Skip to content

Bausch + Lomb SE5420MRecalls & reported problems

Bausch & Lomb Incorporated · 20 ga Posterior Mid-Field Elite Pack · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Aug 8, 2018Class IITerminated

    The back cap separates from the body of the vitrectomy cutter.

    Z-3266-2018 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code HQE (instrument, vitreous aspiration and cutting, ac-powered) under 21 CFR 886.4150 as a Class II device. 21 CFR 886.4150 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Bausch + Lomb SE5420M

Has the Bausch + Lomb SE5420M been recalled?

Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-3266-2018 (Class II), began on Aug 8, 2018: The back cap separates from the body of the vitrectomy cutter. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Who makes the Bausch + Lomb SE5420M, and is it still sold?

The FDA's device database lists Bausch & Lomb Incorporated as the company responsible for it (brand Bausch + Lomb, model SE5420M). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Instrument, Vitreous Aspiration And Cutting, Ac-Powered devices

All Instrument, Vitreous Aspiration And Cutting, Ac-Powered devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page