ARROW recallZ-2376-2026 · Arrow International, LLC · initiated Apr 21, 2026
Class I Open, Classified All recalls posted in June 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Arrow International, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 4/21/2026, correction notices (Firm Event Number: EIF-000608, lidocaine and bupivacaine) were mailed to customers who were asked to do the following: Arrow International LLC is voluntarily recalling product due to the Huons Co. Ltd. recall of Lidocaine and Bupivacaine, which is included in impacted Arrow Kits & Sets. Medical facilities 1. Provide this correction notice to all persons who need to be aware within your organization or to any organization where the potentially affected product has been transferred. 2. Check your inventory for product within the scope of this Medical Device Notification and place a copy of this notice with all affected product. 3. Complete and return the response form virtually via: https://fca.myteleflex.com/en/recall/000608 Distributors: 1. Provide Medical Device Notification to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you as their distributor. 2. Check your inventory for affected product. Place a copy of this notification with all affected product and ensure that the notification is included with future distributions of affected product. 3. Complete and return the response form via link: https://fca.myteleflex.com/en/recall/000608 4. If you have further distributed product outside of your country, please notify Teleflex Customer Service by return email. If you have questions contact firm's Customer Service: Link: https://fca.myteleflex.com/en/recall/000608, Customer Service: Telephone: 1-866-396-2111, FAX: 1-855-419-8507 Email: [email protected] On 5/29/2026, additional amended notices (Firm Event Number: EIF-000608-01) were mailed to include additional affected devices. On 5/1/2026, additional notices (Firm Event Number: EIF-000609, sodium chloride) were mailed and emailed to customers who have 0.9% sodium chloride (10mL saline solution ampule) in kits and sets.
Recall details
- Recall number
- Z-2376-2026
- Recall class
- Class I (most serious)
- Status
- Open, Classified
- Initiated by the company
- Apr 21, 2026
- Posted by the FDA
- Jun 13, 2026
- Recalling firm
- Arrow International, LLC
- Quantity
- 308,263
- Product code
- LRS
- FDA submissions
- K071538, K900263, K970864, K993691
- Distribution
- US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
Product description
Description/REF: CVC/PSI KIT: 2 LUMEN 8 FR X 6 IN/ASK-42802-PLH; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-12123-F; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-15123-F; 3L CVC KIT: 7 FR X 16 CM/ASK-12703-PMAH; 3L PI CVC KIT: 7 FR X 16 CM/ASK-12703-PNM; 3L CVC KIT: 7 FR X 16 CM/ASK-12703-PUMH; MULTILUMEN PI CVC KIT: 3L 7 FR X 8 IN/ASK-15703-PRH; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/CDC-15703-P1A; PI CVC KIT: 3-L 7 FR X 6IN (16 CM)/NA-12703-PCDC; PI CVC KIT: 3-L 7 FR X 8IN (20 CM)/NA-15703-PCDC; CVC KIT: 3-LUMEN 5.5 FR X 8 CM/AK-25553; 2-L CVC KIT: 4 FR X 5 CM/CDC-22402-X1A; 2-L CVC KIT: 5 FR X 5 CM/CDC-22502-X1A; 2-L CVC KIT: 4 FR X 8 CM/CDC-25402-X1A; 2-L CVC KIT: 5 FR X 8 CM/CDC-25502-X1A; 3-L CVC KIT: 5.5 FR X 8 CM/CDC-25553-X1A; 3-L CVC KIT: 5.5 FR X 13 CM/CDC-26553-X1A; CVC KIT 3L: 12 FR X 16 CM AGB/AK-22123-CDC; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-22123-F; CVC KIT: 2-LUMEN 4 FR X 5 CM/AK-22402; CVC KIT: 2-LUMEN 5 FR X 5 CM/AK-22502; CVC KIT: 16 GA X 8IN/AK-24301-SK; CVC KIT: 2-LUMEN 4 FR X 13 CM/AK-24402; CVC KIT: 3-LUMEN 7 FR X 30 CM/AK-24703-ASK; CVC KIT 3L: 12FR X 20 CM AGB/AK-25123-CDC; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-25123-F; CVC KIT: 2-LUMEN 4 FR X 8 CM/AK-25402; CVC KIT: 2-LUMEN 5 FR X 8 CM/AK-25502; CVC KIT: 3-LUMEN 5.5 FR X 13 CM/AK-26553; LBCVC KIT: 3L 12 FR X 16 CM AGB/CDC-22123-1A; LBCVC KIT: 3L 12 FR X 16 CM AGB/CDC-22123-X1A; CVC KIT: 16 GA X 20 CM/CDC-24301-1A; PI CVC KIT: 2L 7 FR X 16 CM AGB/AK-22702-P1A; PI CVC KIT: 2L 7 FR X 16 CM AGB/CDC-22702-XP1A; CVC KIT: 3L 7 FR X 20 CM/ASK-45703-CPR; MULTILUMEN PI CVC KIT: 3L 7 FR X 16 CM/ASK-42703-AKP1A; MULTI-LUMEN PI CVC KIT: 3-L 7 FR X 6IN/ASK-42703-KH; PI CVC KIT: 3L 7FR X 16CM AGB+/ASK-42703-MGH1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6IN/ASK-42703-PCCH2; PI CVC KIT: 3LUMEN 7 FR X 6IN (16 CM)/ASK-42703-PFS1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PNO2; MULTI LUMEN PI CVC KIT: 3L 7 FR/ASK-42703-PSCH1; MULTILUMEN PI CVC KIT: 2L 8 FR X 16 CM/ASK-42802-AKP1A; CVC KIT: 2L 8 FR X 16CM/ASK-42802-LMDTG1; PI CVC KIT: 2L 8 FR X 6 IN (16 CM) AGB+/ASK-42802-PUCI1; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-PUMH; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-UMM; PI CVC KIT: 4-L 8.5 FR X 6 IN/ASK-42854-AKP1A; PI CVC KIT: 4L 8.5FR X 16CM AGB+/ASK-42854-MGH; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-AKP1A; PI CVC KIT: 3L 7 FR X 20CM AGB+/ASK-45703-MGH; MULTI LUMEN PI CVC KIT: 3L 7 FR X 8 IN/ASK-45703-PCCH2; MULTI LUMEN PI CVC KIT: 3L 7 FR X 8 IN/ASK-45703-PECL1; PI CVC KIT: 3 LUMEN 7 FR X 8 IN (20 CM)/ASK-45703-PFS1; PI CVC KIT: 3L 7 FR X 8 IN (20 CM) AGB+/ASK-45703-PHM1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PHU; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PMSG1; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PNCB; MULTILUMEN CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PPCM; PI CVC KIT: 3-LUMEN 7 FR X 8 IN (20 CM)/ASK-45703-PRH; MULTILUMEN PI CVC KIT: 3L 7 FR X 8 IN/ASK-45703-PSJ2; PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PSVM; MULTI LUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PUHS; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PUKL; PI CVC KIT: 3L 7FR X 20CM AGB+/ASK-45703-PUMF1; PI CVC KIT: 3L 7 FR X 8 IN (20 CM)/ASK-45703-PUR1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-RHR1; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-WA; MULTI LUMEN PI CVC KIT: 2L 8 FR X 8 IN/ASK-45802-AKP1A; PI CVC KIT: 4L 8.5 FR X 20 CM/ASK-45854-AKP1A; PI CVC KIT: 4L 8.5FR X 20CM AGB+/ASK-45854-MGH; PI CVC KIT: 4L 8.5 FR X 8IN/ASK-45854-PCCH2; CVC KIT: 4L 8.5 FR X 20 CM AGB/ASK-45854-PNM; MULTI LUMEN CVC KIT: 4L 8.5 FR X 8 IN/ASK-45854-PPCM; PI CVC KIT: 4L 8.5 FR X 20 CM AGB+/ASK-45854-PUCH2; MULTILUMEN PI CVC KIT: 2L 7 FR X 20 CM/ASK-46702-PWBH1; MULTILUMEN CVC KIT: 2L 8 FR X 16 CM/CDC-42802-P1A; PI CVC KIT: 3-L 7 FR X 20 CM AGB/CDC-45703-PA1A; CVC KIT: 2L 8 FR X 6IN (16 CM)/ASK-42802-ECL4; PI CVC KIT: 3-L 7 FR X 16 CM AGB+/AK-42703-P1A; PI CVC KIT: 2-L 8 FR X 16 CM AGB+/AK-42802-P1A; PI CVC KIT: 4-L 8.5 FR X 16 CM AGB+/AK-42854-P1A; PI CVC KIT: 3-L 7 FR X 20 CM AGB+/AK-45703-P1A; PI CVC KIT: 2-L 8 FR X 20 CM AGB
Code information (models, lots, serial numbers)
Show the full code information (13k characters)
REF(Material)/UDI-DI/Lot(Batch): ASK-42802-PLH/10801902193275, 10801902133110, 10801902212013/33F25L0140 33F26B0222, 33F24F0799; AK-12123-F/10801902168464, 10801902193275, 10801902186338/33F24B0185, 33F24C0102, 33F24D0077, 33F24F0323, 33F24G0346, 33F24K0112, 33F25C0929, 33F25E0062, 33F25H0028, 33F25J0833, 33F24G0129; AK-15123-F/10801902168525, 10801902193275, 10801902186499/33F24C0108, 33F24E0349, 33F24G0353, 33F24L0650, 33F25A0478, 33F25C0779, 33F25E0068, 33F25M0946, 33F24E0713; ASK-12703-PMAH/10801902193275/33F24J0116; ASK-12703-PNM/10801902193275, 10801902214338/33F24L0197, 33F24D0725; ASK-12703-PUMH/10801902193275, 10801902164602/33F25L0030, 33F24F0914; ASK-15703-PRH/10801902193275, 10801902133226/33F25C0271, 33F25F0168, 33F25J0101, 33F25J0368, 33F26A1500, 33F24C0858; CDC-15703-P1A/10801902116793, 10801902193275/33F24D0980, 33F24E0236, 33F24E0832, 33F24G0252, 33F24H0074, 33F24H0401, 33F24H0457, 33F24J0195, 33F24K0024, 33F24M0016, 33F24M0017, 33F24M0064, 33F25B0116, 33F25C0577, 33F25D0719, 33F25G0427, 33F25G0441, 33F25G0663, 33F25H0361, 33F25J1039, 33F25M0495, 33F25M0512, 33F25B0670; NA-12703-PCDC/10801902193275, 10801902119978, 10801902119985/33F24G0555, 33F24K0298, 33F25A0602, 33F25E1055, 33F25H0180, 33F25J0942, 33F25L0474, 33F26B0598, 33F24D0414; NA-15703-PCDC/10801902119985, 10801902193275, 10801902119985, 10801902210217/33F24F0191, 33F24G0556, 33F24J0336, 33F24K0572, 33F25A0157, 33F25C0200, 33F25C0833, 33F25E0275, 33F25H0181, 33F25H0762, 33F25J0943, 33F25L0456, 33F26B0587, 33F25A0603; AK-25553/10801902193275, 10801902167801/33F24H0250, 33F24M0062, 33F25A0382, 33F25B0310, 33F25C0069, 33F25C0506, 33F26C0405, 33F24G0377; CDC-22402-X1A/)10801902193275/33F25K0177, 33F26B0893, 33F25E1192; CDC-22502-X1A/10801902193275/33F25F0216, 33F25C0331; CDC-25402-X1A/10801902193275, 20801902116974/33F25A0287, 33F25D0505, 33F25E0824, 33F25H0569, 33F26A1470, 33F26B0216, 33F24L0486; CDC-25502-X1A/10801902193275/33F24L0980, 33F25D0506, 33F25E0698, 33F25G0315, 33F25J0429, 33F26A1472, 33F24L0832; CDC-25553-X1A/10801902193275, 10801902168129, 60801902123611/33F25E0890, 33F25E1191, 33F26A1384, 33F26B0708, 