Skip to content

Monarch Platform, Bronchoscopy 4.1 MON-000008Recalls & reported problems

Auris Health, Inc. · The Monarch Bronchoscope enables electro-mechanical articulation and precise control of a flexible endoscope (bronchoscope) under continuous and direct control by a physician operator. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10810068810988), received between Feb 26, 2026 and Aug 27, 2026 — classified as 2 injuries.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 21, 2026Class IIOpen, Classified

    Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.

    Z-1383-2026 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • Ischemic Heart Disease1
  • Pneumothorax1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 27, 2026Adverse Event Without Identified Device or Use Problem

    During a robot-assisted bronchoscopy, anesthesia noted ST-segment elevation after approximately six needle biopsy passes. The procedure was paused, and evaluation for pneumothorax was negative. An EKG assessment was initiated but temporarily interrupted by a building power outage. The patient was transferred to the Cath Lab, evaluated, and cleared. The patient had a history of cardiac issues and stent placement approximately eight months earlier. No device issues were reported. The procedure was aborted, and the patient was discharged home on (b)(6) 2026.

    Patient problems: Ischemic Heart Disease

    MAUDE report 50480931 ↗
  • InjuryFeb 26, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A MONARCH BRONCHOSCOPY PROCEDURE, THE PATIENT DEVELOPED A PNEUMOTHORAX THAT REQUIRED CHEST TUBE. NO DEVICE ISSUES WERE REPORTED.

    Patient problems: Pneumothorax

    MAUDE report 24457326 ↗

Other Bronchoscope (Flexible Or Rigid) devices

All Bronchoscope (Flexible Or Rigid) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page