Skip to content

Animas Vibe 11620000Recalls & reported problems

Animas Corporation · REP KIT, VIBE PEDS, P2 (OLDER), BLU, mg, US · Class 3

All models of Animas Vibe (45)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Animas Vibe insulin pump delivers insulin and is used with a glucose sensor. Reports mostly describe moisture getting into the pump's casing, cracked battery compartments, electronic and display problems, and alarms that keep appearing. There are 46 reports, with 2 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
46
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 46 adverse event reports that list this exact device (DI 10840406100280), received between Aug 9, 2016 and May 24, 2019 — classified as 2 injuries, 44 malfunctions.

The most frequently reported specific device problem is “Moisture Damage” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Moisture Damage13
  • Structural Problem12
  • Electrical /Electronic Property Problem7
  • No Display/Image6
  • Device Displays Incorrect Message5
  • Fail-Safe Problem4
  • Failure to Prime4
  • Failure to Read Input Signal3

Patient problems

  • No Consequences Or Impact To Patient43
  • Hyperglycemia2
  • Dehydration1
  • Dizziness1
  • Elevated Ketones/ Diabetic Ketoacidosis1
  • Hypoglycemia1
  • No Code Available1

Recent adverse event reports

The 25 most recent of 46 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 24, 2019Moisture Damage; Moisture Damage

    ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (MOISTURE INGRESS) ISSUE' MOISTURE BEHIND THE DISPLAY AND IN THE INFRA-RED WINDOW. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION CAN RESULT IN A DELAY IN TREATMENT OR LONG TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE POWER CIRCUIT OR CARTRIDGE COMPARTMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8639805 ↗
  • MalfunctionJan 23, 2019Pumping Problem; Pumping Problem

    ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS, ALLEGING A OTHER (UNUSUAL ISSUE) ISSUE. THE REPORTER ALLEGED THE PUMP WILL EMIT A LOW BATTERY WARNING AND THE USER WILL CHANGE OUT THE BATTERY AND THE LOW BATTERY WARNING WILL RE-APPEAR AGAIN. DURING TROUBLESHOOTING WITH ANIMAS CUSTOMER SUPPORT, THE ALARM HISTORY IN THE PUMP WAS REVIEWED AND THERE WERE NO LOW BATTERY WARNINGS IN THE HISTORY AS THE REPORTER ALLEGED. THE REPORTER THEN ALLEGED THAT ALL LOW BATTERY WARNINGS THAT WERE EMITTED FROM THE PUMP FROM (B)(6) 2019, TO THE CURRENT TIME WERE NOT RECORDED IN THE PUMP'S ALARM HISTORY AFTER THE USER CONFIRMED THE WARNING. DURING TROUBLESHOOTING, THE PUMP EMITTED A LOW BATTERY WARNING AND IT WAS CORRECTLY RECORDED IN THE PUMP'S ALARM HISTORY AS EXPECTED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG-TERM CESSATION OF INSULIN DELIVERY DUE TO UNEXPECTED POWER LOSS IF THE USER IS UNABLE TO RESOLVE THE ALLEGED LOW BATTERY WARNING RECORDING ERROR ISSUE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8271077 ↗
  • MalfunctionNov 14, 2018Structural Problem; Structural Problem

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE; CRACKED BATTERY COMPARTMENT. THE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION CAN RESULT IN A DELAY IN TREATMENT OR LONG TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE POWER CIRCUIT OR CARTRIDGE COMPARTMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8068753 ↗
  • MalfunctionSep 21, 2018Device Displays Incorrect Message; Fail-Safe Problem; Device Displays Incorrect Message; Fail-Safe Problem

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS AND REPORTED THAT CALL SERVICE ALARM [064] HAD BEEN EMITTED BY THE PUMP MORE OFTEN THAN EXPECTED BY THE USER. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7898157 ↗
  • MalfunctionAug 28, 2018Partial Blockage; Device Displays Incorrect Message; Partial Blockage; Device Displays Incorrect Message