33F24F0084; CDC-26553-X1A/10801902193275, 10801902168136, 50801902117019/33F25E0749, 33F25E1190, 33F25J0432, 33F26B0807, 33F26C0210, 33F26D0063, 33F24D0807; AK-22123-CDC/10801902193275, 10801902167740/33F24K0127, 33F25A0483, 33F25E0076, 33F25E0938, 33F25H0032, 33F25J0480, 33F25K0398, 33F25M0960, 33F24E0358; AK-22123-F/10801902193275/33F24G0362, 33F24J0526, 33F24K0128, 33F24L0762, 33F25C0963, 33F25E0939, 33F25H0033, 33F24G0363; AK-22402/10801902193275, 10801902167764/33F24L0704, 33F25A0484, 33F25C0817, 33F25E0079, 33F25H0035, 33F25L0373, 33F25M0879, 33F24B0203; AK-22502/10801902193275, 10801902167764/33F24E0363, 33F24G0366, 33F24K0131, 33F24L0708, 33F25C0885, 33F25J0842, 33F24K0487; AK-24301-SK/10801902193275/33F24G0367, 33F24J0528, 33F24K0132, 33F24L0680, 33F25C0118, 33F25C0775, 33F25E0080, 33F25G0092, 33F25H0036, 33F25H0724, 33F25J0482, 33F25L0411, 33F25M0915, 33F24K0133; AK-24402/10801902193275/33F24L0059, 33F24L0727, 33F25C0119, 33F25C0976, 33F25E0081, 33F24K0134; AK-24703-ASK/10801902193275, 10801902118971, 60801902123178/33F24L0778, 33F25C0928, 33F25E0082, 33F25H0037, 33F25J0843, 33F26B0495, 33F24F0281; AK-25123-CDC/10801902193275/33F24K0136, 33F24L0730; AK-25123-F/10801902193275/33F25C0913, 33F25E0085, 33F25E0943, 33F25J0484, 33F26B0496, 33F24G0374; AK-25402/10801902193275, 10801902167795/33F24G0376, 33F24J0282, 33F24K0139, 33F24L0752, 33F25C0121, 33F25C0968, 33F25G0093, 33F25J0486, 33F25L0381, 33F26B0532, 33F24G0688; AK-25502/10801902193275, 10801902180381/33F24H0051, 33F24M0001, 33F25A0125, 33F25B0333, 33F25F0256, 33F25F0500, 33F25H0807, 33F25K0228, 33F25L0163, 33F26A0051, 33F26B0266, 33F24G0689; AK-26553/10801902193275, 20801902117506/33F24K0140, 33F24L0776, 33F25E0089, 33F25G0094, 33F25J0847, 33F25L0410, 33F25M0914, 33F24D0024; CDC-22123-1A/10801902168044, 10801902193275/33F24F0943, 33F24L0472, 33F25A0325, 33F25B0672, 33F25E0653, 33F25E0877, 33F25M0114, 33F25M0424, 33F26A1508, 33F25A0326; CDC-22123-X1A/10801902193275, 20801902116851, 20801902116851,10801902213591/33F25B0673, 33F25D0446, 33F25D0667, 33F25F0212, 33F25G0413, 33F25J0317, 33F25M0396, 33F26B0442, 33F24D0842; CDC-24301-1A/30801902116902, 10801902193275, 30801902116902, 30801902116919/33F24D0874, 33F24F0076, 33F24F0863, 33F24K0501, 33F24L0979, 33F25B0617, 33F25D0494, 33F25D0668, 33F25E0697, 33F25E0820, 33F25H0571, 33F25J0249, 33F25K0181, 33F25L0084, 33F25L0104, 33F26A1413, 33F26B0784, 33S25E0697, 33F24C0702; AK-22702-P1A/10801902193275, 30801902122378/33F25D0322, 33F25H0424, 33F25J0097, 33F25K0085, 33F24E0364; CDC-22702-XP1A/10801902193275, 30801902116896/33F25D0451, 33F25F0217, 33F25H0573, 33F25J0322, 33F26A1386, 33F26B0706, 33F25K0180; ASK-45703-CPR/10801902193275/33F24J0477, 33F24L1068; ASK-42703-AKP1A/10801902193275, 10801902202519/33F24J0324, 33F25A0577, 33F25C0841, 33F25E0233, 33F25E1027, 33F25G0140, 33F25H0148, 33F25H0754, 33F25J0924, 33F25M0869, 33F24F0754; ASK-42703-KH/10801902193275, 10801902205657/33F24G0269, 33F24L0087, 33F25E0235, 33F25G0141, 33F25L0213, 33F24D0891; ASK-42703-MGH1/10801902193275, 10801902223682, 10801902193275/33F25J0735, 33F26C0344, 33F25E0509; ASK-42703-PCCH2/10801902193275, 10801902141030/33F24L0231, 33F24L0416, 33F24L0945, 33F25B0443, 33F25B0561, 33F25D0352, 33F25E0726, 33F25G0316, 33F25J0155, 33F25M0013, 33F26B0792, 33F26C0228, 33F25D0540; ASK-42703-PFS1/10801902193275/33F25E0598, 33F25B0275; ASK-42703-PNO2/10801902193275/33F26A1661, 33F24L0335; ASK-42703-PSCH1/10801902137330, 10801902193275/33F24F1038, 33F24H0204, 33F25B0566, 33F25H0491, 33F25J0274, 33F25M0092, 33F26B0124, 33F24H0100; ASK-42802-AKP1A/10801902193275, 10801902217155/33F24G0511, 33F24K0264, 33F24L0745, 33F25C0931, 33F25H0150, 33F25J0561, 33F25L0481, 33F25M0971, 33F24F0676; ASK-42802-LMDTG1/10801902193275, 10801902205701/33F26A1254, 33F26B0814, 33F24H0215; ASK-42802-PUCI1/10801902193275/33F24L0195; ASK-42802-PUMH/10801902193275, 10801902164251/33F25H0493, 33F25K0059, 33F25L0141, 33F25M0674, 33F24D0712; ASK-42802-UMM/10801902159653, 10801902193275/33F24F1027, 33F24G0513; ASK-42854-AKP1A/10801902193275, 10801902202540/33F24J0635, 33F24K0265, 33F24L0610, 33F25A0580, 33F25E0239, 33F25H0152, 33F25J0562, 33F24E0632; ASK-42854-MGH/10801902223668, 10801902193275/33F26C0876, 33F25E1198; ASK-45703-AKP1A/10801902193275, 10801902158120/33F24G0515, 33F24J0636, 33F24K0266, 33F24L0577, 33F25A0581, 33F25C0843, 33F25E0241, 33F25G0142, 33F25H0154, 33F25J0564, 33F25L0238, 33F25M0902, 33F24F0106; ASK-45703-MGH/10801902223675/33F26C0875, 33F25B0580; ASK-45703-PCCH2/10801902193275, 