Animas Vibe 11520000Recalls & reported problems

Animas Corporation · REP KIT, VIBE, ADULT, BLU,mg,US · Class 3

All models of Animas Vibe (45)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Animas Vibe insulin pump is a device worn on the body to deliver insulin. Many of its 472 reports describe cracked or structural parts, moisture damage, electrical problems, and screen problems such as incorrect messages or faded displays, and a few mention high or low blood sugar in patients. The reports include 22 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
472
Deaths reported
0
Injuries reported
22
Recalls naming it
0

Summary

The FDA's MAUDE database contains 472 adverse event reports that list this exact device (DI 10840406100198), received between Aug 10, 2016 and Jul 26, 2019 — classified as 22 injuries, 450 malfunctions.

The most frequently reported specific device problem is “Structural Problem” (157 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Structural Problem157
  • Moisture Damage85
  • Electrical /Electronic Property Problem67
  • Device Displays Incorrect Message45
  • Fail-Safe Problem42
  • Failure to Read Input Signal38
  • Image Display Error/Artifact38
  • No Display/Image35

Patient problems

  • No Consequences Or Impact To Patient450
  • Hyperglycemia16
  • Hypoglycemia5
  • Nausea4
  • Abdominal Pain2
  • Ambulation or Postural Difficulties2
  • Lethargy2
  • Polydipsia2

Recent adverse event reports

The 25 most recent of 472 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 26, 2019Use of Device Problem; Use of Device Problem

    THE REPORTER CONTACTED ANIMAS ON 25-JUL-2019 ALLEGING THE PATIENT EXPERIENCED HYPERGLYCEMIA WHILE ON INSULIN PUMP THERAPY EVIDENCED BY BLOOD GLUCOSE MEASURING 599/MG/DL WITH LARGE KETONES, ABDOMINAL PAIN AND NAUSEA. THE PATIENT WAS REPORTEDLY HOSPITALIZED AND RECEIVED TREATMENT OF INTRAVENOUS FLUIDS AND INSULIN VIA THE PUMP. DURING TROUBLESHOOTING WITH ANIMAS CUSTOMER SUPPORT, IT WAS REVEALED THAT THE PATIENT¿S BLOOD GLUCOSE ELEVATED TO 599 MG/DL AFTER INFUSING 18 UNITS OF INSULIN THAT HAD BEEN STORED INSIDE OF A HOT CAR VIA THE PUMP. THE REPORTER STATED THAT THE HEATED INSULIN THAT FILLED THE CARTRIDGE WAS CLOUDY IN APPEARANCE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT EXPERIENCED SERIOUS INJURY WHILE ON INSULIN PUMP THERAPY ASSOCIATED WITH TRAINING/MISUSE; NOT STORING INSULIN PER MANUFACTURER INSTRUCTION AND USING SAID INSULIN IN THE INSULIN PUMP FOR DIABETES MANAGEMENT. NO PRODUCT IS BEING RETURNED ASSOCIATED WITH THIS COMPLAINT; NO COMPLAINT AGAINST THE PERFORMANCE OF THE INSULIN PUMP.

    Patient problems: Abdominal Pain; Hyperglycemia; Nausea

    MAUDE report 8831747 ↗
  • MalfunctionApr 26, 2019Failure to Read Input Signal; Failure to Read Input Signal

    ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS ALLEGING THE OK BUTTON ON THE KEYPAD WAS UNDER RESPONSIVE. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE CAN RESULT IN INADVERTENT OR INCORRECT INSULIN DELIVERY OR THE INABILITY TO USE THE PUMP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8557521 ↗
  • MalfunctionFeb 26, 2019Image Display Error/Artifact; Image Display Error/Artifact

    ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE THAT WAS NOT IMPROVED WITH CONTRAST ADJUSTMENT DURING TROUBLESHOOTING WITH ANIMAS CUSTOMER SUPPORT. THE REPORTER STATED THE DISPLAY SCREEN COULD BE READ SAFELY AND DECLINED TO RETURN THE PUMP AND RECEIVE A REPLACEMENT PUMP. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8369987 ↗
  • MalfunctionFeb 26, 2019Electrical /Electronic Property Problem; Structural Problem; Electrical /Electronic Property Problem; Structural Problem