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS ALLEGING THAT THE PATIENT EXPERIENCED A BLOOD GLUCOSE OF 350 MG/DL WITH MODERATE KETONES AND NAUSEA ASSOCIATED WITH AN ALLEGED OCCLUSION ISSUE. REPORTEDLY, THE PATIENT DISCONTINUED PUMP THERAPY AND DID NOT RECEIVE ANY TREATMENT ABOVE AND BEYOND THE USUAL ROUTINE OF DIABETES CARE AND MANAGEMENT BUT WAS GIVEN MULTIPLE MANUAL INJECTIONS TO BRING THE BLOOD GLUCOSE BACK DOWN TO THE TARGET RANGE. DURING TROUBLESHOOTING WITH CUSTOMER TECHNICAL SUPPORT, IT WAS NOTED THERE WAS AN OCCLUSION AND A LOSS OF PRIME. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT EXPERIENCED HYPERGLYCEMIA ASSOCIATED WITH AN ALLEGED OCCLUSION ISSUE.

    Patient problems: Hyperglycemia; No Code Available

    MAUDE report 7822818 ↗
  • MalfunctionJul 20, 2018Electrical /Electronic Property Problem; Structural Problem; Electrical /Electronic Property Problem; Structural Problem

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE PUMP HAD NO POWER AND THE BATTERY COMPARTMENT WAS DAMAGED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE USER MAY BE UNAWARE THAT THE PUMP HAS LOST POWER, LEADING TO UNDER DELIVERY. THERE WAS NO SERIOUS INJURY ASSOCIATED WITH THIS COMPLAINT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7706578 ↗
  • MalfunctionJul 10, 2018Structural Problem; Structural Problem

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE. THE BATTERY COMPARTMENT WAS ALLEGED TO BE CRACKED/DAMAGED. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION HAS THE ABILITY TO RESULT IN A DELAY IN TREATMENT OR LONG TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE POWER CIRCUIT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7673491 ↗
  • MalfunctionJun 19, 2018Device Displays Incorrect Message; Fail-Safe Problem; Device Displays Incorrect Message; Fail-Safe Problem

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CS 078) ISSUE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7613701 ↗
  • MalfunctionMay 22, 2018Date/Time-Related Software Problem; Date/Time-Related Software Problem

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE TIME AND DATE RESET TO DEFAULT RANDOMLY. THERE WAS NO INDICATION THAT THE DEVICE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE OF A TIME CHANGE DURING USE, WITHOUT USER INTERVENTION, MAY RESULT IN THE USER RUNNING THE INCORRECT BASAL SEGMENT LEADING TO OVER OR UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7533155 ↗
  • MalfunctionMar 15, 2018Electrical /Electronic Property Problem; Structural Problem; Electrical /Electronic Property Problem; Structural Problem

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (DAMAGE) ISSUE. THE REPORTER ALLEGED THE PUMP WAS UNABLE TO POWER ON, AND THE BATTERY COMPARTMENT WAS CRACKED WITH THE BATTERY COMPARTMENT CAP UNABLE TO TIGHTEN TO THE PUMP. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE USER MAY BE UNAWARE THAT THE PUMP HAS LOST POWER, LEADING TO UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7343141 ↗
  • MalfunctionFeb 23, 2018Moisture Damage; Failure to Read Input Signal; Moisture Damage; Failure to Read Input Signal

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING THE AUDIO BOLUS BUTTON WAS MISSING WITH MOISTURE PRESENT BEHIND THE DISPLAY AND NEAR THE AUDIO BOLUS BUTTON. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THE COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE HAS THE ABILITY TO RESULT IN INADVERTENT OR INCORRECT BOLUSES WHICH MAY RESULT IN OVER OR UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7292555 ↗
  • MalfunctionFeb 6, 2018No Display/Image; No Display/Image

    ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (CLOUDY) ISSUE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL OF THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7244946 ↗
  • MalfunctionJan 12, 2018Failure to Read Input Signal; Failure to Read Input Signal

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THERE IS NO INDICATION THAT THE PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THE COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE HAS THE ABILITY TO RESULT IN INADVERTENT OR INCORRECT INSULIN DELIVERY OR THE INABILITY TO USE THE PUMP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7186131 ↗
  • MalfunctionJan 8, 2018Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (NO POWER) ISSUE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE USER MAY BE UNAWARE THAT THE PUMP HAS LOST POWER, LEADING TO UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7173414 ↗
  • MalfunctionJan 5, 2018Failure to Prime; Failure to Prime

