The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10841156102043), received between Nov 13, 2020 and Aug 27, 2025 — classified as 4 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Gas/Air Leak” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Gas/Air Leak2
- Patient Device Interaction Problem2
- Adverse Event Without Identified Device or Use Problem1
- Device Slipped1
- Inflation Problem1
- Material Twisted/Bent1
Patient problems
- Hematoma5
- Hemorrhage/Blood Loss/Bleeding4
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 27, 2025Device Slipped; Device Slipped
VASC BAND HEMOSTAT OPENED TO STERILE FIELD AND TESTED BY INFLATING WITH 20ML AIR. PLASTIC PIECE FELL OFF OF VASC BAND.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22901644 ↗MalfunctionMar 18, 2025
VASC BAND HEMOSTAT OPENED TO STERILE FIELD AND TESTED BY INFLATING WITH 20ML AIR. PLASTIC PIECE FELL OFF OF VASC BAND.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21630788 ↗InjuryNov 15, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT: THE VASCBAND AIR SELF-DEFLATING. HEMATOMA FORMED COMPRESSION HELD.
Patient problems: Hematoma
MAUDE report 18145977 ↗InjurySep 26, 2023Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT: CUSTOMER STATES "VAC BAND LEAKING AIR. ISSUES WITH REBLEEDING, HEMATOMA AND VASCULAR SURGERY." ADDITIONAL INFORMATION: IT WAS CONFORMED THAT THE ISSUE WAS WITH 1 PATIENT AND 1 VASCBAND. THE PATIENT EXPERIENCED A HEMATOMA/BLEED IN THE RECOVERY UNIT POST PROCEDURE. VASCULAR SURGEON WAS CONSULTED BUT IT IS UNKNOWN IF THE PATIENT RETURNED TO SURGERY. ADDITIONAL INFORMATION HAS BEEN REQUESTED FROM THE ACCOUNT.
Patient problems: Hematoma; Hemorrhage/Blood Loss/Bleeding
MAUDE report 17816882 ↗InjuryAug 15, 2023Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT: THIS COMPLAINT IS FOR THE FIRST VASCBAND THAT WAS APPLIED. ASSOCIATED TO. LHC PCI WAS PERFORMED ACCESS VIA THE RIGHT RADIAL ARTERY ON AN 80 Y/O FEMALE WITH FRAGILE SKIN INTEGRITY. A 6F SHEATH WAS USED IN THE PROCEDURE. PCI PROCEDURE DID NOT HAVE ANY KNOWN ISSUES/COMPLICATIONS. IT WAS REPORTED THAT THERE WAS A SKIN TEAR BESIDE THE ACCESS AREA PREVIOUS TO THE PCI PROCEDURE. POST PCI PROCEDURE, SHEATH PULLED BY MD. VASCBAND PLACED. THE PATIENT WAS TRANSFERRED FROM TABLE TO STRETCHER. IT WAS THEN REPORTED THAT THERE WAS A LARGE HEMATOMA PRESENT PROXIMAL TO THE VASCBAND. A SECOND VASCBAND WAS PLACED ABOVE THE FIRST VASCBAND. HEMATOMA CONTINUED TO GROW. BOTH BANDS WERE REMOVED, AND MANUAL PRESSURE WAS APPLIED. IT WAS REPORTED THAT THE PATIENT'S SKIN TORE AND CAUSED MODERATE BLEEDING. NO BLOOD TRANSFUSION WAS REQUIRED. VASCULAR SURGERY WAS CONSULTED AND DID A PRESSURE DRESSING. WOUND CONSULT WAS SCHEDULED FOR THE PATIENT. ADDITIONAL INFORMATION. THE CASE WAS A PCI. THE HEMATOMA WAS THE SIZE OF A TENNIS BALL. THERE WAS NO VISIBLE AIR LEAK. THE DEVICE WILL NOT BE RETURNED.