    ON (B)(6) 2019 THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (DAMAGE) ISSUE. THE REPORTER STATED THAT THE PUMP WOULD NOT POWER ON UNLESS THEY TAPED THE BATTERY CAP ON TO THE PUMP. THE BATTERY COMPARTMENT IS REPORTEDLY CRACKED. THE REPORTER DENIED DAMAGE TO THE BATTERY CAP AND STATED THAT THE CAP DOES NOT SECURE PROPERLY TO THE PUMP. THE REPORTER CONFIRMED THE BATTERY CAP HAD NOT BEEN REPLACED ON THIS PUMP FOR SEVEN-TO-TWELVE MONTHS. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE USER MAY BE UNAWARE THAT THE PUMP HAS LOST POWER, LEADING TO UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8369946 ↗
  • MalfunctionFeb 7, 2019Structural Problem; Structural Problem

    THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THE BATTERY COMPARTMENT WAS CRACKED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 10-OCT-2018.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8318073 ↗
  • MalfunctionFeb 6, 2019Structural Problem; Structural Problem

    THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THE BATTERY COMPARTMENT WAS CRACKED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 08-JAN-2019.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8311870 ↗
  • MalfunctionJan 28, 2019Image Display Error/Artifact; Image Display Error/Artifact

    ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8281718 ↗
  • MalfunctionJan 28, 2019Structural Problem; Structural Problem

    ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE; A CRACKED BATTERY COMPARTMENT. THE REPORTER DENIED DAMAGE TO THE BATTERY CAP. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION CAN RESULT IN A DELAY IN TREATMENT OR LONG TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE POWER CIRCUIT OR CARTRIDGE COMPARTMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8281715 ↗
  • MalfunctionDec 17, 2018Structural Problem; Structural Problem

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE; CRACKED BATTERY COMPARTMENT. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION CAN RESULT IN A DELAY IN TREATMENT OR LONG TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE POWER CIRCUIT OR CARTRIDGE COMPARTMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8170446 ↗
  • MalfunctionOct 28, 2018No Display/Image; Moisture Damage; No Display/Image; Moisture Damage

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (FLASHING DISPLAY W/MOISTURE) ISSUE WITH MOISTURE INSIDE THE BATTERY COMPARTMENT. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8011380 ↗
  • MalfunctionOct 28, 2018Moisture Damage; Structural Problem; Moisture Damage; Structural Problem

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CASE DAMAGE W/MOISTURE) ISSUE; CRACKED BATTERY COMPARTMENT WITH MOISTURE IN THE BATTERY COMPARTMENT. THE REPORTER DENIED DAMAGE TO THE BATTERY CAP. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION CAN RESULT IN A DELAY IN TREATMENT OR LONG TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE POWER CIRCUIT OR CARTRIDGE COMPARTMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8011218 ↗
  • MalfunctionSep 20, 2018Pumping Problem; Pumping Problem

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A HISTORY/SETTINGS (HISTORY/SETTINGS ISSUE) ISSUE. THE PATIENT REPORTED NO ALARMS WERE STORED IN THE ALARM HISTORY. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS ISSUE IS BEING REPORTED AS AN ISSUE WITH THE PUMP NOT PERFORMING AS INTENDED WITH REGARDS TO THE HISTORY OR THE SETTINGS MAY RESULT IN OVER OR UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7894810 ↗
  • MalfunctionSep 11, 2018Structural Problem; Structural Problem

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE; CARTRIDGE COMPARTMENT CRACK. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION CAN RESULT IN A DELAY IN TREATMENT OR LONG TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE POWER CIRCUIT OR CARTRIDGE COMPARTMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7865291 ↗
  • MalfunctionSep 7, 2018Moisture Damage; Structural Problem; Moisture Damage; Structural Problem

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CASE DAMAGE WITH MOISTURE) ISSUE. IT WAS ALLEGED THAT THE BATTERY COMPARTMENT WAS DAMAGED WITH MOISTURE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION HAS THE ABILITY TO RESULT IN A DELAY IN TREATMENT OR LONG TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE POWER CIRCUIT OR CARTRIDGE COMPARTMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7854820 ↗
  • MalfunctionSep 6, 2018Electrical /Electronic Property Problem; Structural Problem; Electrical /Electronic Property Problem; Structural Problem

