VELA 16530-00 (UDI 10846446001600)Recalls & reported problems

Carefusion Corporation · VELA CF,DISS,GRN O2,TYPE K,ENG OVERLAY · Class 2

All models of VELA (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Vyaire VELA ventilator is a continuous ventilator for facility use. Reports describe mechanical problems, alarms such as motor fault and vent inoperative triggering during setup, failed leak tests and delivered volumes lower than the set amount. There are 111 reports, including 41 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
111
Deaths reported
0
Injuries reported
41
Recalls naming it
0

Summary

The FDA's MAUDE database contains 111 adverse event reports that list this exact device (DI 10846446001600), received between Nov 2, 2018 and Apr 18, 2024 — classified as 41 injuries, 70 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (98 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem98
  • Calibration Problem10
  • Leak/Splash1
  • No Device Output1
  • Use of Device Problem1

Patient problems

  • No Consequences Or Impact To Patient51
  • No Clinical Signs, Symptoms or Conditions33
  • No Patient Involvement28

Recent adverse event reports

The 25 most recent of 111 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 18, 2024No Device Output; No Device Output

    IT WAS REPORTED TO VYAIRE MEDICAL THE VELA VENTILATOR PRESENTED WITH A VENT INOP ALARM. ALARM OCCURRED DURING SETUP. NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19139592 ↗
  • MalfunctionDec 21, 2023Use of Device Problem; Use of Device Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VELA VENTILATOR HAD A MOTOR FAULT ALARM. FURTHERMORE, THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18375399 ↗
  • MalfunctionOct 19, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VELA VENTILATOR IS FAILING LEAK TEST. ADDITIONALLY, THE UNIT INFLATES THE LUNG BUT READS 0.0 FOR THE RESULT. FURTHERMORE, THERE WAS NO INFORMATION REGARDING PATIENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17969451 ↗
  • MalfunctionJul 28, 2023Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT A LOW TIDAL VOLUME ON THE VELA VENTILATOR WAS BEING 250 WHILE SET AT 400 PRIOR TO PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17411143 ↗
  • MalfunctionMar 2, 2023Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VELA VENTILATOR HAS VENT INOP, MOTOR FAULT, LOW PIP(PEAK INSPIRATORY PRESSURE), HIGH PIP(PEAK INSPIRATORY PRESSURE) AND TRANSDUCER FAULT ALARMS TRIGGERED WHEN UNIT WAS POWERED ON. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16473061 ↗
  • MalfunctionJan 24, 2023Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT IMMEDIATELY WHEN THE VELA VENTILATOR WAS POWERED ON, HIGH RATE, HIGH PIP (PEAK INSPIRATORY PRESSURE), LOW VE (MINUTE VENTILATION) AND XDCR (TRANSDUCER) FAULT ALARMS WERE TRIGGERED DURING SET UP PRIOR TO PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16237244 ↗
  • InjuryOct 18, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT TRANSDUCER FAULT ALARM OCCURRED DURING USE ON A PATIENT ON THE VELA VENTILATOR. THE CUSTOMER REPORTED THE PATIENT WAS PUT ON A DIFFERENT VENTILATOR. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15629310 ↗
  • MalfunctionOct 18, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT TRANSDUCER FAULT, LOW PEAK INSPIRATORY PRESSURE, AND LOW MINUTE VENTILATION ALARMS OCCURRED ON THE VELA VENTILATOR. THE CUSTOMER REPORTED THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15629262 ↗
  • InjuryOct 15, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT VENTILATOR INOPERABLE, MOTOR FAULT, LOW PEAK INSPIRATORY PRESSURE, LOW MINUTE VENTILATION AND HIGH RATE ALARMS OCCURRED DURING USE ON A PATIENT ON THE VELA VENTILATOR. THE CUSTOMER REPORTED THE PATIENT WAS PUT ON A DIFFERENT VENTILATOR. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15612409 ↗
  • MalfunctionAug 30, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VELA VENTILATOR HAD EXPERIENCED LOW O2 INLET PRESSURE . THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15325361 ↗
  • MalfunctionAug 24, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VELA VENTILATOR ALARMED VENT INOP (VENTILATOR INOPERABLE). THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15289090 ↗