Patient problems: Hematoma; Hemorrhage/Blood Loss/Bleeding
MAUDE report 17547696 ↗InjuryAug 14, 2023Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT: THIS COMPLAINT IS FOR THE SECOND VASCBAND THAT WAS APPLIED. ASSOCIATED TO 2134812-2023-00037. LHC PCI WAS PERFORMED ACCESS VIA THE RIGHT RADIAL ARTERY ON AN 80 Y/O FEMALE WITH FRAGILE SKIN INTEGRITY. A 6F SHEATH WAS USED IN THE PROCEDURE. PCI PROCEDURE DID NOT HAVE ANY KNOWN ISSUES/COMPLICATIONS. IT WAS REPORTED THAT THERE WAS A SKIN TEAR BESIDE THE ACCESS AREA PREVIOUS TO THE PCI PROCEDURE. POST PCI PROCEDURE, SHEATH PULLED BY MD. VASCBAND PLACED. THE PATIENT WAS TRANSFERRED FROM TABLE TO STRETCHER. IT WAS THEN REPORTED THAT THERE WAS A LARGE HEMATOMA PRESENT PROXIMAL TO THE VASCBAND (RELATED TO (B)(4)). A SECOND VASCBAND WAS PLACED ABOVE THE FIRST VASCBAND. (RELATED TO (B)(4)) HEMATOMA CONTINUED TO GROW. BOTH BANDS WERE REMOVED, AND MANUAL PRESSURE WAS APPLIED. IT WAS REPORTED THAT THE PATIENT'S SKIN TORE AND CAUSED MODERATE BLEEDING. NO BLOOD TRANSFUSION WAS REQUIRED. VASCULAR SURGERY WAS CONSULTED AND DID A PRESSURE DRESSING. WOUND CONSULT WAS SCHEDULED FOR THE PATIENT. ADDITIONAL INFORMATION. THE CASE WAS A PCI. THE HEMATOMA WAS THE SIZE OF A TENNIS BALL. THERE WAS NO VISIBLE AIR LEAK. THE DEVICE WILL NOT BE RETURNED.
Patient problems: Hematoma; Hemorrhage/Blood Loss/Bleeding
MAUDE report 17541133 ↗MalfunctionAug 4, 2023Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT: THE VASCBAND AIR SELF-DEFLATING. HEMATOMA WAS FORMED AND COMPRESSION WAS HELD. THE CURRENT PATIENT CONDITION IS REPORTED AS "FINE". ADDITIONAL INFORMATION WAS REQUESTED AND NO INFORMATION WAS RECEIVED REGARDING MEDICAL INTERVENTION.
Patient problems: Hematoma
MAUDE report 17458148 ↗MalfunctionNov 13, 2020Inflation Problem; Material Twisted/Bent
ELDERLY MALE WITH HISTORY OF HYPERTENSION AND NEW ONSET OF CHEST PRESSURE. HAVING A HEART CATH, THE VASC BAND WAS USED IN THE RIGHT, RADIAL ARTERY. BALLOON WOULD NOT GET FIRM AND THE PATIENT KEPT BLEEDING. WHEN THE JUNCTURE OF THE TUBING WAS MOVED, IT WAS DISCOVERED THAT THE INFLATABLE CUFF WAS KINKED AT THE CONNECTION.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 10835630 ↗
Other CLAMP, VASCULAR devices
- Heartstring III System 3.8MM HSK-3038Maquet Cardiovascular LLC · Clamp, Vascular1 recall4 deaths809 reports
- Heartstring III System 4.3MM HSK-3043Maquet Cardiovascular LLC · Clamp, Vascular1 recall265 reports
- TR BAND TRB24-REGTerumo Medical Corporation · CLAMP, VASCULAR284 reports
- TR BAND TRB24-REGTerumo Medical Corporation · CLAMP, VASCULAR256 reports
- HeartString III Proximal Seal HS-3045Maquet Cardiovascular LLC · Clamp, Vascular1 recall187 reports
- TR BAND TRB29-LRGTerumo Medical Corporation · CLAMP, VASCULAR67 reports
- TR BAND TRB29-LRGTerumo Medical Corporation · CLAMP, VASCULAR65 reports
- Flo-Thru Intraluminal Shunt® FT-12225Synovis Life Technologies, Inc. · CLAMP, VASCULAR1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page