ETHICON EC3D60CRecalls & reported problems
Ethicon Endo-Surgery, LLC · ETHICON 4000 STAPLER STRAIGHT 60 COMPACT · Class 2
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10705036043539), received between Jan 8, 2026 and Sep 30, 2026 — classified as 1 injury, 11 malfunctions.
The most frequently reported specific device problem is “Failure to Form Staple” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Form Staple9
- Failure to Fire2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50952215 ↗MalfunctionSep 28, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50916879 ↗MalfunctionSep 1, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50543916 ↗MalfunctionAug 20, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50349960 ↗MalfunctionJun 22, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING A RIGHT HEMICOLECTOMY, MALFORMED STAPLES WERE OBSERVED ON THE RESECTED TISSUE FORM THE FIRST OR SECOND FIRING. THE MALFORMED AREA WAS REPAIRED WITH BURIED SUTURES, AND THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25580472 ↗InjuryJun 8, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A COLECTOMY PROCEDURE, THE STAPLE LINE WAS BLEEDING. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 25432446 ↗MalfunctionMay 28, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING A THORACOSCOPIC LUNG RESECTION PROCEDURE, ETHICON 4000 WITH A GREEN CARTRIDGE WAS USED ON THE LEFT LOWER BRONCHUS. WHEN CHECKING THE STAPLE LINE, SIX MALFORMED STAPLES (APPROXIMATELY 80-90%) WERE OBSERVED AT THE 2D STAPLE LINE ON THE MAIN BRONCHUS SIDE OF THE STUMP. THE STAPLES FORMED IN A U-SHAPE AND WERE CONSIDERED NOT TO HAVE CONTACTED THE ANVIL. ADDITIONAL SUTURING AND REINFORCEMENT WITH MEDIASTINAL FAT TISSUE WERE PERFORMED, AND THE PROCEDURE WAS COMPLETED. THE PATIENT CONDITION AFTER SURGERY WAS STABLE. THE CARTRIDGE COLOR WAS GREEN. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25318396 ↗MalfunctionMay 12, 2026Failure to Form Staple
IT WAS REPORTED THAT THE DEVICE WAS USED THE PREVIOUS DAY AND MISFIRED, WITH IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE MISFIRED AS A FEW STAPLES WERE STICKING STRAIGHT OUT AND NOT FORMING A CURVED SHAPE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25150060 ↗MalfunctionApr 13, 2026Failure to Form Staple
IT WAS REPORTED THAT THORACIC SURGEON AND PA WERE COMPLETING VATS WEDGE RESECTION USING EC3D60C. WHEN FIRING ACROSS THE PARENCHYMA USING ER60G, THEY NOTICED THE STAPLES WERE NOT PROPERLY FORMED AND APPEARED AS QUILL-LIKE PROTRUDING EXTERNALLY FROM THE TISSUE. THEY DID NOT FIRE ACROSS EXISTING STAPLE LINES WHEN IT OCCURRED. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24864390 ↗MalfunctionMar 11, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING A WHIPPLE PROCEDURE, THEY HAD FIRED THE DEVICE 8-9 TIMES AND ON THE LAST FIRING, THE STAPLES LOOKED LIKE A V SHAPE. THE SURGEON SAID THEY FELT LIKE THE STAPLE LINE WAS IN TACT, SO NO PATIENT HARM, BUT WAS STILL VERY UNSETTLING. THE SURGEON BELIEVED THERE WERE ABOUT 6 THAT LOOKED LIKE THE V SHAPE AT THE DISTAL END OF THE JAWS. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24574452 ↗MalfunctionMar 3, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING A WHIPPLE PROCEDURE, IT WAS OBSERVED THAT THE INNER STAPLE LINE DID NOT FORM APPROPRIATELY. THE STAPLES WERE STICKING OUT, AND THE SURGEON HAD TO OVER SEW THE STAPLE LINE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24495911 ↗MalfunctionJan 8, 2026Failure to Fire
IT WAS REPORTED THAT DURING A WHIPPLE PROCEDURE, IT WAS OBSERVED THAT THE DEVICE DID NOT ACTIVATE WHEN THE SURGEON ATTEMPTED TO FIRE IT. THE RELOAD WAS THEN REPLACED, AND THE STAPLER FUNCTIONED PROPERLY. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24010393 ↗
Other Stapler, surgical devices
- ECHELON ECH60SEthicon Endo-Surgery, LLC · Stapler, surgical1 death631 reports
- ECHELON ECH60LEthicon Endo-Surgery, LLC · Stapler, surgical467 reports
- ECHELON ECH45SEthicon Endo-Surgery, LLC · Stapler, surgical269 reports
- EEA TRIEEA28XTCovidien LP · Stapler, surgical220 reports
- ECHELON ECH60CEthicon Endo-Surgery, LLC · Stapler, surgical197 reports
- ECHELON ECH45CEthicon Endo-Surgery, LLC · Stapler, surgical138 reports
- EEA TRIEEA31XTCovidien LP · Stapler, surgical73 reports
- ETHICON EC3D60LEthicon Endo-Surgery, LLC · Stapler, surgical39 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.