10801902141047, 10801902193275/33F24L0356, 33F24L1092, 33F25B0582, 33F25D0550, 33F25E0615, 33F25G0352, 33F25J0163, 33F25K0154, 33F26B0772, 33F24L0088; ASK-45703-PECL1/10801902193275, 10801902130386/33F25E0618, 33F25F0204, 33F25G0289, 33F25J0167, 33F25M0192, 33F26B0191, 33F26B0765, 33F25C0313; ASK-45703-PFS1/10801902193275/33F25E0510, 33F25B0274; ASK-45703-PHM1/10801902193275, 10801902116229/33F25E0620, 33F25G0290, 33F25H0572, 33F25J0405, 33F26B0715, 33F26B0854, 33F25K0313; ASK-45703-PHU/10801902193275/33F25L0135, 33F26A0243, 33F25G0685; ASK-45703-PMSG1/10801902193275, 30801902116315/33F25E1199, 33F25F0357, 33F26A0411, 33F24D0674; ASK-45703-PNCB/30801902116315, 10801902193275/33F24F0818, 33F24L0256, 33F24L1075, 33F25B0592, 33F25D0486, 33F25E0731, 33F25H0433, 33F25M0159, 33F25M0426, 33F26B0760, 33F25B0456; ASK-45703-PPCM/10801902193275, 10801902116335/33F25F0358, 33F25G0688, 33F25K0048, 33F25L0028, 33F26B0903, 33F24K0067; ASK-45703-PRH/10801902193275, 10801902116359, 10801902116373/33F25A0275, 33F25C0318, 33F25E0807, 33F25G0358, 33F25J0301, 33F26A1420, 33F26B0799, 33F24E0560; ASK-45703-PSJ2/10801902193275, 10801902119732/33F25C0319, 33F25E0733, 33F25G0357, 33F25H0641, 33F25J0302, 33F26A1507, 33F26B0878, 33F24L0259; ASK-45703-PSVM/10801902119749, 10801902193275, 10801902119749/33F24F0755, 33F24L0446, 33F25B0459, 33F25C0320, 33F25D0551, 33F25F0120, 33F25G0356, 33F25J0303, 33F26A1523, 33F26B0762, 33F25D0237; ASK-45703-PUHS/10801902193275, 10801902114058/33F24L0942, 33F25B0596, 33F25E0687, 33F25G0303, 33F25H0642, 33F26A1424, 33F24F0717; ASK-45703-PUKL/10801902193275, 10801902114058, 10801902193275/33F24H0185, 33F24K0433, 33F24L0262, 33F25B0597, 33F25D0372, 33F25E0628, 33F25H0566, 33F25J0173, 33F26B0233, 33S25E0628, 33F25B0461; ASK-45703-PUMF1/10801902193275/33F25K0060, 33F25L0134, 33F26A0235, 33F24K0567; ASK-45703-PUR1/10801902116465, 10801902193275/33F24F0790, 33F24H0206, 33F24L0266, 33F24L0922, 33F25B0601, 33F25D0375, 33F25E0630, 33F25E0870, 33F25J0408, 33F25M0128,33F26A1432, 33F26B0836, 33F24G0616; ASK-45703-RHR1/10801902116465, 10801902193275/33F24F0790, 33F24H0206, 33F24L0266,33F24L0922, 33F25B0601,33F25D0375, 33F25E0630, 33F25E0870, 33F25J0408, 33F25M0128, 33F26A1432, 33F26B0836, 33F24G0616; ASK-45703-WA/10801902193275, 10801902116564, 30801902119774/33F25A0443, 33F25E0011, 33F25K0309, 33F26B0909, 33F24C0746; ASK-45802-AKP1A/10801902193275, 30801902119774/33F24L0023, 33F26A1657, 33F26B0277, 33F25D0034; ASK-45854-AKP1A/10801902193275, 30801902119781/33F24G0671, 33F24L0039, 33F25A0366, 33F25B0094, 33F25D0715, 33F25G0495, 33F25J0033, 33F25K0349, 33F25L0226, 33F26A1654, 33F26B0025, 33F26B0698, 33F24J0130; ASK-45854-MGH/10801902223699, 10801902193275/33F26C0688, 33F25C0363; ASK-45854-PCCH2/10801902193275/33F25J0234, 33F25E1175; ASK-45854-PNM/10801902193275/33F24K0570, 33F25B0431, 33F25E0006, 33F25F0377, 33F25H0780, 33F25L0261; ASK-45854-PPCM/10801902143867/33F24F0736; ASK-45854-PUCH2/10801902193275/33F24M0177; ASK-46702-PWBH1/10801902193275/33F24K0321, 33F24K0269; CDC-42802-P1A/10801902170870, 10801902193275, 10801902170870/33F24F0244, 33F24H0070, 33F24H0426, 33F24J0223, 33F24J0468,33F24L0009,33F24M0079, 33F24M0139, 33F25B0319, 33F25C0533, 33F25E1094, 33F25E1156,33F25F0384, 33F25H0338, 33F25J0766, 33F25J0990, 33F25L0086, 33F25M0016, 33F26A1619, 33F26B1051, 33F25H0880; CDC-45703-PA1A/10801902193275, 30801902117190, 10801902117202/33F24M0081, 33F25F0241, 33F26A1656, 33F26B0334, 33F24C1072; ASK-42802-ECL4/10801902224283, 10801902205664/33F26B0489, 33F24D0651; AK-42703-P1A/10801902193275, 10801902171372/33F25D0323, 33F25D0585, 33F25F0122, 33F25F0153, 33F25G0376, 33F25J0135, 33F26B0459, 33F24L1104; AK-42802-P1A/10801902193275/33F25B0539, 33F25D0386, 33F25F0137, 33F25J0136, 33F25B0512; AK-42854-P1A/10801902193275, 10801902168167, 10801902119152/33F25E0664, 33F25H0442, 33F25J0341, 33F26B0752, 33F23H0557; AK-45703-P1A/10801902193275, 10801902171389/33F25C0341, 33F25D0387, 33F25D0643, 33F25F0154, 33F25G0257, 33F25G0375, 33F25J0137, 33F25J0342, 33F26A1416, 33F26B0150, 33F26B0766, 33F25C0191; AK-45802-P1A/10801902193275, 10801902168181/33F25E0665, 33F25J0191, 33F26B0754, 33F25A0223; AK-45854-P1A/10801902193275, 10801902168198/33F25D0459, 33F25F0138, 33F26B0145, 33F25B0540; AK-47702-P1A/10801902193275/33F25E0502, 33F25G0162, 33F25H0407, 33F25J0343, 33F24G0383; ASK-42703-PECL3/10801902193275/33F26A1397, 33F25B0562; ASK-42703-PUPM3/10801902225150, 10801902156966/33F26C0238, 33F24F0977; ASK-42802-PUPM3/10801902225143, 