SPECTRUM C6381Recalls & reported problems

Conmed Corporation · SPECTRUM II SUTURE HOOK, DISPOSABLE, 45° LEFT · Class 1

All models of SPECTRUM (12)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10845854019863), received between Feb 27, 2019 and Dec 14, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation2
  • Break1
  • Dull, Blunt1

Patient problems

  • No Consequences Or Impact To Patient2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 14, 2023Material Fragmentation; Material Fragmentation

    THE CUSTOMER REPORTED THAT THE C6381, SPECTRUM II SUTURE HOOK, DISPOSABLE, 45 DEGREE LEFT, DEVICE WAS BEING USED DURING A SHOULDER ARTHROSCOPY PROCEDURE ON 23OCT23 WHEN IT WAS REPORTED ¿DURING SHOULDER ARTHROSCOPIC SURGERY, THE SPECTRUM HOOK BROKE INTO TWO PARTS AT THE TUBE/HOOK JUNCTION. IT FELL INTO THE PATIENT'S SHOULDER. THE SURGEON REMOVED THE PIECE WITH IMPORTANT DIFFICULTY GIVEN THE SMALL SIZE OF THE DEVICE.¿. THE PROCEDURE WAS COMPLETED AND THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION OR EXTENDED HOSPITALIZATION FOR THE PATIENT. THIS REPORT IS BEING RAISED ON THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18324213 ↗
  • MalfunctionJan 29, 2020Material Fragmentation; Material Fragmentation

    THIS DISTRIBUTOR REPORTED ON BEHALF OF THEIR CUSTOMER THAT THE C6381 DEVICE WAS BEING USED DURING A BANKART PROCEDURE ON (B)(6) 2020 WHEN THE DEVICE TIP BROKE IN THE PATIENT. IT WAS REPORTED THAT ALL PIECES WERE REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE DEVICE AFTER A 2-MINUTE DELAY. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9642224 ↗
  • MalfunctionFeb 27, 2019Break; Dull, Blunt; Break; Dull, Blunt

    THE CUSTOMER REPORTED AN ISSUE WITH THE SPECTRUM II SUTURE HOOK, DISPOSABLE, 45 DEGREE LEFT, ITEM # C6381, LOT 966582, THAT OCCURRED ON (B)(6) 2019 DURING USE IN A SHOULDER STABILIZATION PROCEDURE. IT WAS REPORTED THAT THE C6381 SUTURE HOOK BROKE OFF IN A PATIENT'S SHOULDER. THE HOOK WAS RETRIEVED. IT IS INDICATED THAT THERE WAS NO IMPACT OR INJURY TO THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A REPORTED 20-MINUTE DELAY. ADDITIONAL INFORMATION WAS RECEIVED FROM THE REPORTER AND THE HOSPITAL ON (B)(6) 2019. THE ARTHROSCOPIC BANKART PROCEDURE WAS APPROXIMATELY MIDWAY THROUGH, HAVING MOBILIZED THE TISSUE AND PREPARED THE BONE SURFACE. THE DOCTOR WAS HANDED THE LIMITED USE SUTURE HOOK FOR THE FIRST ATTEMPT AT PASSING THE SUTURE THROUGH TISSUE. THIS HOOK WAS BLUNT, THEREFORE SHE WAS OFFERED THE SINGLE USE SUTURE HOOK, WHICH SHE ACCEPTED AND ATTEMPTED TO USE TO PASS THROUGH THE SAME TISSUE AS THE FIRST ATTEMPT. THIS HOOK WAS REMOVED FROM THE PACKAGING JUST PRIOR TO THIS AND WAS PREVIOUSLY UNUSED. THE DOCTOR FOUND THAT THIS WAS ALSO BLUNT, AND THAT IS WHEN THE SUTURE HOOK BROKE OFF. IT TOOK APPROXIMATELY 20-MINS TO RETRIEVE THE BROKEN PART WHICH WAS RETRIEVED WHOLE. NO FRA

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8375605 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →