Skip to content

Spectrum MVP C6400Recalls & reported problems

Conmed Corporation · Spectrum MVP Suture Passer, Disposable, Crescent, Medium · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10845854019894), received between Dec 16, 2024 and Dec 16, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Material Fragmentation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 16, 2024Material Fragmentation; Material Fragmentation

    A SALES REPRESENTATIVE REPORTED ON BEHALF OF A CUSTOMER THAT THE C6400, SPECTRUM MVP SUTURE PASSER, DISPOSABLE, CRESCENT, MEDIUM DEVICE WAS USED ON 21NOV24, AND ¿DURING A LABRAL REPAIR, A VERY TINY PIECE OF THE "FORK" THAT PASSES THE SUTURE THROUGH THE MVP BROKE OFF WHICH WE ASSUME BROKE INSIDE THE PATIENT. SURGEON DID NOT NOTICE IT BROKE UNTIL WE WENT TO RELOAD ANOTHER SUTURE. THE SMALL PIECE WAS NOT RECOVERED BUT TO OUR KNOWLEDGE NO HARM HAS BEEN DONE TO THE PATIENT. THE PIECE WAS MUCH TOO SMALL FOR FLUOROSCOPY SO WE DID NOT ATTEMPT TO FIND IT IN THE PATIENT.¿. FOLLOW-UP ASSESSMENT FOUND THAT "NO X-RAY OR ULTRASOUND WAS ATTEMPTED, SURGEON FELT THAT THE PIECE WAS WAY TO SMALL TO FIND AND FELT THE SCOPE WOULD SUCK IT UP WITH SUCTION. THE PIECE WAS EXTREMELY EXTREMELY SMALL, THE SURGEON BELIEVES IT BROKE OFF INSIDE THE PATIENT BUT SINCE IT WAS SO SMALL HE FEELS THERE IS A VERY GOOD CHANCE THAT THE SALINE THAT IS PUMPED IN DURING ARTHROSCOPY¿S JUST NATURALLY SUCKED THE FRAGMENT UP INTO THE NEPTUNE AND IS HOPING IT WAS REMOVED FROM THE PATIENTS BODY THAT WAY". THE FRAGMENT WAS SEARCHED FOR OUTSIDE OF THE PATIENT BUT NOT FOUND. THE SURGERY WAS COMPLETED AS NORMAL, WITH THE USE OF "ANOT

    Patient problems: Foreign Body In Patient

    MAUDE report 20937813 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →