 injuries, 5,005 malfunctions.

The most frequently reported specific device problem is “Failure to Read Input Signal” (1,413 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 66 reportsJan 2024: 65 reportsFeb 2024: 57 reportsMar 2024: 52 reportsApr 2024: 29 reportsMay 2024: 27 reportsJun 2024: 51 reportsJul 2024: 26 reportsAug 2024: 32 reportsSep 2024: 32 reportsOct 2024: 26 reportsNov 2024: 21 reportsDec 2024: 19 reportsJan 2025: 34 reportsFeb 2025: 14 reportsMar 2025: 25 reportsApr 2025: 20 reportsMay 2025: 30 reportsJun 2025: 35 reportsJul 2025: 47 reportsAug 2025: 35 reportsSep 2025: 28 reportsOct 2025: 20 reportsNov 2025: 21 reportsDec 2025: 23 reportsJan 2026: 55 reportsFeb 2026: 23 reportsMar 2026: 20 reportsApr 2026: 29 reportsMay 2026: 34 reportsJun 2026: 24 reportsJul 2026: 1 reportsAug 2026: 2 reportsSep 2026: 30 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202666
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 20, 2021Class IIOpen, Classified

    An internal component may be cracked and become lodged behind the energy button potentially resulting in continuous activation of the device, may result in burns

    Z-2356-2021 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Read Input Signal1,413
  • Material Separation1,132
  • Detachment of Device or Device Component586
  • Device Displays Incorrect Message491
  • Break367
  • Difficult or Delayed Activation255
  • Material Integrity Problem229
  • Device Alarm System119

Patient problems

  • No Clinical Signs, Symptoms or Conditions3,061
  • No Consequences Or Impact To Patient1,949
  • No Code Available18
  • Blood Loss10
  • Hemorrhage/Blood Loss/Bleeding9
  • No Patient Involvement9
  • Insufficient Information7
  • Foreign Body In Patient6

Recent adverse event reports

The 25 most recent of 5,036 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Material Separation; Failure to Read Input Signal

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950628 ↗
  • MalfunctionSep 30, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948657 ↗
  • MalfunctionSep 30, 2026Difficult or Delayed Activation

    It was reported that during an unknown procedure the device could not be activated. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948444 ↗
  • MalfunctionSep 29, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934847 ↗
  • MalfunctionSep 29, 2026Activation, Positioning or Separation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933861 ↗
  • MalfunctionSep 29, 2026Difficult or Delayed Activation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932790 ↗
  • MalfunctionSep 24, 2026Difficult or Delayed Activation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889255 ↗
  • MalfunctionSep 24, 2026Material Integrity Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50885559 ↗
  • MalfunctionSep 24, 2026Material Separation

    It was reported that during an unknown procedure the titanium tip broke off. The broken piece was retrieved by forceps and was discarded. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50883320 ↗
  • MalfunctionSep 23, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50873328 ↗
  • MalfunctionSep 23, 2026Material Separation

    It was reported that during an unknown procedure the titanium tip broke off. The broken piece was retrieved by forceps and was discarded. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50873155 ↗
  • MalfunctionSep 23, 2026Failure to Read Input Signal

    It was reported that during a Liver Resection procedure the ¿Replace Instrument¿ alert screen was displayed. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872787 ↗
  • MalfunctionSep 23, 2026Failure to Read Input Signal

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872669 ↗
  • MalfunctionSep 23, 2026Failure to Read Input Signal

    It was reported that during a Right Hemicolectomy procedure the ¿Replace Instrument¿ alert screen was displayed. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50869282 ↗
  • MalfunctionSep 23, 2026Activation, Positioning or Separation Problem

    It was reported that during an unknown procedure the scalpel was hard to cut. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50867511 ↗
  • MalfunctionSep 22, 2026Failure to Read Input Signal

    It was reported that during an unknown procedure the ¿Replace Instrument¿ alert screen is displayed. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50857292 ↗
  • MalfunctionSep 22, 2026Failure to Read Input Signal

    It was reported that during an unknown procedure the ¿Replace Instrument¿ alert screen was displayed. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50857176 ↗
  • MalfunctionSep 22, 2026Difficult or Delayed Activation

    It was reported that during an unknown procedure the device could not be activated. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50854193 ↗
  • MalfunctionSep 21, 2026Difficult or Delayed Activation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50842305 ↗
  • MalfunctionSep 18, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50825387 ↗
  • MalfunctionSep 18, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50824860 ↗
  • MalfunctionSep 17, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50809843 ↗
  • MalfunctionSep 16, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50793793 ↗
  • MalfunctionSep 16, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50789759 ↗
  • MalfunctionSep 15, 2026Failure to Read Input Signal

    It was reported that during an unknown procedure the ¿Device error¿ alert screen displayed by the generator. And then ¿Replace Instrument¿ alert screen is displayed. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50772951 ↗

FDA clearance and classification

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Harmonic HARHD36

Has the Harmonic HARHD36 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2356-2021 (Class II), began on Jul 20, 2021: An internal component may be cracked and become lodged behind the energy button potentially resulting in continuous activation of the device, may result in burns. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Harmonic HARHD36?

The FDA's MAUDE database lists 5,036 adverse event reports for this device, received between Nov 20, 2017 and Sep 30, 2026. In those reports, the most common specific device problems are “Failure to Read Input Signal” (1,413), “Material Separation” (1,132) and “Detachment of Device or Device Component” (586); the most common patient problems are “Blood Loss” (10), “Hemorrhage/Blood Loss/Bleeding” (9) and “Foreign Body In Patient” (6). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Harmonic HARHD36?

2 reports are classified as deaths, 29 as injuries and 5,005 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Harmonic HARHD36, and is it still sold?

The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand Harmonic, model HARHD36). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Instrument, ultrasonic surgical devices

All Instrument, ultrasonic surgical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page