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. THERE WAS NO INDICATION THAT THE ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG-TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7168590 ↗
  • MalfunctionDec 15, 2017Moisture Damage; Moisture Damage

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (MOISTURE INGRESS) ISSUE. IT WAS REPORTED THAT THERE WAS MOISTURE IN THE BATTERY COMPARTMENT WITHOUT DAMAGE. THERE IS NO INDICATION THAT THE ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION HAS THE ABILITY TO RESULT IN A DELAY IN TREATMENT OR LONG-TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE POWER CIRCUIT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7122768 ↗
  • MalfunctionOct 23, 2017No Display/Image; No Display/Image

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE DISPLAY SCREEN WAS BLANK. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN THE INABILITY TO USE THE PRODUCT WHICH MAY LEAD TO LONG TERM CESSATION OF DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6969930 ↗
  • MalfunctionOct 13, 2017No Display/Image; No Display/Image

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (CLOUDY) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL OF THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6948905 ↗
  • MalfunctionOct 11, 2017Failure to Prime; Failure to Prime

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. IT WAS REPORTED THAT THE PUMP REPEATEDLY EMITTED LOSS OF PRIME WARNINGS DESPITE CARTRIDGE CHANGES. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS ISSUE IS BEING REPORTED BECAUSE IT MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6943271 ↗
  • MalfunctionOct 2, 2017Moisture Damage; Moisture Damage

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (MOISTURE INGRESS) ISSUE. THE REPORTER ALLEGED THERE WAS MOISTURE PRESENT IN THE BATTERY COMPARTMENT. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION HAS THE ABILITY TO RESULT IN A DELAY IN TREATMENT OR LONG TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE POWER CIRCUIT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6907131 ↗
  • MalfunctionSep 11, 2017Premature Discharge of Battery; Moisture Damage; Premature Discharge of Battery; Moisture Damage

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE W/ MOISTURE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6857162 ↗
  • MalfunctionSep 5, 2017Device Displays Incorrect Message; Fail-Safe Problem; Device Displays Incorrect Message; Fail-Safe Problem

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CS 064) ISSUE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6844107 ↗
  • MalfunctionSep 5, 2017Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. THE REPORTER WAS NOT ABLE TO PROVIDE FURTHER INFORMATION DURING THE INITIAL COMPLAINT. CUSTOMER TECHNICAL SUPPORT MADE ATTEMPTS TO CONTACT THE REPORTER FOR ADDITIONAL INFORMATION AND TROUBLESHOOTING, WITHOUT SUCCESS. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS ISSUE IS BEING REPORTED BECAUSE THE USER MAY BE UNAWARE THAT THE PUMP HAS LOST POWER, LEADING TO UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6844106 ↗
  • MalfunctionAug 23, 2017Premature Discharge of Battery; Moisture Damage; Premature Discharge of Battery; Moisture Damage

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE WITH MOISTURE) ISSUE. IT WAS ALLEGED THAT THERE WAS MOISTURE BEHIND THE DISPLAY. THE REPORTER WAS NOT ABLE TO PROVIDE FURTHER INFORMATION DURING THE INITIAL COMPLAINT. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THE ISSUE IS BEING REPORTED BECAUSE IT MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6816011 ↗
  • MalfunctionAug 23, 2017Electrical /Electronic Property Problem; Moisture Damage; Electrical /Electronic Property Problem; Moisture Damage

    ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (MOISTURE INGRESS) ISSUE. IT WAS ALLEGED THAT THERE WAS MOISTURE BEHIND THE DISPLAY AND THERE WAS INTERMITTENT POWER. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS ISSUE IS BEING REPORTED BECAUSE THE USER MAY BE UNAWARE THAT THE PUMP HAS LOST POWER, LEADING TO UNDER DELIVERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6816001 ↗

Other Pump, infusion, insulin, to be used with invasive glucose sensor devices

All Pump, infusion, insulin, to be used with invasive glucose sensor devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page