Skip to content

Inspire 900-007-010Recalls & reported problems

Inspire Medical Systems, Inc. · The Model 3028 IPG contains electronics and a battery that are sealed inside a titanium case. The IPG is implanted subcutaneously, below the clavicle in the upper chest, and connect to the stimulation lead and sensing lead. The Model 3028 does not contain any software or firmware. All functions including the telemetry and algorithm have been designed into the hardware of the IPG. The algorithm synchronizes stimulation of the hypoglossal nerve with respiration signals. The IPG processes the same dynamic range of pressure signals (2-48 cmH2O) in order to account for pressure reading variability. Based on typical settings from the STAR pivotal trial, the longevity of the Model 3028’s battery will average 10 years although the device is smaller than the currently approved version (Model 3024). In addition the Model 3028 IPG will allow patients to safely undergo magnetic resonance imaging (MRI) under specified conditions. · Class 3 · Implant

All models of Inspire Sleep Apnea Implant (24)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10855728005762), received between Jun 22, 2021 and Jan 14, 2026 — classified as 8 injuries.

3 other recalls cover devices cleared under the same FDA submission (P130008); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem8

Patient problems

  • Pain3
  • Unspecified Infection3
  • Wound Dehiscence2
  • Erosion1
  • Speech Disorder1
  • Swelling/ Edema1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 14, 2026Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH A LUMP AT THE NECK INCISION SITE FROM THE LEAD WIRES AND PAIN AT THE IPG SITE FOLLOWING WEIGHT LOSS. PHYSICIAN BROUGHT THE PATIENT INTO THE OR AND REMOVED THE ENTIRE INSPIRE SYSTEM.

    Patient problems: Pain

    MAUDE report 24075892 ↗
  • InjuryJan 6, 2026Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH TENDERNESS AND SWELLING AT THE CHEST INCISION SITE. PHYSICIAN PRESCRIBED ANTIBIOTICS.

    Patient problems: Swelling/ Edema

    MAUDE report 23978911 ↗
  • InjuryJul 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH PERSISTENT DYSARTHRIA, NECK PAIN, AND CHEST PAIN. DUE TO THE SYMPTOMS AND THE PATIENT'S WEIGHT LOSS WITH A REDUCTION IN APNEA-HYPOPNEA INDEX (AHI), THE PATIENT ELECTED SYSTEM REMOVAL. PHYSICIAN BROUGHT THE PATIENT INTO THE OR AND REMOVED THE ENTIRE INSPIRE SYSTEM.

    Patient problems: Pain; Speech Disorder

    MAUDE report 22559623 ↗
  • InjuryOct 31, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH AN INFECTION AT THE IPG SITE. PHYSICIAN BROUGHT THE PATIENT INTO THE OR, REMOVED THE IPG, CAPPED THE REMAINING LEADS, AND CLOSED.

    Patient problems: Unspecified Infection

    MAUDE report 20577165 ↗
  • InjuryMay 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH PAIN AT THE IPG SITE. PHYSICIAN BROUGHT THE PATIENT INTO THE OPERATING ROOM AND EXPLANTED THE IPG, SENSING LEAD, AND A MAJORITY OF THE STIMULATION LEAD.

    Patient problems: Pain

    MAUDE report 19288968 ↗
  • InjuryFeb 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE ER PHYSICIAN WITH CONCERNS OF INFECTION. ER PHYSICIAN DETERMINED THAT THE PATIENT DID NOT HAVE AN INFECTION. PATIENT PRESENTED TO THEIR INSPIRE-TRAINED PHYSICIAN WITH CONCERNS OF INFECTION. PHYSICIAN EVALUATED THE PATIENT AND DETERMINED THE IPG HAD MIGRATED AND THAT THERE WAS SIGNIFICANT TETHERING FROM THE STIMULATION LEAD. PHYSICIAN BROUGHT THE PATIENT TO THE OR AND CONFIRMED INFECTION AT THE IPG SITE. PHYSICIAN REMOVED THE ENTIRE INSPIRE SYSTEM, FLUSHED THE AREA WITH ANTIBIOTICS, AND PRESCRIBED ORAL ANTIBIOTICS AFTER THE PROCEDURE WAS COMPLETED.

    Patient problems: Unspecified Infection

    MAUDE report 18785010 ↗
  • InjuryAug 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH WOUND DEHISCENCE AND DRAINAGE. PHYSICIAN PRESCRIBED ANTIBIOTICS.

    Patient problems: Wound Dehiscence

    MAUDE report 17498377 ↗
  • InjuryJun 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH EROSION AT THE SENSE LEAD INCISION SITE THROUGH WHICH THE SUTURE ANCHOR WAS VISIBLE. PHYSICIAN SURGICALLY REPLACED THE SENSE LEAD ON (B)(6) 2021, RESOLVING THE ISSUE.

    Patient problems: Erosion; Unspecified Infection; Wound Dehiscence

    MAUDE report 12042434 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P130008. One submission often covers many models, so check the product and lot details in each record.

  • Z-2516-2026 · May 11, 2026 · Open, Classified
    Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire stimulation Lead Model/Catalog Number: 4063 Product Description: The Inspire Stimulation Lead (Model 4063) is design
  • Z-2515-2026 · May 11, 2026 · Open, Classified
    Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The Inspire Respirato
  • Z-2969-2018 · Jun 12, 2018 · Terminated
    Inspire II Model 3024 Implantable Pulse Generator (IPG) for Obstructive Sleep Apnea, Inspire Part #900-003-020, Product Usage: The device is an implanted component of the Inspir

Other Stimulator, Hypoglossal Nerve, Implanted, Apnea devices

All Stimulator, Hypoglossal Nerve, Implanted, Apnea devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page