Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10845854045756), received between Mar 3, 2017 and Mar 19, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Material Fragmentation1

Patient problems

  • No Consequences Or Impact To Patient2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 19, 2026Material Fragmentation

    A DISTRIBUTOR REPORTED ON BEHALF OF A CUSTOMER THAT THE SMI-02N, SPECTRUM AUTOPASS SUTURE PASSER NEEDLE DEVICE WAS USED IN A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2026, AND ¿THE NEEDLE BECAME JAMMED AND BROKE INSIDE THE PATIENT.¿. THE PROCEDURE WAS COMPLETED WITH THE USE OF AN ALTERNATE SAME DEVICE. FURTHER ASSESSMENT FOUND ¿THE NEEDLE FRAGMENTED INSIDE THE PATIENT, BUT THE FRAGMENTS WERE REMOVED USING HOSPITAL INSTRUMENTS.¿. THERE WAS NO REPORT OF INJURY, MEDICAL/SURGICAL INTERVENTION OR EXTENDED HOSPITALIZATION REQUIRED FOR THE PATIENT. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24638964 ↗
  • MalfunctionDec 5, 2017Break; Break

    A CONMED AFFILIATE REPORTED ON BEHALF OF A CUSTOMER THAT THE TIP OF THE SMI-02N NEEDLE BROKE OFF DURING PRE-OP TESTING PRIOR TO A ROTATOR CUFF PROCEDURE. THIS ALLEGED MALFUNCTION DID NOT CAUSE HARM TO THE PATIENT OR A SURGICAL DELAY AS IT WAS FOUND BEFORE THE PROCEDURE BEGAN. TO DATE, ADDITIONAL INFORMATION REGARDING THE PATIENT AND EVENT HAVE NOT BEEN PROVIDED. THIS REPORT IS RAISED ON THE BASIS OF A REPORTED DEVICE MALFUNCTION WITH POTENTIAL FOR PATIENT INJURY WITH RECURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7084681 ↗
  • MalfunctionMar 3, 2017Break; Break

    THE SALES REPRESENTATIVE REPORTED THAT WHILE ATTENDING A SHOULDER ARTHROSCOPY PROCEDURE, THE SPECTRUM AUTOPASS SUTURE PASSER AND THE SPECTRUM AUTOPASS NEEDLE WERE USED TO PASS HIFI SUTURE. WHILE PASSING THE SUTURE THE TIP OF THE NEEDLE BROKE OFF INTO THE PATIENT'S SHOULDER. THE TIP WAS RETRIEVED USING A GRASPER. THE PROCEDURE WAS COMPLETED WITH A 1-MINUTE DELAY REPORTED AND NO PATIENT INJURY OR ANY FURTHER SURGICAL COMPLICATION. THIS REPORT IS FILED ON THE BASIS OF POTENTIAL FOR PATIENT INJURY WITH RECURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6375951 ↗