Harmonic HAR1136Recalls & reported problems
Ethicon Endo-Surgery, LLC · HARMONIC 1100 Shears, 36cm Shaft Length · Class U
All models of Harmonic (13)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon Harmonic shears is an ultrasonic surgical instrument. Its 1,578 reports mostly describe the instrument failing to read its input signal and parts separating, and some describe delayed activation or the device stopping during surgery. The reports mention 2 deaths and 36 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,532
- Deaths reported
- 2
- Injuries reported
- 36
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,532 adverse event reports that list this exact device (DI 10705036026167), received between Apr 12, 2021 and Sep 29, 2026 — classified as 2 deaths, 36 injuries, 1,494 malfunctions.
The most frequently reported specific device problem is “Failure to Read Input Signal” (873 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Read Input Signal873
- Material Separation548
- Difficult or Delayed Activation119
- Insufficient Device Problem Information70
- Detachment of Device or Device Component61
- Material Integrity Problem48
- Activation, Positioning or Separation Problem32
- Noise, Audible23
Patient problems
- No Clinical Signs, Symptoms or Conditions1,495
- Hemorrhage/Blood Loss/Bleeding17
- Insufficient Information7
- Burn(s)4
- Failure to Anastomose4
- Abdominal Pain3
- Foreign Body In Patient3
- Perforation3
Recent adverse event reports
The 25 most recent of 1,532 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935730 ↗MalfunctionSep 28, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50920984 ↗MalfunctionSep 28, 2026Failure to Read Input Signal
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917414 ↗MalfunctionSep 28, 2026Material Integrity Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50916899 ↗MalfunctionSep 28, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50915938 ↗MalfunctionSep 28, 2026Failure to Read Input Signal
It was reported that during an unknown procedure there was an issue with the device. It could not be registered by the generator. They tried a different device and there was no issue. No patient consequences reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50913230 ↗MalfunctionSep 24, 2026Failure to Read Input Signal
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50890042 ↗MalfunctionSep 24, 2026Failure to Read Input Signal
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50889526 ↗MalfunctionSep 24, 2026Failure to Read Input Signal
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50889205 ↗MalfunctionSep 24, 2026Failure to Read Input Signal
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50885314 ↗MalfunctionSep 24, 2026Failure to Read Input Signal
It was reported that during an unknown procedure the ¿Replace Instrument¿ alert screen was displayed. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50885213 ↗MalfunctionSep 23, 2026Material Integrity Problem
It was reported that during an unknown procedure the device blade fractured. The device could not be used further and was replaced with a new device. No patient consequences reported when this complaint event occurred.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50869591 ↗MalfunctionSep 23, 2026Failure to Read Input Signal
It was reported that during an unknown procedure the ¿Replace Instrument¿ alert screen was displayed. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50869458 ↗MalfunctionSep 23, 2026Failure to Read Input Signal
It was reported that during an unknown procedure the screen displayed the message "Replace instrument". A new device was installed; however, the same message alarm recurred shortly after use. A second replacement with another new device was performed, but the issue persisted. The procedure was ultimately completed successfully using an alternative manufacturer's product. No patient consequences reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50869404 ↗MalfunctionSep 22, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50860066 ↗MalfunctionSep 22, 2026Failure to Read Input Signal
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50859720 ↗MalfunctionSep 22, 2026Device Markings/Labelling Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50859340 ↗MalfunctionSep 22, 2026Material Separation; Failure to Read Input Signal
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50858961 ↗MalfunctionSep 22, 2026Activation, Positioning or Separation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50857545 ↗MalfunctionSep 22, 2026Difficult or Delayed Activation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50856781 ↗MalfunctionSep 22, 2026Failure to Read Input Signal
It was reported that during an unknown procedure the generator displayed the message ¿Clean the blade.¿ The blade was cleaned according to the on-screen instructions, and the device was activated for approximately 2 seconds with the jaws open. The generator then displayed a second message indicating ¿Damaged blade ¿ replace Harmonic instrument.¿The device was disconnected and reconnected; however, the same error message reappeared. The device had been used correctly throughout the procedure, and cleaned regularly with a moist gauze pad in accordance with recommended practice. There was no impact on the patient. The procedure was only slightly prolonged due to troubleshooting and replacement of with a new device.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50856414 ↗MalfunctionSep 22, 2026Activation, Positioning or Separation Problem
It was reported that during a Lobectomy procedure the scalpel was hard to cut. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50854513 ↗MalfunctionSep 22, 2026Difficult or Delayed Activation
It was reported that during an unknown procedure the device did not work in the middle of use. No patient consequences reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50854120 ↗MalfunctionSep 22, 2026Failure to Read Input Signal
It was reported that during an unknown procedure the ¿Replace Instrument¿ alert screen was displayed. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50853745 ↗MalfunctionSep 22, 2026Failure to Read Input Signal
It was reported that during an unknown procedure the ¿Relax Pressure on Blade¿ alert screen displayed by the generator. Changed to another device to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50853367 ↗
FDA clearance and classification
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Harmonic HAR1136
Has the Harmonic HAR1136 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705036026167), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Harmonic HAR1136?
The FDA's MAUDE database lists 1,532 adverse event reports for this device, received between Apr 12, 2021 and Sep 29, 2026. In those reports, the most common specific device problems are “Failure to Read Input Signal” (873), “Material Separation” (548) and “Difficult or Delayed Activation” (119); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (17), “Burn(s)” (4) and “Failure to Anastomose” (4). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Harmonic HAR1136?
2 reports are classified as deaths, 36 as injuries and 1,494 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Harmonic HAR1136, and is it still sold?
The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand Harmonic, model HAR1136). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Instrument, ultrasonic surgical devices
- Harmonic ACE HARH36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical4 deaths7,598 reports
- Harmonic HARHD36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical1 recall2 deaths5,036 reports
- Harmonic ACE HAR36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical2,728 reports
- THUNDERBEAT 5 mm,35 cm, Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical3 recalls1,372 reports
- THUNDERBEAT 5 mm,35 cm,Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical3 recalls2 deaths1,128 reports
- THUNDERBEAT 5 mm,35 cm,Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical2 recalls1 death967 reports
- Harmonic ACE HARH23Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical934 reports
- Harmonic Ace ACE36EEthicon Endo-Surgery, LLC · Instrument, ultrasonic surgical860 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.