Skip to content

Medline DYND160116CRecalls & reported problems

Medline Industries, Inc. · TRAY,FOLEY,ERASE CAUTI,LTX,COUD,16FR10ML · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10884389118757), received between Dec 5, 2025 and Dec 11, 2025 — classified as 7 injuries, 1 malfunction.

The most frequently reported specific device problem is “Migration” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 8 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available3
  • Migration2
  • Device Dislodged or Dislocated1
  • Fluid/Blood Leak1
  • Inflation Problem1
  • Patient Device Interaction Problem1

Patient problems

  • Pain7
  • Swelling/ Edema4
  • Genital Bleeding1
  • Hemorrhage/Blood Loss/Bleeding1
  • High White Blood Cell Count1
  • Pelvic Pain1
  • Unspecified Blood or Lymphatic problem1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 11, 2025Migration; Migration

    ACCORDING TO THE FACILITY, ON (B)(6) 2025, PATIENT WENT TO THE EMERGENCY ROOM A FEW HOURS AFTER CHANGING CATHETER AND WAS EXPERIENCING "SUDDEN NEW ONSET OF PENILE PAIN." CATHETER BALLOON WAS BELIEVED TO BE IN THE URETHRA.

    Patient problems: Pain; Unspecified Blood or Lymphatic problem

    MAUDE report 23782600 ↗
  • InjuryDec 11, 2025Migration; Migration

    ACCORDING TO THE CUSTOMER, ON (B)(6) 2025 WITHIN "A FEW DAYS" AFTER CATHETER INSERTION, THE PATIENT BEGAN COMPLAINING OF "PENILE PAIN", AND AN ULTRASOUND SHOWED THAT THE BALLOON "IN THE PROSTATE".

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Swelling/ Edema

    MAUDE report 23780210 ↗
  • InjuryDec 11, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    ACCORDING TO THE FACILITY ON (B)(6) 2025, THE PATIENT RETURNED FROM THE EMERGENCY ROOM AFTER CATHETER MANIPULATION TO CHANGE POSITION AND A FEW HOURS LATER, WATER WAS NOTED ON BED. THE CATHETER WAS REMOVED AND WAS NOTED TO HAVE SLIT IN IT AND THEREFORE ALL THE "WATER" WAS MISSING.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 23779787 ↗
  • InjuryDec 11, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    A CATHETER WAS REMOVED AND REPLACED DUE TO SEVERE PENILE AND PELVIC PAIN.

    Patient problems: Pain; Pelvic Pain

    MAUDE report 23779267 ↗
  • InjuryDec 11, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    ACCORDING TO THE FACILITY ON (B)(6) 2025, "HE WAS COMPLAINING OF PENILE PAIN AND DISCOMFORT".

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 23778718 ↗
  • MalfunctionDec 8, 2025Inflation Problem; Inflation Problem

    BALLOON BECAME DEFLATED WHILE IN USE.

    Patient problems: Pain

    MAUDE report 23745021 ↗
  • InjuryDec 5, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    "CT IMAGING SHOWED CATHETER BALLOON WAS IN THE PROSTATE".

    Patient problems: Pain

    MAUDE report 23732473 ↗
  • InjuryDec 5, 2025Fluid/Blood Leak; Device Dislodged or Dislocated; Fluid/Blood Leak; Device Dislodged or Dislocated

    ACCORDING TO THE FACILITY, A FOLEY CATHETER WAS INSERTED "WITHOUT ISSUE," AND WITHIN A FEW DAYS, THE PATIENT WAS COMPLAINING OF "PENILE PAIN" AND LAB WORK SHOWED "INCREASED WHITE BLOOD CELL COUNT WITHOUT KNOWN CAUSE." A CT WAS PERFORMED AND THE "BALLOON WAS NOTED TO BE IN THE PROSTATE."

    Patient problems: Genital Bleeding; High White Blood Cell Count; Pain; Swelling/ Edema

    MAUDE report 23729918 ↗

Other Catheter care tray devices

All Catheter care tray devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page