Skip to content

AIRSEAL IAS5-120LPRecalls & reported problems

Conmed Corporation · AIRSEAL 5 MM ACCESS PORT AND LOW PROFILE OBTURATOR WITH BLADELESS OPTICAL TIP, 120 MM LENGTH · Class 2

All models of AIRSEAL (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10845854045657), received between Dec 27, 2022 and May 28, 2026 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (4 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 14, 2026Class IIOpen, Classified

    Potential for overpressure alerts.

    Z-2467-2026 · Root cause: Process change control · Recall details · FDA record ↗

  2. May 14, 2026Class IIOpen, Classified

    Potential for overpressure alerts.

    Z-2470-2026 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Fragmentation4

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Foreign Body In Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 28, 2026Material Fragmentation

    THE CUSTOMER REPORTED THAT THE DEVICE IAS5-120LP, 5 MM ACCESS PORT & LOW-PROFILE OBTURATOR WAS BEING USED ON (B)(6) 2026 DURING A HYSTERECTOMY AND THE ¿CANNULA BROKE MIDWAY UP TO THE SHAFT OUTER LUMEN CRACKED WITH SMALL 2X8 PIECE MISSING. MISSING PIECE COULD NOT BE LOCATED. X-RAY TAKEN WITHOUT IDENTIFICATION OF PIECE IN PATIENT ABDOMEN.¿ PER FURTHER ASSESSMENT IT WAS REPORTED "WE CANNOT CONFIRM WHETHER OR NOT THE FRAGMENT OF THE CANNULA WAS RETAINED OR NOT RETAINED WITHIN THE PATIENT. A KUB INTRAOPERATIVE X-RAY WAS TAKEN IN AN ATTEMPT TO LOCATE THE FRAGMENT. HOWEVER, CONFIDENCE WAS LOW FOR SEEING THE FRAGMENT ON RADIOGRAPHY DUE TO THE LACK OF OR LIMITED RADIOPACITY OF THE MATERIAL AND SMALL SIZE OF THE FRAGMENT. THE PIECE COULD POSSIBLY HAVE BEEN RETAINED." THERE WAS NO IMPACT OR INJURY TO THE PATIENT OR USER. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE UNKNOWN DEVICE. THERE WAS A 60 TO 90-MINUTE DELAY TO THE PROCEDURE. THIS REPORT IS BEING RAISED DUE TO THE REPORTED INJURY OF PATIENT COULD POSSIBLY HAVE RETAINED A DEVICE FRAGMENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 25316322 ↗
  • MalfunctionApr 30, 2025Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE IAS5-120LP, 5 MM ACCESS PORT & LOW-PROFILE DEVICE WAS BEING USED DURING AN UNKNOWN PROCEDURE ON (B)(6) 2025 WHEN IT WAS REPORTED ¿THE OUTER CANULA OF THE TROCAR BROKE IN HALF AND A PIECE FELL INTO THE PATIENT¿S ABDOMEN. ALL OF THE PIECES WERE RETRIEVED FROM THE PATIENT WITH NO INJURY. THERE WAS A 5-10 MINUTE DELAY FOR RETRIEVAL.¿. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21933476 ↗
  • MalfunctionFeb 21, 2024Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER, THAT THE IAS5-120LP, AIRSEAL 5/120MM PORT WAS BEING USED DURING AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE WHEN IT WAS REPORTED ¿THE OUTER DIAMETER OF IAS5-120LP WHICH IS A PLASTIC PIECE BROKE OFF AND FELL INTO THE PATIENT DURING SURGERY. THE ITEM WAS RETRIEVED WITH NO INJURY TO THE PATIENT WITH MINIMAL DELAY.¿. THE PROCEDURE WAS COMPLETED WITHOUT THE USE OF AN ALTERNATE DEVICE, AND MINIMAL DELAY WAS REPORTED. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION TO THE PATIENT OR USER. THIS REPORT IS BEING RAISED ON THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18751231 ↗
  • MalfunctionDec 27, 2022Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE DEVICE, IAS5-120LP, AIRSEAL 5/120MM PORT WAS BEING USED ON 7DEC22 DURING A FIBROID HYSTERECTOMY PROCEDURE AND ¿DR. WAS ASSISTING ME IN A FIBROID HYSTERECTOMY TURNED ADVANCED ENDOMETRIOSIS CASE YESTERDAY IN THE RICHMOND OR. WE WERE USING THE AIRSEAL SYSTEM FOR INSUFFLATION. WHILE IT WAS A DIFFICULT CASE, THERE WAS NOT UNDUE TORQUE OR PRESSURE PLACED ON THE AIRSEAL TROCAR. ABOUT 4 HOURS INTO THE CASE WE SUDDENLY LOST INSUFFLATION AND AN ERROR MESSAGE OCCURRED WITH THE MACHINE. WE RESTARTED THE AIRSEAL SYSTEM BUT IT TOOK TOO LONG ¿ 5 MINUTES?- TO INITIALIZE SO WE DECIDED TO MOVE TO CONVENTIONAL INSUFFLATION. SHORTLY THEREAFTER THE INNER TUBE AND EXTERNAL PORTION OF THE PORT FELL OUT OF THE ABDOMEN. WHEN I PICKED IT UP I REALIZED THERE WAS A PIECE MISSING. WE SEARCHED THE ABDOMEN AND FOUND IT IN THE RIGHT UPPER QUADRANT. DR. WAS ABLE TO REMOVE THE PIECE THROUGH THE OPEN VAGINAL CUFF AND PUZZLE-PIECE IT WITH THE PORTION THAT FELL OUT TO CONFIRM THERE WERE NO SHARDS IN THE ABDOMEN. WE COMPLETED THE CASE WITH CONVENTIONAL INSUFFLATION.¿. THERE WAS NO IMPACT OR INJURY TO THE PATIENT. THE AIRSEAL WILL BE LISTED AS A CONCOM

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16050324 ↗

Other Laparoscope, general & plastic surgery devices

All Laparoscope, general & plastic surgery devices →

Sources and updates