Skip to content

ARROW IPN036979Recalls & reported problems

Teleflex Incorporated · Arterial Catheterization · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10801902118087), received between May 17, 2024 and Sep 23, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Physical Resistance/Sticking” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K810962); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 21, 2026Open, Classified

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

    Z-2391-2026 · Root cause: Nonconforming Material/Component · FDA record ↗

Most reported problems

Device problems

  • Physical Resistance/Sticking3
  • Material Twisted/Bent2
  • Unraveled Material2
  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2024Material Twisted/Bent; Physical Resistance/Sticking; Material Twisted/Bent; Physical Resistance/Sticking

    IT WAS REPORTED THAT: " THE NEEDLE WENT INTO THE ARTERY JUST FINE. SHE WAS ABLE TO PUT THE GUIDEWIRE THROUGH THE NEEDLE SMOOTHLY. THE WIRE KINKED AT THE TIP OF THE NEEDLE SO SHE COULD NOT MOVE THE WIRE AND NEEDLE. DUE TO THE KINKED WIRE, SHE REMOVED THE NEEDLE AND WIRE TOGETHER AS ONE UNIT OUT THE PATIENT. TREATMENT WAS DELAYED BUT THERE WAS NO OTHER PATIENT HARM. ANOTHER DEVICE WAS USED INSTEAD."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20277206 ↗
  • MalfunctionSep 19, 2024Difficult to Remove; Material Twisted/Bent; Physical Resistance/Sticking; Difficult to Remove; Material Twisted/Bent; Physical Resistance/Sticking

    IT WAS REPORTED "PATIENTS RADIAL ATRIAL LINE WAS ACCIDENTALLY REMOVED WHILE ON 3 VASOPRESSORS. ATTEMPTED TO REESTABLISH RADIAL ARTERIAL ACCESS WITH NO SUCCESS. WHEN ATTEMPT WAS MADE TO PLACE FEMORAL ARTERIAL LINE, I WAS ABLE TO EASILY CANNULATE THE ARTERY WITH MY NEEDLE, HAD GREAT PULSATILE BLOOD FLOW, BUT WHEN I ATTEMPT TO THREAD THE GUIDE WIRE, I WAS UNABLE TO ADVANCE AND HAD DIFFICULTY REMOVING THE WIRE. WHEN I WAS FINALLY ABLE TO REMOVE THE WIRE, IT WAS KINKED". THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20255567 ↗
  • MalfunctionMay 21, 2024Unraveled Material; Physical Resistance/Sticking; Unraveled Material; Physical Resistance/Sticking

    IT WAS REPORTED THAT: "THE INTRODUCER NEEDLE GOES INTO THE ARTERY AND THEN THEY PUT THE GUIDEWIRE THROUGH THE NEEDLE. WHEN SHE GOES TO PULL THE WIRE OUT, THE TIP GOT STUCK ON THE NEEDLE. THE WIRE WAS ABLE TO BE REMOVED IN FULL FROM THE PATIENT. IT WAS UNRAVELED." THE PATIENT IS REPORTED AS FINE. ASSOCIATED COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19357885 ↗
  • MalfunctionMay 17, 2024Unraveled Material; Unraveled Material

    IT WAS REPORTED THAT: "THE INTRODUCER NEEDLE GOES INTO THE ARTERY AND THEN THE GUIDEWIRE IS PUT THROUGH THE NEEDLE. WHEN THE WIRE IS PULLED OUT THE TIP GETS STUCK ON THE NEEDLE. THE WIRE WAS ABLE TO BE REMOVED FROM THE PATIENT BUT GOT CAUGHT UP WHEN REMOVING." THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19339170 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K810962. One submission often covers many models, so check the product and lot details in each record.

  • Z-2377-2026 · Apr 21, 2026 · Open, Classified
    Description/REF: ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM6 RA CATH KIT: 20 GA X 1-3/4IN/NA-04020-X1ASP ACCESS TRAY/ASK-04001-MS RA CATH KIT: 20 GA X 8 CM/ASK-00820-UM1 VESSE
  • Z-1960-2018 · Apr 11, 2018 · Terminated
    Spring Wire Guide Introducer Needle; Product Codes: AN-04320 and AN-04418

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.