10801902193275/33F26C0239, 33F25B0498; ASK-45703-PECL3/10801902193275, 10801902224313/33F26A1577, 33F26B0773, 33F25C0234; ASK-45703-PUPM3/10801902225181, 10801902116465/33F26C0219, 33F24D0840; ASK-45854-PECL3/10801902193275/33F24L0191; ASK-45854-PUPM3/10801902225167, 10801902193275/33F26C0240, 33F24H0111; ASK-45854-SMW/10801902224436, 10801902214888/33F26B0830, 33F24F0872; CDC-42802-XP1A/10801902193275, 10801902168365, 10801902162103/33F25J0295, 33F25J0728, 33F25K0304, 33F25K0390, 33F25L0228, 33F25M0037, 33F25M0485, 33F25M0717, 33F26A1630, 36F25C0008, 36F25D0001, 36F25D0034, 36F25E0001, 36F25G0022, 36F25H0015, 33F24D0258; CDC-42854-XP1A/10801902193275, 20801902117148, 10801902162134/33F25J0681, 33F26A0390, 33F26C0209, 33F26D0062, 36F25D0003, 36F25H0014, 33F24D0827; CDC-45703-XP1A/10801902193275, 20801902117216/33F25G0286, 33F25H0876,33F25J0089,33F25J0682, 33F25J0702, 33F25J0705, 33F25J0711,33F25J0718, 33F25J0729, 33F25J0733, 33F25K0305, 33F25L0063,33F25L0093, 33F25L0094, 33F25L0231, 33F25L0252, 33F25L0253, 33F25L0254, 33F25L0286, 33F25L0302, 33F25L0322, 33F25M0487, 33F26A0062, 33F26B0637, 33F26B0638, 33F26D0185, 36F25A0008, 36F25A0014, 36F25D0010, 36F25D0011, 36F25D0014, 36F25D0021, 36F25D0022, 36F25D0023, 36F25D0028, 36F25D0041, 33F25C0372; CDC-45854-XP1A/10801902193275, 10801902169720, 10801902162158/33F25J0695, 33F25J0730, 33F25K0394, 33F26A1620, 33F26B0572, 33F26B0735, 36F25G0003, 36F25G0021, 36F25G0024, 36F25H0011, 36F25H0017, 33F24D0424 ASK-42703-PGM3/10801902213751/33F24F0951, 33F24K0743; ASK-42802-UH1/10801902156386/33F24H0122; ASK-45703-PGM3/10801902213768/33F24D0844, 33F24F0840;
Devices named in this recall
87 devices in the FDA device registry whose identifier appears in the recall's code information.
- ARROW IPN925037Teleflex Incorporated · CATHETER, CONDUCTION, ANESTHETIC7 recalls0 reports
- ARROW IPN038522Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days3 recalls1 report
- ARROW IPN931283Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days3 recalls0 reports
- ARROW IPN931284Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days3 recalls0 reports
- ARROW IPN931286Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days3 recalls0 reports
- ARROW IPN036445Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls8 reports
- ARROW IPN915628Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls5 reports
- ARROW IPN916994Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls4 reports
- ARROW IPN036830Teleflex Incorporated · INTRODUCER, CATHETER2 recalls3 reports
- ARROW IPN036452Teleflex Incorporated · INTRODUCER, CATHETER2 recalls3 reports
- ARROW IPN915630Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls2 reports
- ARROW IPN036474Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 report
- ARROW IPN904896Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 report
- ARROW IPN038523Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls1 report
- ARROW IPN923102Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls1 report
- ARROW IPN927593Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls1 report
- ARROW IPN036596Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls0 reports
- ARROW IPN037185Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls0 reports
- ARROW IPN037016Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls0 reports
- ARROW IPN037790Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls0 reports
- ARROW IPN055589Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN915625Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN056570Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN916803Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN917020Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN917026Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN919358Teleflex Incorporated · Catheter, hemodialysis, triple lumen, non-implanted2 recalls0 reports
- ARROW IPN922680Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN922682Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN923103Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN927740Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPNB00178Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPNB00181Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPNB00182Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPNB00180Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN036428Teleflex Incorporated · CATHETER, HEMODIALYSIS, NON-IMPLANTED2 recalls0 reports
