Ethicon Endo-Surgery, LLC · ETHICON 4000 STAPLER STRAIGHT 60 STANDARD · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
20
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 10705036043522), received between Oct 31, 2025 and Sep 30, 2026 — classified as 2 injuries, 18 malfunctions.

The most frequently reported specific device problem is “Failure to Form Staple” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 3 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 3 reportsJul 2026: 2 reportsAug 2026: 3 reportsSep 2026: 5 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Form Staple10
  • Ejection Problem3
  • Positioning Failure3
  • Adverse Event Without Identified Device or Use Problem2
  • Difficult to Open or Close2
  • Failure to Fire2
  • Material Integrity Problem2
  • Mechanical Jam2

Patient problems

  • No Clinical Signs, Symptoms or Conditions18
  • Failure to Anastomose1
  • Hemorrhage/Blood Loss/Bleeding1
  • Insufficient Information1

Recent adverse event reports

The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem

    It was reported that postop to an unspecified surgical procedure, the patient experienced unspecified postoperative complications. No additional information was provided. The device was discarded and, therefore, will not be returned.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Insufficient Information

    MAUDE report 50952435 ↗
  • InjurySep 30, 2026Failure to Form Staple; Adverse Event Without Identified Device or Use Problem

    It was reported that, following an unspecified surgical procedure, the patient developed a leak at the proximal aspect of the gastric sleeve, resulting in prolonged hospitalization. No additional information was provided. The device was discarded and, therefore, unavailable for return.

    Patient problems: Failure to Anastomose

    MAUDE report 50952413 ↗
  • MalfunctionSep 25, 2026Mechanical Jam

    It was reported that, during an unknown surgery, it was observed that the knife did not return at the 2nd firing of Functional End to End Anastomosis. The home button was used. There were no adverse consequences to the patient. No further information is available.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50898484 ↗
  • MalfunctionSep 25, 2026Failure to Form Staple

    It was reported that during an unknown surgery, it was observed that 5-6 staples were malformed at the 1st firing. The device was used on gastric body. The cartridge color was Green. Another competitive device was used to complete the case. There were no adverse consequences to the patient. No further information is available.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50898415 ↗
  • MalfunctionSep 17, 2026Failure to Fire; Difficult to Open or Close; Activation Problem

    It was reported that during a Liver Resection procedure, it was observed that the stapler got stuck on tissue. It also wouldn't fire or open jaw with battery in place or manually. There was no patient consequence nor surgical delay reported. No additional information provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50803400 ↗
  • MalfunctionAug 6, 2026Difficult or Delayed Activation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50160324 ↗
  • MalfunctionAug 4, 2026Positioning Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50132753 ↗
  • MalfunctionAug 3, 2026Material Integrity Problem; Ejection Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50114665 ↗
  • MalfunctionJul 27, 2026Failure to Fire

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50024106 ↗
  • MalfunctionJul 27, 2026Failure to Form Staple

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50019491 ↗
  • MalfunctionJun 29, 2026Failure to Form Staple; Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING A WHIPPLE PROCEDURE, WHILE CREATING THE ANASTOMOSIS, THE SURGICAL TEAM STAPLED ALONG THREE ALLIS CLAMPS. UPON COMPLETION OF THE STAPLING, THE SURGEON REPORTED FEELING SHARP AREAS ALONG THE STAPLE LINE. IT WAS NOTED THAT WHEN THE STAPLER WAS CLAMPED, THERE APPEARED TO BE A GAP BETWEEN THE ANVIL AND THE CARTRIDGE, WHICH MAY HAVE RESULTED IN MALFORMED STAPLES. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25650534 ↗
  • MalfunctionJun 22, 2026Positioning Failure; Difficult to Open or Close; Noise, Audible

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC NEPHRECTOMY PROCEDURE, IT WAS OBSERVED THAT THE STAPLER LOCKED ONTO THE TISSUE OR VESSEL PRIOR TO FIRING. THE USER ATTEMPTED TO REOPEN THE DEVICE BUT WAS UNABLE TO DO SO. THE DEVICE HAD NOT BEEN FIRED. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25578801 ↗
  • MalfunctionJun 1, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING A ROBOT-ASSISTED GASTRECTOMY, WHEN THE DEVICE WAS USED FOR DUODENAL TRANSECTION, PARTIAL MALFORMED STAPLE FORMATION WAS OBSERVED. HOWEVER, AS THE STAPLE LINE WAS REMOVED DURING ANASTOMOSIS, THERE WAS NO DIRECT IMPACT ON THE PATIENT. NO ADDITIONAL ACTION WAS REQUIRED SINCE THE AREA WITH MALFORMED STAPLES WAS REMOVED DURING ANASTOMOSIS. THE PRODUCT WAS USED ON THE DUODENUM. THE CARTRIDGE COLOR WAS BLUE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25351287 ↗
  • MalfunctionApr 15, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING A RROBOT-ASSISTED DISTAL GASTRECTOMY FOR STOMACH CANCER. DURING TRANSECTION OF THE DUODENUM, ONE STAPLE AT THE JAWS TIP (KNIFE SIDE) WAS NOT FORMED. THE MALFORMED ONE STAPLE WAS REMOVED AND THE SURGERY WAS CONTINUED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24887549 ↗
  • MalfunctionFeb 20, 2026Positioning Failure; Ejection Problem

