Inspire 900-002-009Recalls & reported problems
Inspire Medical Systems, Inc. · The sensing lead incorporates a differential pressure sensor that detects respiratory cycles. The connector end of the lead is connected to the IPG. The sensor is implanted in between the intercostal muscle layers. The device senses relative pressure variations that correspond to the respiration cycle. The pressure waveform is monitored by the IPG and triggers stimulation therapy synchronous with inspiration · Class 3 · Implant
All models of Inspire Sleep Apnea Implant (24)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10855728005496), received between Oct 4, 2018 and Nov 19, 2025 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Sensing Problem” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (P130008); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Sensing Problem2
- Improper or Incorrect Procedure or Method1
- Malposition of Device1
- Use of Device Problem1
Patient problems
- Pneumothorax2
- Infiltration into Tissue1
- No Known Impact Or Consequence To Patient1
- Pain1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryNov 19, 2025Device Sensing Problem; Device Sensing Problem
DURING A ROUTINE FOLLOW-UP EXAMINATION, PHYSICIAN OBSERVED A DEVICE SENSING ISSUE. THE PATIENT REPORTED SEVERE, PAINFUL SHOCKING SENSATIONS DEEPER IN THE CHEST, PREVENTING THEM FROM LYING DOWN. IMAGING WAS PERFORMED, AND THE DISTAL TIP OF THE SENSING LEAD WAS SUSPECTED OF SEPARATING FROM THE LEAD BODY. PHYSICIAN BROUGHT THE PATIENT IN FOR SURGERY AND CONFIRMED THE DISTAL SENSOR TIP HAD SEPARATED FROM THE LEAD BODY AND MIGRATED. THE PHYSICIAN CONSULTED WITH A CARDIOTHORACIC SURGEON AND DEEMED RETRIEVAL WAS A HIGHER RISK TO THE PATIENT, SO THE SENSOR WAS ABANDONED. THE PHYSICIAN REMOVED THE SENSE LEAD BODY AND IMPLANTED A NEWER IPG MODEL, WHICH DOES NOT REQUIRE A SENSING LEAD, WITH NO FURTHER ISSUES.
Patient problems: Pain
MAUDE report 23596407 ↗InjuryMar 5, 2019Malposition of Device; Malposition of Device
DURING IMPLANT, THE SENSING LEAD PENETRATED THE SPACE BETWEEN THE INTERNAL INTERCOSTALS AND LUNG. THE ISSUE WAS RECOGNIZED, THE AREA WAS OVERSEWN WITH TWO LAYERS OF SUTURE. THE AREA WAS FLOODED WITH SALINE, ANESTHESIA PERFORMED VALSALVA WITHOUT ANY BUBBLES BEING SEEN. THE LEAD WAS THEN MOVED TO A DIFFERENT RIB SPACE SUCCESSFULLY. POST-OP CHEST X-RAY DID NOT SHOW PNEUMOTHORAX. THE PATIENT WAS CLINICALLY ASYMPTOMATIC AND DISCHARGED PER USUAL.
Patient problems: Infiltration into Tissue
MAUDE report 8391027 ↗InjuryOct 8, 2018Use of Device Problem; Use of Device Problem
DURING TUNNELLING, PRIOR TO PLACEMENT OF THE SENSOR LEAD, PHYSICIAN WAS USING A MALLEABLE RIBBON RETRACTOR WHICH IS NOT AN INSPIRE PRODUCT. THE RETRACTOR WAS PLACED INTO THE INTERCOSTAL SPACE BUT PLANED SLIGHTLY BELOW THE INTERNAL INTERCOSTAL FIBERS. A SLIGHT NOISE WAS NOTED AND SOME SMALL BUBBLES WERE OBSERVED. CORRECTIVE MEASURES WERE TAKEN TO MANAGE WHAT APPEARED TO BE A SMALL PNEUMOTHORAX. NO LUNG COLLAPSE OR VOLUME CHANGE WAS NOTED AND ONCE IT WAS DETERMINED THE SITUATION WAS STABLE THE SENSOR WAS PLACED AND THE CASE COMPLETED AS NORMAL. THE PATIENT WAS OBSERVED AND NO FURTHER COMPLICATIONS WERE NOTED.
Patient problems: Pneumothorax
MAUDE report 7945364 ↗InjuryOct 8, 2018Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
DURING A SURGICAL PROCEDURE UNRELATED TO INSPIRE THERAPY THE SURGEON OBSERVED THAT THE SENSING LEAD, IMPLANTED 2.5 YEARS EARLIER, WAS PLACED IN THE PLEURAL SPACE. THERE IS NO PATIENT INJURY AND THE THERAPY IS WORKING WELL, BUT IT APPEARS THAT THE PATIENT EXPERIENCED A SELF HEALING PNEUMOTHORAX DURING THE IMPLANT PROCEDURE.
Patient problems: Pneumothorax
MAUDE report 7945197 ↗MalfunctionOct 4, 2018Device Sensing Problem; Device Sensing Problem
INITIAL REPORT WAS THAT PATIENT REQUIRED DEVICE SETTING ADJSUTMENTS BUT SUBSEQUENTLY IT WAS MADE KNOWN TO INSPIRE THAT ISSUES WITH SENSING AND WAVEFORMS LED TO TWO REVISION SURGERIERS, ONE TO REPLACE THE SENSING LEAD (B)(6) 2018), AND THEN LATER (B)(6) 2018) THE IPG WAS REPLACED. THIS RESOLVED THE ISSUES AND THERE WAS NO PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7936254 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P130008. One submission often covers many models, so check the product and lot details in each record.
- Z-2515-2026 · May 11, 2026 · Open, ClassifiedBrand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The Inspire Respirato
- Z-2516-2026 · May 11, 2026 · Open, ClassifiedBrand Name: Inspire Medical Systems, Inc. Product Name: Inspire stimulation Lead Model/Catalog Number: 4063 Product Description: The Inspire Stimulation Lead (Model 4063) is design
- Z-2969-2018 · Jun 12, 2018 · TerminatedInspire II Model 3024 Implantable Pulse Generator (IPG) for Obstructive Sleep Apnea, Inspire Part #900-003-020, Product Usage: The device is an implanted component of the Inspir
Other Stimulator, Hypoglossal Nerve, Implanted, Apnea devices
- Inspire 900-007-014Inspire Medical Systems, Inc. · Stimulator, Hypoglossal Nerve, Implanted, Apnea3 deaths1,012 reports
- Inspire 900-013-001Inspire Medical Systems, Inc. · Stimulator, Hypoglossal Nerve, Implanted, Apnea2 recalls359 reports
- Inspire 900-007-008Inspire Medical Systems, Inc. · Stimulator, Hypoglossal Nerve, Implanted, Apnea1 death248 reports
- Inspire 900-009-002Inspire Medical Systems, Inc. · Stimulator, Hypoglossal Nerve, Implanted, Apnea138 reports
- Inspire 900-001-009Inspire Medical Systems, Inc. · Stimulator, Hypoglossal Nerve, Implanted, Apnea125 reports
- Inspire 900-001-011Inspire Medical Systems, Inc. · Stimulator, Hypoglossal Nerve, Implanted, Apnea56 reports
- Inspire 900-014-005Inspire Medical Systems, Inc. · Stimulator, Hypoglossal Nerve, Implanted, Apnea1 recall0 reports
- Inspire 900-013-003Inspire Medical Systems, Inc. · Stimulator, Hypoglossal Nerve, Implanted, Apnea1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page