OverTube Endoscopic Access System OVT-027-160Recalls & reported problems
Apollo Endosurgery, Inc. · OverTube (Non-Sterile) · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10811955020701), received between Jan 10, 2020 and May 28, 2026 — classified as 2 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Material Integrity Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Integrity Problem4
- Device-Device Incompatibility3
- Adverse Event Without Identified Device or Use Problem2
- Crack1
- Device Slipped1
- Inflation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Perforation of Esophagus2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 28, 2026Material Integrity Problem
NOTE: THIS RECORD REPRESENTS THE THIRD OF THREE OVERTUBE DEVICES. IT WAS REPORTED TO BOSTON SCIENTIFIC THAT AN OVERTUBE WAS USED IN AN ENDOSCOPIC SLEEVE GASTROPLASTY (ESG) PROCEDURE ON (B)(6) 2026, FOR THE TREATMENT OF OBESITY. DURING THE PROCEDURE, THE INFLATION CUFFS ON THREE OVERTUBE DEVICES WOULD NOT INFLATE. THE PROCEDURE WAS CANCELLED DUE TO PATIENT'S DISTORTED ANATOMY. NO PATIENT CONSEQUENCES WERE REPORTED AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25325163 ↗MalfunctionMay 28, 2026Material Integrity Problem
NOTE: THIS RECORD REPRESENTS THE FIRST OF THREE OVERTUBE DEVICES. IT WAS REPORTED TO BOSTON SCIENTIFIC THAT AN OVERTUBE WAS USED IN AN ENDOSCOPIC SLEEVE GASTROPLASTY (ESG) PROCEDURE ON (B)(6) 2026, FOR THE TREATMENT OF OBESITY. DURING THE PROCEDURE, THE INFLATION CUFFS ON THREE OVERTUBE DEVICES WOULD NOT INFLATE. THE PROCEDURE WAS CANCELLED DUE TO PATIENT'S DISTORTED ANATOMY. NO PATIENT CONSEQUENCES WERE REPORTED AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25325161 ↗MalfunctionMay 28, 2026Material Integrity Problem
NOTE: THIS RECORD REPRESENTS THE SECOND OF THREE OVERTUBE DEVICES. IT WAS REPORTED TO BOSTON SCIENTIFIC THAT AN OVERTUBE WAS USED IN AN ENDOSCOPIC SLEEVE GASTROPLASTY (ESG) PROCEDURE ON (B)(6) 2026, FOR THE TREATMENT OF OBESITY. DURING THE PROCEDURE, THE INFLATION CUFFS ON THREE OVERTUBE DEVICES WOULD NOT INFLATE. THE PROCEDURE WAS CANCELLED DUE TO PATIENT'S DISTORTED ANATOMY. NO PATIENT CONSEQUENCES WERE REPORTED AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25325160 ↗MalfunctionMay 9, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT ON (B)(6) 2024 THAT THE OVERTUBE AND NXT SYSTEM WOULD NOT FIT TOGETHER. ON (B)(6) 2024 THE PHYSICIAN REPORT WHEN TRYING TO INTUBATE WITH THE SCOPE AFTER RE-TAPING THE DEVICE TO THE SCOPE AND ADDING ADDITIONAL LUBRICATION, THAT THE SCOPE AND NXT WOULD NOT FIT THROUGH THE OVERTUBE. THE PHYSICIAN TRIED MULTIPLE ATTEMPTS. THE PHYSICIAN REMOVED THE OVERTUBE AND THE PROCEDURE WAS COMPLETED WITH THE NXT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19281823 ↗MalfunctionMay 8, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT ON APRIL 16, 2024 THAT THE OVERTUBE AND NXT SYSTEM WOULD NOT FIT TOGETHER. ON (B)(6) 2024 THE PHYSICIAN REPORT WHEN TRYING TO INTUBATE WITH THE SCOPE AFTER RE-TAPING THE DEVICE TO THE SCOPE AND ADDING ADDITIONAL LUBRICATION, THAT THE SCOPE AND NXT WOULD NOT FIT THROUGH THE OVERTUBE. THE PHYSICIAN TRIED MULTIPLE ATTEMPTS. THE PHYSICIAN REMOVED THE OVERTUBE AND THE PROCEDURE WAS COMPLETED WITH THE NXT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19268093 ↗MalfunctionMar 22, 2024Device-Device Incompatibility; Material Integrity Problem; Device-Device Incompatibility; Material Integrity Problem
NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURING REPORT NUMBER 3005099803-2024-01049 FOR THE OVERSTITCH NXT ENDOSCOPIC SUTURING SYSTEM AND 3005099803-2024-01220 FOR THE OVERTUBE ENDOSCOPIC ACCESS SYSTEM. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN OVERSTITCH NXT ENDOSCOPIC SUTURING SYSTEM WAS USED IN A TRANSORAL OUTLET REDUCTION (TORE) PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THERE WAS SOME RESISTANCE WHEN MANEUVERING THE OVERSTITCH THROUGH THE OVERTUBE AND IT WAS NOTED THAT THE OVERTUBE WAS TORN. IT WAS REPORTED THAT THERE WAS LESS INFLATION THAN THEY WANTED. IN THE PHYSICIAN'S ASSESSMENT THE RESISTANCE AND MANEUVERING MAY HAVE CAUSED THE TEAR. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH THESE DEVICES. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18959055 ↗InjuryApr 30, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON 31/MAR/2021 RECEIVED UF/IMPORTER REPORT # (B)(4). REPORTED: DURING THE INSERTION OF THE OR'S SCOPE INTO THE OVERTUBE DEVICE, AN ESOPHAGEAL PERFORATION WAS NOTED JUST DISTAL TO THE OVERTUBE. THE DEVICE WAS IMMEDIATELY REMOVED AND ENDOSCOPIC CLIPS WERE PLACED TO CLOSE THE PERFORATION.
Patient problems: Perforation of Esophagus
MAUDE report 11756200 ↗InjuryDec 10, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
LITERATURE REVIEW REPORTED: THERE WAS ONE CASE OF SUPERFICIAL ESOPHAGEAL TEAR FROM USING THE OVERTUBE WITH THE OVERSTITCH DEVICE WHICH WAS CONTROLLED ENDOSCOPICALLY WITHOUT NEED FOR REPEAT PROCEDURE.
Patient problems: Perforation of Esophagus
MAUDE report 10987184 ↗MalfunctionJan 10, 2020Crack; Inflation Problem; Device Slipped
THE INSIDE OF THE TUBE WAS CRACKED, CAUSING THE DEVICE TO NOT INFLATE AND THEREFORE, COULD NOT HOLD TUBE IN PLACE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9574572 ↗
Other Endoscopic Access Overtube, Gastroenterology-Urology devices
- EVIS EXERA II SMALL INTESTINAL VIDEOSCOPE OLYMPUS SIF TYPE Q180Olympus Medical Systems Corp. · endoscopic access overtube, gastroenterology-urology1 recall42 reports
- Trokamed WA2PS20STROKAMED GmbH · Endoscopic Access Overtube, Gastroenterology-Urology1 recall2 deaths2 reports
- Trokamed WA2PS20LTROKAMED GmbH · Endoscopic Access Overtube, Gastroenterology-Urology1 recall2 deaths2 reports
- Trokamed WA2PS18LTROKAMED GmbH · Endoscopic Access Overtube, Gastroenterology-Urology1 recall0 reports
- Trokamed WA2PS18STROKAMED GmbH · Endoscopic Access Overtube, Gastroenterology-Urology1 recall0 reports
- Navigator HD M0062502220Boston Scientific Corporation · endoscopic access overtube, gastroenterology-urology1 death31 reports
- Navigator HD M0062502230Boston Scientific Corporation · endoscopic access overtube, gastroenterology-urology29 reports
- Proxis™ Ureteral Access Sheath 231045C. R. Bard, Inc. · endoscopic access overtube, gastroenterology-urology22 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page