Skip to content

BioScrew HyperFlex C8027DRecalls & reported problems

Conmed Corporation · BIOSCREW HYPERFLEX GUIDEWIRE, NITINOL 18.5 IN, STERILE, 1.1 x 470 mm (.045 x 18.5 in) · Class 1

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10845854041727), received between Mar 25, 2024 and Mar 25, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Fragmentation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 25, 2024Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER, THAT THE C8027D BIOSCREW HYPERFLEX GUIDEWIRE NITINOL 18.5 IN STER WAS BEING USED DURING AN ACL RECONSTRUCTION PROCEDURE ON (B)(6) 2024 DATE WHEN IT WAS REPORTED ¿THE ACL GUIDE WIRE WAS IN THE TIBIA, SCREW WAS INSERTED, THE GUIDE WIRE WAS STUCK, UPON PULLING THE TIP BROKE AND HAD TO BE RETRIEVED. FLUOROSCOPY HAD TO BE BROUGHT IN TO CERTIFY THE FRAGMENT WAS REMOVED.¿. IT WAS ALSO REPORTED ¿THEY BACKED OUT THE SCREW AND GRABBED THE FRAGMENT WITH A SNAP.¿ THE PROCEDURE WAS COMPLETED WITH A 10-MINUTE DELAY. THE CURRENT STATUS OF THE PATIENT WAS REPORTED AS ¿UNHARMED, WILL BE DISCHARGED AS NORMAL.¿. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION TO THE PATIENT OR USER. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18967568 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.