    ON (B)(6) 2018, A REPORTER CONTACTED ANIMAS ALLEGING THAT A PATIENT HAD EXPERIENCED A HYPERGLYCEMIC EVENT WHILE ON THE PUMP. THE REPORTER STATED THAT THE PATIENT HAD A BLOOD GLUCOSE OF OVER 500 MG/DL WITH SYMPTOMS OF DROWSINESS AND EXCESS URINATION. THE PATIENT DID NOT RECEIVE ANY TREATMENT ABOVE AND BEYOND THE NORMAL COURSE OF DIABETES MANAGEMENT. THE REPORTER STATED THAT THE PATIENT EXPERIENCED THE BLOOD GLUCOSE EXCURSION AFTER THE PUMP STOPPED WORKING. THE REPORTER ALLEGED THAT THIS WAS RELATED TO A POWER ISSUE WITH THE PUMP. THE BATTERY CAP WAS FOUND TO BE DAMAGED. THIS COMPLAINT IS BEING REPORTED BASED ON THE ALLEGATION THAT THE PATIENT EXPERIENCED A HYPERGLYCEMIC EVENT WHILE ON THE PUMP DUE TO A POWER ISSUE.

    Patient problems: Hyperglycemia; Lethargy; Urinary Frequency / Polyuria

    MAUDE report 7850149 ↗
  • MalfunctionAug 30, 2018Structural Problem; Structural Problem

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7834383 ↗
  • MalfunctionAug 28, 2018No Display/Image; No Display/Image

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (CLOUDY) ISSUE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL OF THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7822298 ↗
  • MalfunctionAug 14, 2018Structural Problem; Structural Problem

    THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A CASE/CONDITION ISSUE. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 09-AUG-2018.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7782035 ↗
  • MalfunctionAug 13, 2018No Display/Image; Structural Problem; No Display/Image; Structural Problem

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DAMAGED) ISSUE. IT WAS REPORTED THAT THE DISPLAY SCREEN WAS SCRATCHED AND CRACKED AND UNABLE TO BE READ SAFELY. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL OF THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7779777 ↗
  • MalfunctionAug 10, 2018Failure to Read Input Signal; Failure to Read Input Signal

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. IT WAS REPORTED THAT THE CONTRAST, UP ARROW, DOWN ARROW, AND OK/ENTER BUTTONS WERE UNDER-RESPONSIVE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE HAS THE ABILITY TO RESULT IN INADVERTENT OR INCORRECT INSULIN DELIVERY OR THE INABILITY TO USE THE PUMP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7774620 ↗
  • MalfunctionAug 7, 2018Image Display Error/Artifact; Image Display Error/Artifact

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (LINE THROUGH DISPLAY) ISSUE. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7758974 ↗
  • MalfunctionAug 2, 2018Structural Problem; Structural Problem

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE. THE REPORTER ALLEGED THAT THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION HAS THE ABILITY TO RESULT IN A DELAY IN TREATMENT OR LONG TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE POWER CIRCUIT OR THE CARTRIDGE COMPARTMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7744244 ↗
  • InjuryJul 31, 2018Electrical /Electronic Property Problem; Moisture Damage; Electrical /Electronic Property Problem; Moisture Damage

    ON (B)(6) 2018, A REPORTER CONTACTED ANIMAS ALLEGING THAT A PATIENT HAD EXPERIENCED A HYPERGLYCEMIC EVENT WHILE ON THE PUMP. THE REPORTER STATED THAT THE PATIENT HAD A BLOOD GLUCOSE OF 500 MG/DL WITH SYMPTOMS OF EXTREME DROWSINESS AND NAUSEA. THE PATIENT DID NOT RECEIVE ANY TREATMENT ABOVE AND BEYOND THE NORMAL COURSE OF DIABETES MANAGEMENT. THE REPORTER STATED THAT THE PUMP HAD LOST POWER. UPON INSPECTION OF THE PUMP, MOISTURE CORROSION WAS FOUND IN THE BATTERY COMPARTMENT. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGATION THAT THE PUMP LOSING POWER UNEXPECTEDLY CAUSED THE BLOOD GLUCOSE EXCURSION.

    Patient problems: Hyperglycemia; Lethargy; Nausea

    MAUDE report 7735340 ↗
  • MalfunctionJul 27, 2018No Display/Image; No Display/Image

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (MISALIGNED) ISSUE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL OF THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7728226 ↗
  • MalfunctionJul 27, 2018Premature Discharge of Battery; Structural Problem; Premature Discharge of Battery; Structural Problem

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE W/ DAMAGE) ISSUE. IT WAS NOTED THAT THE BATTERY HAS TO BE REPLACED EVERY DAY AND THE PUMP IS CRACKED. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE USER MAY BE UNAWARE THAT THE PUMP HAS LOST POWER, LEADING TO UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7725663 ↗

Other Pump, infusion, insulin, to be used with invasive glucose sensor devices

All Pump, infusion, insulin, to be used with invasive glucose sensor devices →