Animas Vibe 11420500Recalls & reported problems

Animas Corporation · Insulin Pump & CGM System, Adult Green · Class 3

All models of Animas Vibe (45)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Animas Vibe insulin pump, used with a continuous glucose monitor, is a body-worn device that delivers insulin. The 60 reports mostly describe physical damage and reliability problems: cracked casings and battery compartments, moisture getting inside, buttons that no longer respond, dim or blank screens, alarms too quiet to hear, and delays or interruptions in insulin delivery. Two reports involved injuries, and no deaths were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
60
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 60 adverse event reports that list this exact device (DI 10840406100105), received between Aug 12, 2016 and Nov 15, 2017 — classified as 2 injuries, 58 malfunctions.

The most frequently reported specific device problem is “Structural Problem” (19 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Structural Problem19
  • Device Displays Incorrect Message11
  • Fail-Safe Problem11
  • Moisture Damage10
  • Failure to Read Input Signal6
  • No Display/Image6
  • Premature Discharge of Battery5
  • Device Issue3

Patient problems

  • No Consequences Or Impact To Patient58
  • Hyperglycemia2
  • Confusion/ Disorientation1
  • Lethargy1
  • Polydipsia1
  • Urinary Frequency / Polyuria1

Recent adverse event reports

The 25 most recent of 60 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 15, 2017Image Display Error/Artifact; Image Display Error/Artifact

    THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THE DISPLAY WAS DIM. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 10/30/2017.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7036098 ↗
  • MalfunctionOct 30, 2017Device Issue; Device Issue

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A OTHER (UNUSUAL ISSUE) ISSUE. IT WAS NOTED THAT THERE WAS A DELAY IN INSULIN DELIVERY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED ISSUE COULD HAVE AN IMPACT ON INSULIN DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6986631 ↗
  • MalfunctionOct 24, 2017Low Audible Alarm; Low Audible Alarm

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A AUDIO TONE/VIBRATION (QUIET SOUNDS) ISSUE. IT WAS ALLEGED THAT THE VIBRATORY ALARM WAS WORKING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY CAUSE THE USER TO MISS AN ALARM OR WARNING THAT MAY CAUSE CESSATION OF INSULIN DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6975354 ↗
  • MalfunctionOct 16, 2017Failure to Read Input Signal; Failure to Read Input Signal

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER ALLEGED THE OK/ENTER BUTTON WAS UNDER RESPONSIVE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THE COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE HAS THE ABILITY TO RESULT IN INADVERTENT OR INCORRECT INSULIN DELIVERY OR THE INABILITY TO USE THE PUMP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6952269 ↗
  • MalfunctionOct 9, 2017Structural Problem; Structural Problem

    THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A CRACKED BATTERY COMPARTMENT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 9/11/2017.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6933555 ↗
  • MalfunctionSep 18, 2017Premature Discharge of Battery; Premature Discharge of Battery

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE) ISSUE. THERE IS NO REPORTED ADVERSE EVENT WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6873278 ↗
  • MalfunctionAug 25, 2017Structural Problem; Structural Problem

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE. THE CARTRIDGE COMPARTMENT WAS ALLEGED TO BE CRACKED. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION HAS THE ABILITY TO RESULT IN A DELAY IN TREATMENT OR LONG TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE CARTRIDGE COMPARTMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6821962 ↗
  • MalfunctionAug 10, 2017Failure to Read Input Signal; Structural Problem; Failure to Read Input Signal; Structural Problem

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (DAMAGE PRIOR TO TACTILE CHANGE) ISSUE. IT WAS ALSO REPORTED THAT THE KEYPAD WAS MISSING/PEELING AND THAT THE UP ARROW, DOWN ARROW, AND OK/ENTER BUTTONS WERE UNDER-RESPONSIVE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE HAS THE ABILITY TO RESULT IN INADVERTENT OR INCORRECT INSULIN DELIVERY OR THE INABILITY TO USE THE PUMP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6785769 ↗
  • MalfunctionAug 7, 2017Failure to Read Input Signal; Failure to Read Input Signal

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. IT WAS REPORTED THAT THE OK BUTTONS WAS UNDER-RESPONSIVE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE HAS THE ABILITY TO RESULT IN INADVERTENT OR INCORRECT INSULIN DELIVERY OR THE INABILITY TO USE THE PUMP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6769979 ↗
  • MalfunctionAug 1, 2017Device Displays Incorrect Message; Fail-Safe Problem; Device Displays Incorrect Message; Fail-Safe Problem

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CALL SERVICE ALARM ISSUE) ISSUE. THERE WAS NO INDICATION THAT THIS ISSUE LEAD TO AN ADVERSE EVENT. THIS IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6759313 ↗
  • MalfunctionAug 1, 2017Moisture Damage; Structural Problem; Moisture Damage; Structural Problem

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CASE DAMAGE W/MOISTURE) ISSUE. IT WAS REPORTED THAT THE BATTERY COMPARTMENT WAS CRACKED/DAMAGED, WHICH WAS NOTICED AFTER SWIMMING WHILE WEARING THE PUMP. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION HAS THE ABILITY TO RESULT IN A DELAY IN TREATMENT OR LONG TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE POWER CIRCUIT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6758077 ↗
  • MalfunctionAug 1, 2017Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE USER MAY BE UNAWARE THAT THE PUMP HAS LOST POWER, LEADING TO UNDER DELIVERY. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6757207 ↗
  • MalfunctionJul 28, 2017Device Displays Incorrect Message; Fail-Safe Problem; Device Displays Incorrect Message; Fail-Safe Problem

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CALL SERVICE ALARM ISSUE) ISSUE. IT WAS REPORTED THAT THERE WAS A 087 CALL SERVICE ALARM. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6752870 ↗
  • MalfunctionJul 28, 2017Device Displays Incorrect Message; Fail-Safe Problem; Device Displays Incorrect Message; Fail-Safe Problem

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CALL SERVICE ALARM ISSUE) ISSUE. IT WAS REPORTED THAT THERE WAS A 069 CALL SERVICE ALARM. THERE WAS NO INDICATION THAT THIS ISSUE LEAD TO AN ADVERSE EVENT. THIS IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6752869 ↗
  • MalfunctionJul 24, 2017No Display/Image; Structural Problem; No Display/Image; Structural Problem

    THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THE BATTERY COMPARTMENT WAS CRACKED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 06/28/2017.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6739153 ↗
  • MalfunctionJul 24, 2017No Display/Image; Moisture Damage; No Display/Image; Moisture Damage

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (CLOUDY W/ MOISTURE) ISSUE. IT WAS REPORTED THAT MOISTURE WAS LOCATED BEHIND THE DISPLAY. IT WAS ALSO REPORTED THAT THE PATIENT WENT SWIMMING WHILE WEARING THE PUMP ON (B)(6) 2017. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL OF THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6735858 ↗
  • MalfunctionJul 19, 2017Image Display Error/Artifact; Image Display Error/Artifact

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL OF THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6725702 ↗
  • MalfunctionJul 17, 2017Device Displays Incorrect Message; Fail-Safe Problem; Device Displays Incorrect Message; Fail-Safe Problem

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CALL SERVICE ALARM ISSUE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6716710 ↗
  • MalfunctionJul 6, 2017Date/Time-Related Software Problem; Date/Time-Related Software Problem

    THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THE BATTERY COMPARTMENT WAS CRACKED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(6) 2017.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6692035 ↗
  • MalfunctionJul 4, 2017Moisture Damage; Moisture Damage

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (MOISTURE INGRESS) ISSUE. IT WAS ALLEGED THAT THERE WAS MOISTURE IN THE BATTERY COMPARTMENT. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION HAS THE ABILITY TO RESULT IN A DELAY IN TREATMENT OR LONG TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE POWER CIRCUIT OR CARTRIDGE COMPARTMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6685469 ↗
  • MalfunctionJul 3, 2017Device Displays Incorrect Message; Fail-Safe Problem; Device Displays Incorrect Message; Fail-Safe Problem

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CS 078) ISSUE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6685219 ↗
  • MalfunctionJun 14, 2017Premature Discharge of Battery; Premature Discharge of Battery

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE) ISSUE. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE USER MAY BE UNAWARE THAT THE PUMP HAS LOST POWER, LEADING TO UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6642500 ↗
  • MalfunctionJun 14, 2017Electrical /Electronic Property Problem; Moisture Damage; Electrical /Electronic Property Problem; Moisture Damage

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (MOISTURE INGRESS) ISSUE. IT WAS NOTED THAT THERE WAS MOISTURE WITHIN THE BATTERY COMPARTMENT. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE USER MAY BE UNAWARE THAT THE PUMP HAS LOST POWER, LEADING TO UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6642478 ↗
  • MalfunctionJun 12, 2017Premature Discharge of Battery; Moisture Damage; Premature Discharge of Battery; Moisture Damage

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE W/ MOISTURE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6635926 ↗
  • MalfunctionMay 12, 2017Premature Discharge of Battery; Moisture Damage; Premature Discharge of Battery; Moisture Damage

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE W/ MOISTURE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6563020 ↗

Other Pump, infusion, insulin, to be used with invasive glucose sensor devices

All Pump, infusion, insulin, to be used with invasive glucose sensor devices →