  • MalfunctionMay 24, 2022Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED VENTILATOR INOPERABLE AND MOTOR FAULT ALARMS DURING USE ON A PATIENT ON THE VELA VENTILATOR. THE CUSTOMER REPORTED THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14492693 ↗
  • InjuryMay 23, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VELA VENTILATOR ALARMED "XDCR FAULT (TRANSDUCER FAULT), LOW PIP (POSITIVE INSPIRATORY PRESSURE) AND LOW VE (VENTILATION) WHILE ON A PATIENT. THE PATIENT WAS TRANSFERRED TO ANOTHER VENTILATOR. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14478803 ↗
  • MalfunctionMay 23, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VELA VENTILATOR ALARMED TRANSDUCER FAULT. THE CUSTOMER CONFIRMED THAT THERE IS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14478458 ↗
  • InjuryMar 20, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT WHILE THE VELA VENTILATOR WAS RUNNING, "HIGH PIP" (PEAK INSPIRATORY PRESSURE), "LOW PIP", "HIGH PEEP" (POSITIVE END-EXPIRATORY PRESSURE), "HIGH RATE" AND "XDCR FAULT" (TRANSDUCER FAULT) ALARMS OCCURRED. THE ISSUE OCCURRED DURING PATIENT-USE AND THE DEVICE WAS REPLACED WITH ANOTHER VENTILATOR. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13831802 ↗
  • InjuryFeb 10, 2022Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE VELA VENTILATOR ALARMED VENT INOP (VENTILATOR INOPERABLE), MOTOR FAULT, XDCR FAULT (TRANSDUCER FAULT) AND LOW PIP (PEAK INSPIRATORY PRESSURE) WHILE ON A PATIENT. THE PATIENT WAS TRANSFERRED TO ANOTHER VENTILATOR. THE CUSTOMER CONFIRMED THAT THERE IS NO PATIENT HARM ASSOCIATED WITH THIS REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13507078 ↗
  • InjuryFeb 10, 2022Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT VENT INOPERABLE, MOTOR FAULT, TRANSDUCER FAULT, LOW VE (LOW MINUTE VENTILATION), AND LOW PIP (PEAK INSPIRATORY PRESSURE) ALARMS OCCURRED ON VELA VENTILATOR. THE ISSUE OCCURRED DURING PATIENT-USE AND THE DEVICE WAS REPLACED WITH ANOTHER VENTILATOR. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13506985 ↗
  • InjuryJan 6, 2022Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TRANSDUCER FAULT, HIGH RATE, HIGH PEAK INSPIRATORY PRESSURE AND LOW MINUTE VENTILATION ALARMS TRIGGERED DURING USE ON A PATIENT ON THE VELA VENTILATOR. THE CUSTOMER REPORTED THE PATIENT WAS TRANSFERRED TO ANOTHER VENTILATOR. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13178401 ↗
  • MalfunctionNov 3, 2021Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VELA VENTILATOR WAS READING TRANSDUCER FAULT. AT THIS TIME, THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12746237 ↗
  • MalfunctionOct 13, 2021Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT VELA VENTILATOR UNIT IMMEDIATELY WHEN WAS POWERED ON, HIGH RATE, HIGH PIP/PEAK INSPIRATORY PRESSURE, LOW VE/MINUTE VENTILATION AND XDCR/TRANSDUCER FAULT ALARMS TRIGGERED. THE REPORTER CONFIRMED THAT THERE IS NO PATIENT INVOLVEMENT ASSOCIATED ON THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12626235 ↗
  • InjurySep 23, 2021Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED HIGH PEAK INSPIRATORY PRESSURE AND APNEA INTERVAL ALARMS DURING USE ON A PATIENT ON THE VELA VENTILATOR. THE ISSUE OCCURRED DURING PATIENT-USE AND THE DEVICE WAS REPLACED WITH ANOTHER VENTILATOR. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12522394 ↗
  • MalfunctionSep 14, 2021Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE VELA VENTILATOR UNIT HAS XDCR FAULT/TRANSDUCER ALARM. THE REPORTER CONFIRMED THAT THERE IS NO PATIENT INVOLVEMENT ASSOCIATED ON THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12471278 ↗
  • MalfunctionSep 14, 2021Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VELA VENTILATOR HAS CIRCUIT OCCLUSION WHEN POWERED ON. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12471135 ↗
  • MalfunctionAug 11, 2021Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED CIRCUIT DISCONNECT AND LOW PEAK INSPIRATORY PRESSURE DURING USE ON A PATIENT ON THE VELA VENTILATOR. AT THIS TIME, THERE IS NO INFORMATION REGARDING PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12306575 ↗
  • InjuryAug 3, 2021Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TRANSDUCER FAULT ALARM WHILE ON A PATIENT ON THE VELA VENTILATOR. THE CUSTOMER REPORTED THE PATIENT WAS TRANSFERRED TO ANOTHER VENTILATOR. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12268384 ↗

Other Ventilator, Continuous, Facility Use devices

All Ventilator, Continuous, Facility Use devices →