Skip to content

Axonics 9009Recalls & reported problems

Axonics Modulation Technologies, Inc. · Percutaneous Extension · Class 3

Adverse event reports
5
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10810005340288), received between Sep 21, 2020 and Sep 4, 2026 — classified as 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Disconnection” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Disconnection1
  • High impedance1
  • Material Integrity Problem1
  • Migration1

Patient problems

  • Drug Resistant Bacterial Infection1
  • Incontinence1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1
  • Therapeutic Response, Decreased1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjurySep 4, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient with this Trial Stimulator had an overnight hospital stay due to extreme pain. They received pain medication for it. The patient has unknown etiology back pain prior to the trial procedure and the procedure magnified the existing pain. When trying to determine where the pain was, the patient stated the pain was not where they felt the stimulation. The pain was near the pocket and wrapped around their back like a band. The patient did not explain how the pain felt. Additionally, the pain was controlled and the patient moved forward with the Stage 2 implant.

    Patient problems: Pain

    MAUDE report 50602217 ↗
  • InjuryDec 26, 2025High impedance; Material Integrity Problem; High impedance; Material Integrity Problem

    IT WAS REPORTED THAT DURING THE INITIAL IMPLANT PROCEDURE FOR THE EXTERNAL NEUROSTIMULATOR, THE TINED LEAD INITIALLY TESTED WITHIN RANGE, BUT REPEATED IMPEDANCE CHECKS SHOWED OPEN VALUES ON ALL ELECTRODES. MULTIPLE RECONNECTIONS AND IMPEDANCE TESTS WERE ATTEMPTED WITHOUT RESOLUTION. THE PHYSICIAN REOPENED THE INCISION, CONFIRMED INTACT LEAD INTEGRITY. THE PERCUTANEOUS EXTENSION WAS REPLACED AND ALL IMPEDANCES WERE TESTED WITHIN NORMAL LIMITS. THE CASE WAS COMPLETED USING THE REPLACEMENT EXTENSION. THE PATIENT HAD BEEN DOING WELL AND REPORTED BENEFITING FROM THE THERAPY. THIS EVENT OCCURRED DURING THE STAGE 1 TRIAL. NO FURTHER CLINICAL COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23899171 ↗
  • InjuryDec 9, 2024Migration; Migration

    THE PATIENT UNDERWENT REVISION SURGERY DUE TO MIGRATION.

    Patient problems: Incontinence

    MAUDE report 20885879 ↗
  • InjuryAug 3, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE COMPANY WAS MADE AWARE THAT A PATIENT UNDERWENT AN EXPLANT SURGERY DUE TO MRSA INFECTION.

    Patient problems: Drug Resistant Bacterial Infection

    MAUDE report 15156993 ↗
  • MalfunctionSep 21, 2020Disconnection; Disconnection

    ON (B)(6) 2020, CLINICAL SPECIALIST REPORTED THAT TRIAL LEADS HAD DISCONNECTED FROM THE PERCUTANEOUS EXTENSION WHILE HOME CARE NURSE WAS REMOVING THE DRESSINGS. THE PATIENT IS PARAPLEGIC. THE DISCONNECTION DAMAGED THE EXTENSION CABLE WHEN THE PATIENT WAS LIFTED ONTO THE HOYER LIFT. THE EXTENSION WAS REPLACED AND PATIENT IS DOING WELL.

    Patient problems: Therapeutic Response, Decreased

    MAUDE report 10560925 ↗

Other Implanted Electrical Device Intended For Treatment Of Fecal Incontinence devices

All Implanted Electrical Device Intended For Treatment Of Fecal Incontinence devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.