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC GASTRIC BYPASS PROCEDURE, IT WAS OBSERVED THAT THE END EFFECTOR AT THE DISTAL BLACK ARTICULATION JOINT WAS EXTREMELY MOVABLE WHICH MADE IT HARD TO UNLOAD AND RELOAD. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24402036 ↗
  • MalfunctionJan 8, 2026Ejection Problem

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE MIDDLE BASE OF THE RELOAD STAYED WITH THE JAWS OF THE STAPLER WHILE UNLOADING AND GOT STUCK. THIS MADE US UNABLE TO RELOAD THE STAPLER AND HAVE TO PULL NEW RELOAD AND STAPLER. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24006156 ↗
  • MalfunctionJan 5, 2026Failure to Form Staple; Material Integrity Problem; Mechanical Jam

    IT WAS REPORTED THAT DURING AN INTERNAL TRAINING LAB USING A NON-STERILE DEVICE, THE DEVICE WAS FIRED ON THE FUNDUS OF HARVESTED PORCINE STOMACH USING A WHITE RELOAD WITH SLR AND FIRED WITHOUT INCIDENT. ON THE SECOND FIRING ON PORCINE STOMACH ON A THICKER AREA NEAR THE FUNDUS USING A WHITE RELOAD WITH SLR, THE DEVICE WAS BEING PULSE FIRED. THE DEVICE FIRED, BUT THE KNIFE DID NOT RETURN TO THE HOME POSITION. THE HOME BUTTON WAS USED TO RETURN THE KNIFE AND OPEN THE JAWS. UPON OPENING, IT WAS NOTED THAT THE ANVIL WAS DAMAGED (ANVIL PULL THROUGH) AND MALFORMED STAPLES WERE OBSERVED. THE DEVICE WAS FIRED A THIRD TIME WITH A BLUE RELOAD WITH NO SLR ON JEJUNUM JUST TO SEE WHAT WOULD HAPPEN AND FIRED COMPLETELY AND THE KNIFE RETURNED. MALFORMED STAPLES WERE OBSERVED IN THE DAMAGED AREA OF THE JAWS, AS EXPECTED. THE DEVICE WAS LATER BAILED OUT AS A TRAINING EXERCISE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23958326 ↗
  • MalfunctionJan 2, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING A SLEEVE GASTRECTOMY PROCEDURE, AFTER THE FIRST FIRING IT SEEMED THAT THE SURGEON KIND OF OVERLAPPED THE PREVIOUS STAPLE LINES WITH EACH SEQUENTIAL FIRING AND THE STAPLES DID NOT PROPERLY GO INTO THE TISSUE AND WERE KIND OF HANGING OFF AT THE BEGINNING OF EACH STAPLE LINE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23950382 ↗
  • MalfunctionNov 11, 2025Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING A PANCREATECTOMY PROCEDURE, THE SURGEON FIRED ON THE SECOND FIRING OF THE DEVICE; THE STAPLE LINE OPENED AND DID NOT LATCH. THE GREEN RELOAD WAS USED WITH STAPLE-LINE REINFORCEMENT. THE TEAM OVERSEWED THE PANCREAS, AND THE PATIENT REMAINED STABLE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23522743 ↗
  • MalfunctionOct 31, 2025Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT, DURING A VATS LOBECTOMY, THE SURGEON INITIALLY FIRED ACROSS THE BRONCHUS WITH A GREEN RELOAD, WHICH APPEARED COMPLETE AND SATISFACTORY. AT THE END OF THE CASE, NO LEAK WAS FOUND IN THE BRONCHUS DURING HEMOSTASIS. HOWEVER, THE SURGEON LATER DISCOVERED MALFORMED STAPLES IN THE MIDDLE OF THE STAPLE LINE, AND THE LEAK ORIGINATED FROM THE AREA WHERE THE BLUE AND WHITE RELOAD WAS USED ON THE LUNG.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23441701 ↗

FDA clearance and classification

The FDA regulates product code GAG (stapler, surgical) under 21 CFR 878.4740 as a Class II device. 21 CFR 878.4740 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: A surgical stapler for internal use is a specialized prescription device used to deliver compatible staples to internal tissues during surgery for resection, transection, and creating anastomoses.

Common questions about the ETHICON EC3D60S

Has the ETHICON EC3D60S been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705036043522), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the ETHICON EC3D60S?

The FDA's MAUDE database lists 20 adverse event reports for this device, received between Oct 31, 2025 and Sep 30, 2026. In those reports, the most common specific device problems are “Failure to Form Staple” (10), “Ejection Problem” (3) and “Positioning Failure” (3); the most common patient problems are “Failure to Anastomose” (1) and “Hemorrhage/Blood Loss/Bleeding” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ETHICON EC3D60S?

None of the reports are classified as deaths, 2 as injuries and 18 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the ETHICON EC3D60S, and is it still sold?

The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand ETHICON, model EC3D60S). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Stapler, surgical devices