Prelude® Short Sheath 10884450015329Recalls & reported problems
Merit Medical Systems, Inc. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884450015329), received between Apr 6, 2022 and Dec 12, 2023 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Material Separation” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation3
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryDec 12, 2023Material Separation; Material Separation
THE CUSTOMER REPORTS DURING A FISTULAGRAM/ANGIO WHEN THE MD REMOVED THE SHORT SHEATH OUT AT COMPLETION, THE BLACK TIP WAS BROKEN OFF INSIDE OF THE PATIENT. THE MD HAD TO THEN ATTEMPT TO RETRIEVE THE TIP AND WAS SUCCESSFUL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18314755 ↗InjurySep 30, 2022Material Separation; Material Separation
THE ACCOUNT ALLEGES THAT AFTER THE PHYSICIAN REMOVED THE SHEATH, IT WAS NOTED THAT THE SHEATHS BLACK MARKER TIP HAD DETACHED. THE CLINICAL STAFF LOOKED FOR THE FOREIGN BODY UNDER FLUOROSCOPY AND WAS ABLE TO SUCCESSFULLY IDENTIFY THE MARKER TIP WITHIN THE PATIENT'S ARTERIOVENOUS FISTULA. THE PHYSICIAN DEPLOYED A COVERED STENT TO SUCCESSFULLY PIN THE FOREIGN BODY IN PLACE, PREVENTING FOREIGN BODY MIGRATION. THE PATIENT TOLERATED THE PROCEDURE WELL, WITH NO ADDITIONAL CONSEQUENCES TO REPORT.
Patient problems: Foreign Body In Patient
MAUDE report 15510034 ↗MalfunctionApr 6, 2022Fracture; Material Separation; Fracture; Material Separation
THE ACCOUNT ALLEGES THAT WHILE ATTEMPTING VASCULAR ACCESS FOR A ROUTINE AV FISTULAGRAM PROCEDURE, THE HUB OF THE 7F SHEATH INTRODUCER/DILATOR DETACHED WHILE ATTEMPTING TO ACQUIRE VASCULAR ACCESS VIA "SELDINGER TECHNIQUE". NO PATIENT INJURY TO REPORT. THE SHEATH WAS EXCHANGED FOR A NEW ONE AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14024779 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page