Skip to content

VITROS 6802454Recalls & reported problems

Ortho Clinical Diagnostics · VITROS Immunodiagnostic Products Anti-HCV Reagent Pack · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10758750002801), received between Dec 19, 2025 and Dec 19, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “False Negative Result” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 19, 2025False Negative Result; False Negative Result

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT A DISCORDANT, NON-REACTIVE (NEGATIVE) VITROS ANTI-HCV (AHCV) RESULT WAS OBTAINED FROM A NON-VITROS RANDOX INTERNATIONAL QUALITY ASSESSMENT SCHEME (RIQAS) PROFICIENCY FLUID USING VITROS IMMUNODIAGNOSTIC PRODUCTS AHCV REAGENT ON A VITROS XT 7600 INTEGRATED SYSTEM. THE RESULT WAS CONSIDERED DISCORDANT WHEN COMPARED TO THE REPEAT REACTIVE RESULTS FROM THE SAME SAMPLE ACROSS TWO ANALYZERS. RIQAS PROFICIENCY SAMPLE RESULT OF 0.66 S/C (NEGATIVE) VERSUS THE EXPECTED RESULT OF 1.18 S/C (REACTIVE) BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE DISCORDANT, NON-REACTIVE VITROS AHCV RESULT WAS OBSERVED USING A NON-PATIENT FLUID AND WAS NOT REPORTED FROM THE LABORATORY. THERE HAS BEEN NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23848895 ↗

Other ASSAY,ENZYME LINKED IMMUNOSORBENT,HEPATITIS C VIRUS devices

All ASSAY,ENZYME LINKED IMMUNOSORBENT,HEPATITIS C VIRUS devices →