AIRSEAL IAS12-150LPiRecalls & reported problems
Conmed Corporation · AirSeal 12 mm Access Port and Palm Grip Obturator with Bladeless Optical Tip, 150 mm Length · Class 2
All models of AIRSEAL (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10845854052006), received between Mar 2, 2023 and Aug 2, 2024 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Material Fragmentation” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 14, 2026Class IIOpen, Classified
Potential for overpressure alerts.
Z-2472-2026 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Fragmentation4
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 2, 2024Material Fragmentation; Material Fragmentation
THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE IAS12-150LPI, AIRSEAL 12/150MM LPI PORT WAS BEING USED DURING A VIDEO ASSISTED THORACIC SURGERY ON (B)(6) 2024 AND ¿I WAS RECENTLY MADE AWARE OF AN INCIDENT ON (B)(6) 2024 AT (B)(6) HOSPITAL WHERE A CORNER OF THE RUBBER VALVE ON THE SOUND CAP OF AN AIRSEAL 12MM TROCAR RIPPED OFF AND WAS FOUND IN THE PATIENT¿S ABDOMEN. THE CIRCULATION NURSE REPORTS THAT THE RUBBER VALVE RIPPED AFTER INSERTING AND REMOVING A STAPLER FROM THE TROCAR. THEY ALSO REPORT THEY HAVE BEEN OPERATING WITH THE SAME STAPLER FOR A YEAR AND THIS HAS NOT HAPPENED BEFORE. I HAVE RECOMMENDED TO THE ACCOUNT THAT THEY SHOULD REMOVE THE SOUND CAP BEFORE INSERTING THE STAPLER OR REMOVING SPECIMEN¿¿. PER FURTHER ASSESSMENT IT WAS FOUND THAT THE FRAGMENT WAS RETRIEVED AND ¿THE PATIENT IS FINE, NO HARM DONE TO THE PATIENT.¿. THERE WAS NO EXTENDED STAY AT THE HOSPITAL FOR THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A 5 MINUTE DELAY. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19892230 ↗MalfunctionAug 21, 2023Material Fragmentation; Material Fragmentation
THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE IAS12-150LPI, AIRSEAL 12/150MM LPI PORT, WAS BEING USED DURING A GYNECOLOGICAL ROBOT PROCEDURE ON (B)(6) 2023 WHEN IT WAS REPORTED, ¿THE CUSTOMER NOTICED THAT AT THE END OF THE GYNECOLOGICAL ROBOT PROCEDURE, A PIECE OF THE EXTERNAL DOUBLE-WALLED CANNULA WAS MISSING. AFTER SEARCHING INSIDE THE ABDOMEN, THEY FOUND THE PIECE OF PLASTIC.¿. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT OR USER. THE PROCEDURE WAS COMPLETED AS PLANNED. ALL FRAGMENTATION WAS RECOVERED FROM THE PATIENT PER THE REPORTER. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17583219 ↗MalfunctionAug 21, 2023Material Fragmentation; Material Fragmentation
THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE IAS12-150LPI, AIRSEAL 12/150MM LPI PORT, WAS BEING USED DURING A GYNECOLOGICAL ROBOT PROCEDURE ON (B)(6) 2023, WHEN IT WAS REPORTED, ¿THE CUSTOMER NOTICED THAT AT THE END OF THE GYNECOLOGICAL ROBOT PROCEDURE, A PIECE OF THE EXTERNAL DOUBLE-WALLED CANNULA WAS MISSING. AFTER SEARCHING INSIDE THE ABDOMEN, THEY FOUND THE PIECE OF PLASTIC.¿ THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT OR USER. THE PROCEDURE WAS COMPLETED AS PLANNED. ALL FRAGMENTATION WAS RECOVERED FROM THE PATIENT PER THE REPORTER. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17583216 ↗MalfunctionMay 5, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE IAS12-150LPI, AIRSEAL 12/150MM LPI PORT WAS BEING USED ON (B)(6) 2023 DURING A ROBOTIC PARTIAL NEPHRECTOMY AND ¿THE CONMED/AIRSEAL REP WAS PRESENT DURING THIS CASE. THE STAFF WAS USING A CLIP APPLIER INSTRUMENT THROUGH THE IAS12-150LPI AND NOTICED THAT ONE OF THE BLUE PLASTIC FLAPS FROM THE NOISE REDUCTION CAP HAD BECOME DISLODGED FROM THE WHITE CAP HOUSING AND WAS LAYING INSIDE THE ABDOMINAL CAVITY. THIS FLAP WAS RETRIEVED WITH A GRASPER AND REMOVED FROM THE PATIENT. SURGEON DR. SAID THERE WAS NO INJURY TO PATIENT. THE SURGICAL ASSIST WHO WAS HANDLING THE CLIP APPLIER AND PUTTING IT THROUGH THE AIRSEAL PORT, SAID THAT PERHAPS THE BUSINESS END OF THE CLIP APPLIER INSTRUMENT HAD GOTTEN CAUGHT ON THIS BLUE PLASTIC FLAP AND PERHAPS DISLODGED IT. HE NOTICED A PUNCTURE HOLE/MARK ON THE BLUE FLAP.¿. THERE WAS NO INJURY OR IMPACT TO THE PATIENT OR USER. THERE WAS NO MEDICAL/SURGICAL INTERVENTION REQUIRED FOR THE PATIENT. THE PROCEDURE WAS COMPLETED WITH AN UNKNOWN ALTERNATE DEVICE. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16877855 ↗MalfunctionMar 2, 2023Material Fragmentation; Material Fragmentation
THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE DEVICE IAS12-150LPI AIRSEAL 12/150MM LPI PORT WAS BEING USED (B)(6) 2023 AND ¿THE BLUE PLASTIC FELL OFF OF THE SOUNDCAP AND INTO THE PATIENT'S ABDOMEN DURING THE CASE. THEY STATED THAT THIS WAS DURING A ROBOTIC CASE BUT CANNOT REMEMBER WHAT INSTRUMENT WAS USED DOWN THE AS PORT. THEY STATED THAT THEY WERE ABLE TO RETRIEVE THE PIECE OF PLASTIC AND THAT THE PATIENT WAS OK.¿ THERE WAS NO REPORT OF INJURY OR IMPACT FOR THE PATIENT. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16467048 ↗
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page