Skip to content

Bausch + Lomb DP5501Recalls & reported problems

Bausch & Lomb Incorporated · Phaco Accessory Packs · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10757770055828), received between Aug 26, 2024 and Aug 26, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Inaccurate Delivery” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inaccurate Delivery2

Patient problems

  • Eye Injury2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 26, 2024Inaccurate Delivery; Inaccurate Delivery

    THE USER VETERINARY FACILITY REPORTS THAT DURING THE PROCEDURE (WHEN SCULPTING A BOWL AND NEARING THE POSTERIOR CORTEX), THE PHACO SLEEVES ARE SPLITTING OPEN NEAR THE POINT WHERE THE CYLINDRICAL PART MEETS THE SCREW-ON BASE. ALTHOUGH THIS SITE IS OUTSIDE OF THE EYE, WHEN IT SPLITS OPEN A FOUNTAIN ERUPTS AND THE ANTERIOR CHAMBER BECOMES VERY SHALLOW, POSTERIOR CAPSULE EMERGES BETWEEN TWO LENS FRAGMENTS AND IS CAUGHT BY PHACO TIP, AND THEN FLUID SURGES (MACHINE SENSES THE LOSS OF PRESSURE) AND THE LENS FRAGMENTS ARE PUSHED THROUGH THE POSTERIOR CAPSULE INTO THE VITREOUS.

    Patient problems: Eye Injury

    MAUDE report 20065088 ↗
  • InjuryAug 26, 2024Inaccurate Delivery; Inaccurate Delivery

    THE USER VETERINARY FACILITY REPORTS THAT DURING THE PROCEDURE (WHEN SCULPTING A BOWL AND NEARING THE POSTERIOR CORTEX), THE PHACO SLEEVES ARE SPLITTING OPEN NEAR THE POINT WHERE THE CYLINDRICAL PART MEETS THE SCREW-ON BASE. ALTHOUGH THIS SITE IS OUTSIDE OF THE EYE, WHEN IT SPLITS OPEN A FOUNTAIN ERUPTS AND THE ANTERIOR CHAMBER BECOMES VERY SHALLOW, POSTERIOR CAPSULE EMERGES BETWEEN TWO LENS FRAGMENTS AND IS CAUGHT BY PHACO TIP, AND THEN FLUID SURGES (MACHINE SENSES THE LOSS OF PRESSURE) AND THE LENS FRAGMENTS ARE PUSHED THROUGH THE POSTERIOR CAPSULE INTO THE VITREOUS.

    Patient problems: Eye Injury

    MAUDE report 20064365 ↗

Other Unit, Phacofragmentation devices

All Unit, Phacofragmentation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page