- ARROW IPN036437Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall33 reports
- ARROW IPN927331Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall3 reports
- ARROW IPN037257Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall1 report
- ARROW IPN037093Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall1 report
- ARROW IPN036540Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall1 report
- ARROW IPN036988Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall1 report
- ARROW IPN036502Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN036534Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN036601Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN036495Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN036566Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN037155Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN037069Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN037166Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN037691Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN038176Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN029652Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN055555Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN914810Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN055840Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN056641Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN916989Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN916991Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN916996Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN916812Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN916813Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN916816Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN916818Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN916819Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN916836Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN917104Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN917373Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN917134Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN917135Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN916995Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN919342Teleflex Incorporated · Catheter, hemodialysis, triple lumen, non-implanted1 recall0 reports
- ARROW IPN923110Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN926612Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN927330Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN928775Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN931285Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPNB00080Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPNB00093Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN036512Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN036454Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN036510Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN036493Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN036468Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN037015Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN037130Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN036961Teleflex Incorporated · Catheter, hemodialysis, triple lumen, non-implanted1 recall0 reports
Product lines
Other devices under the same FDA submission
231 devices cleared under K071538, K900263, K970864, K993691 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ARROW IPN030177Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days254 reports
- ARROW IPN030162Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days163 reports
- ARROW IPN055818Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls1 report
- ARROW IPN055820Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls1 report
- ARROW IPN056571Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls1 report
- ARROW IPN917372Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls1 report
- ARROW IPN055870Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN056182Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN922908Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- Medline Industries, Inc. DYNDM1031AMedline Industries, Inc. · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls0 reports
- ARROW IPN915614Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall48 reports
- ARROW IPN030129Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days71 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98793).
