Skip to content

ACUMED 46-0040-SRecalls & reported problems

Acumed LLC · Acu-Sinch Kntls Mini Sys w/Insrtr 2.3 · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10806378124235), received between Sep 19, 2025 and May 14, 2026 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Off-Label Use” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 4 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Off-Label Use4
  • Implant subsidence1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Implant Pain1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 14, 2026Implant subsidence

    IT WAS REPORTED: 78F PRESENTED WITH PAIN AND THUMB SUBSIDENCE POST-OP. X-RAY SHOWED ROUND BUTTON PULL THROUGH. PATIENT IS OSTEOPOROTIC. PATIENT PRESENTED WITH PAIN PRIOR TO THE 2 WEEK MARK.

    Patient problems: Implant Pain

    MAUDE report 25178711 ↗
  • MalfunctionOct 15, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE ACUSINCH MINI USED ACCORDING TO TECHNIQUE GUIDE, TRIGGER PULLED, FLIP BUTTON SUCCESSFULLY DEPLOY, PULLED BACK ON BLUE HANDLE DEVICE, FLIP BUTTON FELL THROUGH THE DRILL HOLE OPENED SECOND DEVICE TO COMPLETE SURGERY - 12 MINUTE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23292238 ↗
  • MalfunctionSep 19, 2025Off-Label Use; Off-Label Use

    IT WAS REPORTED THROUGH COMMUNICATIONS WITH ACUMED SALES REPRESENTATIVES AND DISTRIBUTORS THAT THE PRODUCT WAS USED IN AN OFF-LABEL CAPACITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23100997 ↗
  • MalfunctionSep 19, 2025Off-Label Use; Off-Label Use

    IT WAS REPORTED THROUGH COMMUNICATIONS WITH ACUMED SALES REPRESENTATIVES AND DISTRIBUTORS THAT THE PRODUCT WAS USED IN AN OFF-LABEL CAPACITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23100995 ↗
  • MalfunctionSep 19, 2025Off-Label Use; Off-Label Use

    IT WAS REPORTED THROUGH COMMUNICATIONS WITH ACUMED SALES REPRESENTATIVES AND DISTRIBUTORS THAT THE PRODUCT WAS USED IN AN OFF-LABEL CAPACITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23100994 ↗
  • MalfunctionSep 19, 2025Off-Label Use; Off-Label Use

    IT WAS REPORTED THROUGH COMMUNICATIONS WITH ACUMED SALES REPRESENTATIVES AND DISTRIBUTORS THAT THE PRODUCT WAS USED IN AN OFF-LABEL CAPACITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23100985 ↗

Other WASHER, BOLT NUT devices

All WASHER, BOLT NUT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page