Magellan 8881850110Recalls & reported problems
Cardinal Health, Inc. · Hypodermic Safety Needle · Class 2
All models of Magellan (29)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10884521000513), received between Sep 15, 2017 and Oct 10, 2022 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Device Contamination with Chemical or Other Material3
- Fail-Safe Did Not Operate1
- Shielding Failure1
- Use of Device Problem1
Patient problems
- No Consequences Or Impact To Patient4
- Needle Stick/Puncture2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 10, 2022Fail-Safe Did Not Operate; Fail-Safe Did Not Operate
THE CUSTOMER REPORTED THAT A PARAMEDIC SUFFERED A NEEDLE STICK FROM A DIRTY NEEDLE. THE SAFETY CAP FAILED ON THE DEVICE. PER ADDITIONAL INFORMATION RECEIVED, THE EMPLOYEE HAS RESUMED WORKING AND IS BEING FOLLOWED BY OCCUPATIONAL HEALTH FOR BLOODBORNE EXPOSURE.
Patient problems: Needle Stick/Puncture
MAUDE report 15573553 ↗MalfunctionNov 7, 2019Use of Device Problem; Use of Device Problem
THE CUSTOMER REPORTED THAT THE DEVICE IS PULLING IN AIR/BUBBLES WHEN DRAWING BLOOD. ADDITIONAL INFORMATION RECEIVED STATED THIS HAS HAPPENED SPORADICALLY OVER THE PAST MONTHS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9290150 ↗MalfunctionDec 26, 2017Shielding Failure; Shielding Failure
THE CUSTOMER REPORTS THAT A NURSE SUSTAINED A NEEDLE STICK WITH PRODUCT WHICH FAILED TO LOCK AFTER SAFETY NEEDLE COVER WAS ADVANCED OVER THE NEEDLE.
Patient problems: Needle Stick/Puncture
MAUDE report 7145586 ↗MalfunctionSep 15, 2017Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
THE CUSTOMER REPORTS THAT AIR IS BEING DRAWN INTO THE SYRINGE WHILST OBTAINING SAMPLES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6870199 ↗MalfunctionSep 15, 2017Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
THE CUSTOMER REPORTS THAT AIR IS BEING DRAWN INTO THE SYRINGE WHILST OBTAINING SAMPLES. THIS AFFECTS THE QUALITY OF THE SAMPLE RESULTING IN THE PATIENT HAVING TO HAVE ANOTHER VENEPUNCTURE TO OBTAIN A SECOND BLOOD SAMPLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6869719 ↗MalfunctionSep 15, 2017Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
THE CUSTOMER REPORTS THAT AIR IS BEING DRAWN INTO THE SYRINGE WHILST OBTAINING SAMPLES. THIS AFFECTS THE QUALITY OF THE SAMPLE RESULTING IN THE PATIENT HAVING TO HAVE ANOTHER VENEPUNCTURE TO OBTAIN A SECOND BLOOD SAMPLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6869405 ↗
FDA clearance and classification
The FDA regulates product code FMI (needle, hypodermic, single lumen) under 21 CFR 880.5570 as a Class II device. 21 CFR 880.5570 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the Magellan 8881850110
Has the Magellan 8881850110 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10884521000513), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Magellan 8881850110?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Sep 15, 2017 and Oct 10, 2022. In those reports, the most common specific device problems are “Device Contamination with Chemical or Other Material” (3), “Fail-Safe Did Not Operate” (1) and “Shielding Failure” (1); the most common patient problem is “Needle Stick/Puncture” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Magellan 8881850110?
None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Magellan 8881850110, and is it still sold?
The FDA's device database lists Cardinal Health, Inc. as the company responsible for it (brand Magellan, model 8881850110). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Needle, hypodermic, single lumen devices
- BD Nano 2nd Gen 320550Becton Dickinson Pen Limited · Needle, hypodermic, single lumen1,892 reports
- BD SafetyGlide 305916Becton, Dickinson And Company · Needle, hypodermic, single lumen1,768 reports
- BD Nano Ultra-Fine 320122Becton Dickinson Pen Limited · Needle, hypodermic, single lumen566 reports
- N/A 305106Becton, Dickinson And Company · Needle, hypodermic, single lumen1 recall465 reports
- N/A 305110Becton, Dickinson And Company · Needle, hypodermic, single lumen418 reports
- BD Safe-Clip 328235Becton, Dickinson and Company Inc · Needle, hypodermic, single lumen268 reports
- BD Ultra-Fine 320109Becton Dickinson Pen Limited · Needle, hypodermic, single lumen208 reports
- BD AutoShield Duo 329515Becton, Dickinson And Company · Needle, hypodermic, single lumen200 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page