- Z-2363-2026 · Open, ClassifiedPARACENTESIS KIT, REF: AK-00376, ASK-00376-VCU
- Z-2364-2026 · Open, ClassifiedDescription/REF: ACCESS TRAY/ASK-04001-SUH-S; ACCESS TRAY/ASK-04001-VM1; STIMUCATH SFTY KIT: 19G CTH, 17Gx4CM NDL/AB-19604-KS; STIMUCATH SFTY KIT: 19G CTH, 17Gx8CM NDL/AB-19608-KS;…
- Z-2365-2026 · Open, ClassifiedDescription/REF: BIOPSY SYSTEM COMPREHENSIVE KIT/ASK-09458-AH, ASK-09458-COC, ASK-09458-EUH, ASK-09458-SD, ASK-09458-VU
- Z-2366-2026 · Open, ClassifiedDescription/REF: PACING PSI KIT: 6 FR/AI-06210-KS, PACING PSI KIT: 5 FR 2-L/AI-07155-KS, PACING PSI KIT: 5 FR 2L/ASK-07155-UPM
- Z-2367-2026 · Open, ClassifiedDescription/REF: AGBA PICC G4 Stylet: 3L 6FR x 55CM BP/DLX-45563-VPSB; AGBA PICC G4 Stylet: 3L 6FR x 55CM TCG/DLX-45563-VPSC; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-UCHD1; ARROW(R) V…
- Z-2368-2026 · Open, ClassifiedDescription/REF: RA CATH KIT: 20 GA X 1 3/4 IN/TJ-04020-1; RA CATH KIT: 20 GA X 1 1/2IN/NA-04220-S1A; ARTERIAL LINE KIT/AK-04510-S; ARTERIAL CATH KIT: 18 GA X 6IN (16 CM)/ASK-04018…
- Z-2369-2026 · Open, ClassifiedDescription/REF: PSI KIT: 7 FR X 4IN ANTIMICROBIAL/ASK-09701-UMH; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-09803-OSU4; PSI/CVC KIT: 9 FR/7 FR X 6IN (16CM)/ASK-09903-AMC1; PSI KIT: 9 …
- Z-2370-2026 · Open, ClassifiedDescription/REF: PI AGBA PICC KIT: 3L 6 FR X 40 CM W/VPS/ASK-44063-NS AGBA PICC/DELTA: 2L 5.5 FR X 50 CM/ASK-45052-NS AGBA PICC/DELTA KIT: 1L 4.5 FR X 55 CM/ASK-45541-MHBV PICC KIT…
- Z-2371-2026 · Open, ClassifiedDescription/REF: EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05502-VM1; EPIDURAL CATHETERIZATION KIT: 19 GA/AK-05501; EPIDURAL CATHETERIZATION KIT/AK-05502; EPIDURAL CATHETERIZATION K…
- Z-2372-2026 · Open, ClassifiedDescription/REF: HEMODIALYSIS KIT: 2L 14 FR X 20 CM AGB CDC-25142-X1A
- Z-2373-2026 · Open, ClassifiedDescription/REF: HEMODIALYSIS KIT: 2-LUMEN 8 FR X 11 CM/AK-11802-F; HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-22122; HEMODIALYSIS KIT 2L: 14 FR X 15CM AGB/AK-22142-CDC; HEMODIALYS…
- Z-2374-2026 · Open, ClassifiedDescription/REF: AHDC KIT: 3-L 12 FR X 16 CM/AK-12123-PCDC; AHDC KIT: 3-L 12 FR X 20 CM/AK-15123-PCDC; AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XP1A; AHDC KIT: 3-L 12 FR X 25 CM/CDC-1…
- Z-2377-2026 · Open, ClassifiedDescription/REF: ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM6 RA CATH KIT: 20 GA X 1-3/4IN/NA-04020-X1ASP ACCESS TRAY/ASK-04001-MS RA CATH KIT: 20 GA X 8 CM/ASK-00820-UM1 VESSE…
- Z-2378-2026 · Open, ClassifiedDescription/REF: PI PICC: 3: 6FR x 55CM w CHG/ DLX-35563-HPKC; AGBA PICC: 2L 5.5 FR X 45 CM/ASK-44552-ABKB; AGBA PICC/DELTA KIT: 5.5 FR X 55 CM/ASK-45552-MHBV; AGBA PICC: 1L 4.5FR …
- Z-2379-2026 · Open, ClassifiedDescription/REF: PI JACC KIT: 2L 5.5 FR X 15 CM CG+/ASK-41552-LTAC; PI JACC KIT: 3L 6 FR X 20 CM CG+/ASK-42063-LTAC; AGBA PI JACC KIT: 1-L 4.5 FR X 15 CM/CDC-41541-JX1A; AGBA PI JA…
