ARROW IPN038521Recalls & reported problems
Teleflex Incorporated · Arrowg+ard Blue(R) PSI Kit for use with 7.5 - 8 Fr. Catheters · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10801902141023), received between Jan 17, 2025 and Jan 17, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Separation” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
17 other recalls cover devices cleared under the same FDA submission (K780532); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Apr 21, 2026Open, Classified
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
Z-2393-2026 · Root cause: Nonconforming Material/Component · FDA record ↗
Most reported problems
Device problems
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 17, 2025Material Separation; Material Separation
IT WAS REPORTED THAT THE SIDE PORT OF AN INTRODUCER COMPLETELY POPPED OFF FROM THE INTRODUCER INTO TWO PIECES. THE CATHETER HAD BEEN IN PLACE FOR SIX DAYS WHEN THE SIDE PORT "POPPED OFF." ANOTHER CATHETER WAS RE-INSERTED TO RESOLVE THE ISSUE. THE PATIENT HAD NO HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21178712 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K780532. One submission often covers many models, so check the product and lot details in each record.
- Z-2385-2026 · Apr 21, 2026 · Open, ClassifiedDescription/REF: ACCESS TRAY/GU-04020; PSI KIT: 8.5 FR/AK-09803-CDC; PSI KIT: 9 FR/AK-09903-CDC; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-09810-PCMH1; PSI KIT: 9 FR X 10 CM/ASK-09903-KH;
- Z-1242-2020 · Dec 20, 2019 · TerminatedArrow PSI Kit for use with 7 - 7.5 Fr. Catheters ASK-09801-UPM
- Z-2576-2018 · May 23, 2018 · TerminatedPercutaneous Sheath Introducer Kit for use with 7 - 7.5 Fr. Catheters (8 Fr. 10 cm sheath length .035 inch dia. spring-wire guide), REF ES-09807. The percutaneous sheath introduc
- Z-2011-2018 · Apr 11, 2018 · TerminatedSpring-Wire Guide Introducer Catheter Assembly; Product Code: CN-04018
- Z-1977-2018 · Apr 11, 2018 · TerminatedPercutaneous Sheath Introducer Kit with Integral Hemostasis Valve/Side Port and Sharps Safety Features (Recommended for use with 7 Fr. Catheters); Product Code: ASK-09907-SB
- Z-2032-2018 · Apr 11, 2018 · Terminated9 Fr. Radiopaque Polyurethane Sheath/Dilator with Integral Hemostasis Valve/Side Port for use with 7 - 8 Fr. Catheters and Arrow Percutaneous Introducer System; Product codes: SS-0
- Z-2031-2018 · Apr 11, 2018 · TerminatedARROWg+ard Blue PSI Kit with Integral Hemostasis Valve/ Side Port for use with 7 - 7.5 Fr. Catheters; Product codes: SF-29803-1
- Z-1976-2018 · Apr 11, 2018 · TerminatedPercutaneous Sheath Introducer Kit with Integral Hemostasis Valve/Side Port for use with 7.5 - 8 Fr. Catheters; Product Code: ASK-09903-UPA
- Z-1329-2018 · Dec 22, 2017 · Terminated(1) Percutaneous Sheath Introducer Kit with Integral Hemostasis Valve/Side Port, Sharps Safety Features, and Maximal Barrier Precautions for use with 7.5 - 8 Fr. Catheters, Part Nu
- Z-1158-2018 · Nov 14, 2017 · TerminatedPercutaneous Sheath Introducer Kit with Integral Hemostasis Valve/Side Port, Sharps Safety Features and Maximal Barrier Precautions for use with 7.5 - 8 Fr. Catheters. Permits ven
- Z-1287-2018 · Oct 26, 2017 · TerminatedArterial Access Tray, introducer catheter; Model: GU-04020
- Z-2329-2017 · May 9, 2017 · Terminated1) Introducer Kit with Integral Hemostasis Valve/Side Port, Sharps Safety Features, and Maximal Barrier Precautions for use with 7.5 - 8 Fr., Introducer Catheter 2) Arrow PSI Kit
- Z-0066-2015 · Aug 6, 2014 · TerminatedArrow Select Kits (ASK) and Maximal Barrier Precautions Kits (CDC) with BD Medical Posiflush 10mL Saline Flush Syringe.
- Z-1713-2014 · Apr 14, 2014 · TerminatedArrow Percutaneous Sheath Introducer Kits The Percutaneous sheath introducer products permit venous or arterial access and device introduction to the central circulation.
- Z-0468-2014 · Oct 9, 2013 · TerminatedArrow Transradial Artery Access products Arrow Transradial Artery Access products are used for percutaneous introduction of devices into the radial artery for diagnostic and thera
…and 2 more.
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.