Animas Vibe 11451400Recalls & reported problems
Animas Corporation · Insulin Pump Replacement, Adult Pink · Class 3
All models of Animas Vibe (45)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Animas Vibe insulin pump has 784 FDA reports, including 749 malfunctions, 35 injuries and no deaths. Many reports describe cracks in the battery compartment, display errors and alarms that could not be cleared, and some mention moisture damage. Some injury reports mention high or low blood sugar. There are no recalls naming this device. The latest report is from August 2019.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 784
- Deaths reported
- 0
- Injuries reported
- 35
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 784 adverse event reports that list this exact device (DI 10840406100068), received between Aug 8, 2016 and Aug 28, 2019 — classified as 35 injuries, 749 malfunctions.
The most frequently reported specific device problem is “Structural Problem” (300 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Structural Problem300
- Device Displays Incorrect Message120
- Fail-Safe Problem120
- Electrical /Electronic Property Problem100
- Moisture Damage89
- Image Display Error/Artifact79
- Failure to Read Input Signal68
- No Display/Image36
Patient problems
- No Consequences Or Impact To Patient745
- Hyperglycemia29
- Hypoglycemia9
- Nausea9
- Polydipsia7
- Confusion/ Disorientation5
- Blurred Vision4
- Dehydration3
Recent adverse event reports
The 25 most recent of 784 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 28, 2019Structural Problem; Structural Problem
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A BATTERY COMPARTMENT CRACK. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 16-AUG-2019.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8939245 ↗MalfunctionAug 28, 2019Structural Problem; Structural Problem
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A BATTERY COMPARTMENT CRACK. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 15-AUG-2019.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8939215 ↗MalfunctionJul 24, 2019Structural Problem; Structural Problem
THE PUMP WAS RETURNED FOR INVESTIGATION. EVALUATION REVEALED A CRACK IN THE BATTERY COMPARTMENT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 19-JUL-2019.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8821134 ↗MalfunctionJul 10, 2019Image Display Error/Artifact; Image Display Error/Artifact
ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8776989 ↗MalfunctionJul 9, 2019Premature Discharge of Battery; Premature Discharge of Battery
ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS ALLEGING THE PUMP HAD A BATTERY LIFE ISSUE; FLUCTUATING BATTERY LIFE LEVELS. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE USER MAY BE UNAWARE THAT THE PUMP HAS LOST POWER, LEADING TO UNDER DELIVERY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8770883 ↗MalfunctionJun 5, 2019Device Displays Incorrect Message; Fail-Safe Problem; Device Displays Incorrect Message; Fail-Safe Problem
ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CS 078) ISSUE. THE ALARM WAS NOT ABLE TO BE CLEARED FROM THE PUMP. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG-TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8671281 ↗MalfunctionJun 5, 2019Moisture Damage; Moisture Damage
ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (MOISTURE INGRESS) ISSUE. THE REPORTER DENIED DAMAGE TO THE PUMP. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION CAN RESULT IN A DELAY IN TREATMENT OR LONG TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE POWER CIRCUIT OR CARTRIDGE COMPARTMENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8671279 ↗MalfunctionMay 6, 2019Structural Problem; Structural Problem
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A CASE/CONDITION ISSUE. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(6) 2019.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8582480 ↗MalfunctionApr 24, 2019Electrical /Electronic Property Problem; Structural Problem; Electrical /Electronic Property Problem; Structural Problem
ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (DAMAGE) ISSUE. THE REPORTER ALLEGED AN INTERMITTENT POWER ISSUE. THE REPORTER ALSO ALLEGED THE BATTERY COMPARTMENT WAS CRACKED. THE REPORTER DENIED DAMAGE TO THE BATTERY CAP AND STATED THE YELLOW O-RING OF THE BATTERY CAP WAS NOT VISIBLE WHEN THE BATTERY CAP IS SECURED TO THE PUMP. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE USER MAY BE UNAWARE THAT THE PUMP HAS LOST POWER, LEADING TO UNDER DELIVERY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8543830 ↗MalfunctionApr 11, 2019Structural Problem; Structural Problem
ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE; A CRACKED BATTERY COMPARTMENT. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION CAN RESULT IN A DELAY IN TREATMENT OR LONG TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE POWER CIRCUIT OR CARTRIDGE COMPARTMENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8503720 ↗MalfunctionMar 27, 2019Structural Problem; Structural Problem
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A CRACKED BATTERY COMPARTMENT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 20-MAR-2019.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8455681 ↗MalfunctionMar 6, 2019Device Displays Incorrect Message; Fail-Safe Problem; Device Displays Incorrect Message; Fail-Safe Problem
ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CALL SERVICE ALARM ISSUE) ISSUE. IT WAS ALLEGED A CALL SERVICE 069 ALARM AND ANOTHER UNKNOWN ALARM OCCURRED. THERE WAS NO INDICATION THAT THIS ISSUE LEAD TO AN ADVERSE EVENT. THIS IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8395955 ↗MalfunctionFeb 12, 2019Structural Problem; Structural Problem
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A KEYPAD BUTTON ISSUE AND A CASE/CONDITION ISSUE. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 06-FEB-2019.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8331044 ↗MalfunctionJan 22, 2019Image Display Error/Artifact; Image Display Error/Artifact
ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8266759 ↗MalfunctionJan 16, 2019No Display/Image; Structural Problem; No Display/Image; Structural Problem
ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DAMAGED) ISSUE; HEAVILY SCRATCHED DISPLAY LENS AND POSSIBLY CRACKED. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8253157 ↗MalfunctionDec 10, 2018Device Displays Incorrect Message; Fail-Safe Problem; Device Displays Incorrect Message; Fail-Safe Problem
ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CALL SERVICE ALARM ISSUE) ISSUE; CALL SERVICE 0069 ALARM. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG-TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8147224 ↗MalfunctionDec 5, 2018Inaccurate Delivery; Inaccurate Delivery
ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS ALLEGING THAT THE PATIENT EXPERIENCED A BLOOD GLUCOSE (BG) OF 36 WITH BLURRED VISION ASSOCIATED WITH AN ALLEGED INACCURATE DELIVERY ISSUE. REPORTEDLY, THE PATIENT REMAINED ON THE PUMP AND DID NOT RECEIVE ANY TREATMENT ABOVE AND BEYOND THE USUAL ROUTINE OF DIABETES CARE AND MANAGEMENT. DURING TROUBLESHOOTING WITH CUSTOMER TECHNICAL SUPPORT, IT WAS REPORTED THAT THE HEALTH CARE PROVIDER (HCP) MADE ADJUSTMENT TO BG. PATIENT HCP HAS INCREASED BG SINCE 11/18/2018 THE REPORTER STATED THAT THE BG WAS INCREASED DUE TO REBOUNDED BG OF 400MG/DL. PATIENT STATED TREATMENT OF LOW BG CONSISTED OF CONSUMING CARBS. PATIENT ALSO STATED IN THE PAST HCP TRIED MAKING ADJUSTMENT TO PUMP AND GOT VERY FRUSTRATED. PATIENT REFUSED TO COMPLETE TROUBLESHOOTING DUE TO BLURRED VISION CAUSED BY LOW BG. PATIENT WAS ADVISED TO DISCONTINUE USE OF PUMP. PATIENT REFUSED TO CONTACT HCP CONCERNING ALTERNATE FORM OF INSULIN DELIVERY. IT WAS REPORTED THAT THE BASAL HISTORY MATCHED THE ACTIVE BASAL PROGRAM SETTINGS. THIS COMPLAINT IS BEING REPORTED BASED ON THE ALLEGATION THAT THE PATIENT EXPERIENCED ISSUES WITH THEIR BLOOD GLUCOSE. THE PUMP COULD NOT BE RULED OUT AS A CONTRI
Patient problems: Blurred Vision; Hypoglycemia
MAUDE report 8132464 ↗MalfunctionOct 28, 2018Structural Problem; Structural Problem
ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE; CRACKED BATTERY COMPARTMENT. THE REPORTER DENIED DAMAGE TO THE BATTERY CAP. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION CAN RESULT IN A DELAY IN TREATMENT OR LONG TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE POWER CIRCUIT OR CARTRIDGE COMPARTMENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8011346 ↗MalfunctionOct 24, 2018Image Display Error/Artifact; Image Display Error/Artifact
ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE DISPLAY SCREEN WAS (DIM/FADING/COLOR SPECTRUM). THERE WAS NO INDICATION THAT THE DEVICE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL OF THE INFORMATION ON THE SCREEN, WHICH MAY RESULT IN OVER OR UNDER DELIVERY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8003358 ↗MalfunctionOct 18, 2018Image Display Error/Artifact; Image Display Error/Artifact
ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7979376 ↗MalfunctionOct 5, 2018Device Displays Incorrect Message; Fail-Safe Problem; Device Displays Incorrect Message; Fail-Safe Problem
ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CALL SERVICE ALARM ISSUE) ISSUE; UNKNOWN CALL SERVICE ALARM REPORTEDLY OCCURRED. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG-TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7938178 ↗MalfunctionOct 4, 2018Device Displays Incorrect Message; Communication or Transmission Problem; Fail-Safe Problem; Device Displays Incorrect Message; Communication or Transmission Problem; Fail-Safe Problem
ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (COMMUNICATION) ISSUE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7934161 ↗MalfunctionSep 12, 2018Electrical /Electronic Property Problem; Structural Problem; Electrical /Electronic Property Problem; Structural Problem
ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (DAMAGE) ISSUE. THE REPORTER STATED THE BATTERY COMPARTMENT WAS CRACKED AND THE BATTERY CAP WOULD NOT REMAIN SECURED TO THE PUMP. THE REPORTER DENIED DAMAGE TO THE BATTERY CAP AND STATED THE YELLOW O-RING WAS VISIBLE WHEN THE CAP WAS ON THE PUMP. THE REPORTER CONFIRMED THE BATTERY CAP HAD NEVER BEEN REPLACED ON THIS PUMP. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE USER MAY BE UNAWARE THAT THE PUMP HAS LOST POWER, LEADING TO UNDER DELIVERY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7868022 ↗MalfunctionSep 7, 2018Image Display Error/Artifact; Image Display Error/Artifact
ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM WITH MOISTURE) ISSUE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL OF THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7855060 ↗MalfunctionAug 2, 2018Premature Discharge of Battery; Structural Problem; Premature Discharge of Battery; Structural Problem
ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE W/ DAMAGE) ISSUE; CRACKED BATTERY COMPARTMENT WITH LESS THAN EXPECTED BATTERY LIFE. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG-TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG-TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM. CONTRIBUTED TO AN ADVERSE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7741823 ↗
Other Pump, infusion, insulin, to be used with invasive glucose sensor devices
- t:slim X2 5.0 Insulin Pump 1000885Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor1 recall2 deaths17,224 reports
- t:slim X2 5.0 Insulin Pump 1000911Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor2 recalls2 deaths15,928 reports
- t:slim X2 5.0 Insulin Pump 1000897Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor1 recall2 deaths6,444 reports
- t:slim X2 5.x Insulin Pump 1003808Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor2 recalls1 death2,526 reports
- t:slim G4 Insulin Pump with Dexcom G4 Platinum CGM 006194Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor4 deaths2,272 reports
- Animas Vibe 11451100Animas Corporation · Pump, infusion, insulin, to be used with invasive glucose sensor2 deaths2,262 reports
- t:slim X2 5.0 Insulin Pump 1000891Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor1 recall1,561 reports
- Animas Vibe 11451000Animas Corporation · Pump, infusion, insulin, to be used with invasive glucose sensor1,122 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page