- Z-2380-2026 · Open, ClassifiedDescription/REF: PERITONEAL LAVAGE CATH KIT:8FR/AK-09000
- Z-2381-2026 · Open, ClassifiedDescription/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-12122-F
- Z-2382-2026 · Open, ClassifiedDescription/REF: DRAINAGE KIT/AK-01600, DRAINAGE KIT/AK-01601, DRAINAGE KIT/ASK-01601-MM, THORACENTESIS KIT/AK-01000-T, THORACENTESIS KIT/AK-01000, THORACENTESIS KIT/ASK-01000-CHP,…
- Z-2383-2026 · Open, ClassifiedDescription/REF: PACING/PSI KIT: 5 FR 2L/ASK-07155-MGU
- Z-2384-2026 · Open, ClassifiedDescription/REF: ACCESS TRAY/ASK-04001-JHH; ACCESS TRAY/ASK-04001-LMDTG; PI CVC KIT: 3 LUMEN 7 FR X 6 IN (16 CM)/ASK-12703-PSPH1; 3L CVC KIT: 7 FR X 16 CM/ASK-12703-WMC2; MULTILUME…
- Z-2385-2026 · Open, ClassifiedDescription/REF: ACCESS TRAY/GU-04020; PSI KIT: 8.5 FR/AK-09803-CDC; PSI KIT: 9 FR/AK-09903-CDC; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-09810-PCMH1; PSI KIT: 9 FR X 10 CM/ASK-09903-KH;…
- Z-2387-2026 · Open, ClassifiedDescription/REF: PI MIDLINE 2L: 5FR X 20CM/CDC-32052-MPKC
- Z-2388-2026 · Open, ClassifiedDescription/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-15122-F, AHDC KIT: 3L 12 FR X 16 CM/ASK-12123-UOC1, AHDC KIT: 3-L 12 FR X 16 CM/ASK-12123-UPM, AHDC KIT: 3L 12 FR X 20 C…
- Z-2389-2026 · Open, ClassifiedDescription/REF: SPINAL ANESTHESIA KIT/ASA-24090-SB; SPINAL ANESTHESIA KIT/ASA-24103-SB; SPINAL ANESTHESIA KIT/ASA-25090-SB; SPINAL ANESTHESIA KIT: 24 GA X 9 CM/ASK-24090-SA; SPINA…
- Z-2390-2026 · Open, ClassifiedDescription/REF: SAC KIT: 18 GA X 12CM W BIO/SAC-01218-B; SAC KIT: 18 GA X 12CM W CHG/SAC-01218-C; SAC KIT: 18 GA X 16CM W BIO/SAC-01618-B
- Z-2391-2026 · Open, ClassifiedDescription/REF: RA CATH KIT: 20 GA X 1-3/4IN/AK-04020-SPC; RA CATH KIT: 20 GA X 5 IN/ASK-04020-MW; ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM5; CATH SL 20 GA X 5 CM HYDROPHIL…
- Z-2393-2026 · Open, ClassifiedDescription/REF: PSI KIT: 7 FR X 4 IN/ASK-09701-UP; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-09803-OSU5; PSI KIT: 8.5 FR/ASK-09807-UC; PSI KIT: 9 FR X 4 1/8IN (10 CM)/ASK-09903-HH2; …
- Z-2394-2026 · Open, ClassifiedDescription/REF: MIDLINE CATHETER KIT: 3 FR X 20 CM/CDC-02031-MK1A; AM/AT PI MIDLINE 1L: 4.5FR X 15CM/CDC-41541-MPK1A; AM/AT PI MIDLINE 2L: 5.5FR X 15CM/CDC-41552-MPK1A; AGBA PI MI…
- Z-2396-2026 · Open, ClassifiedDescription/REF: CVC KIT: 16 GA X 8IN/AK-04210, CVC KIT: 1-L 16 GA X 8IN (20 CM)/AK-04300, CVC KIT: 14 GA X 6 IN/CK-04711
- Z-2397-2026 · Open, ClassifiedDescription/REF: RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-1A; RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-X1A; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-HCA; ARTERIAL CATHETER KIT: 18 …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page