Magellan 8881833558 (UDI 10884521005846)Recalls & reported problems

Cardinal Health, Inc. · 3 mL Syringe with Hypodermic Safety Needle · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
16
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10884521005846), received between Nov 1, 2017 and Apr 12, 2024 — classified as 1 injury, 15 malfunctions.

The most frequently reported specific device problem is “Crack” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack6
  • Break4
  • Detachment of Device or Device Component3
  • Fluid/Blood Leak2
  • Disconnection1
  • Leak/Splash1
  • Material Deformation1

Patient problems

  • No Consequences Or Impact To Patient7
  • No Patient Involvement6
  • Insufficient Information1
  • No Information1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 12, 2024

    THE ISSUE IS THAT THE NEEDLE IS NOT ALWAYS ¿PREATTACHED¿ WELL AND BECOMES DETACHED WHEN TRYING TO DRAW OUT THE MEDICATION FOR USE DURING OUR IONTOPHORESIS TREATMENT, THUS FORCING THE CLINICIAN TO EITHER DUMP IT INTO THE SHARPS CONTAINER AND GET ANOTHER ONE OR TRY TO PUSH THE NEEDLE BACK INTO THE SYRINGE (BY HOLDING ONTO THE NON-SHARP PART OF THE NEEDLE WHERE IT INSERTS INTO THE SYRINGE. THIS IS AN UNSAFE PRACTICE. TWO LOT NUMBERS ARE AFFECTED, 118056 AND S14684. NOT ALL DEVICES IN THESE LOTS EXHIBITED FAILURES. THE FAILURES WERE SPORADIC.

    Patient problems: Insufficient Information

    MAUDE report 19095775 ↗
  • MalfunctionSep 18, 2020Crack; Crack

    THE CUSTOMER REPORTED THAT THE MAGELLAN SYRINGE WAS BROKEN AT THE HUB. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10556916 ↗
  • MalfunctionSep 15, 2020Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED THE NEEDLE WAS NOT FULLY TIGHTENED ON THE SYRINGE AND IT WAS NOT CHECKED BEFORE USING THE SYRINGE/NEEDLE. THE MEDICATION SQUIRTED OUT AROUND THE CONNECTION. THERE WAS NO PATIENT HARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10534618 ↗
  • InjurySep 15, 2020Fluid/Blood Leak; Detachment of Device or Device Component; Fluid/Blood Leak; Detachment of Device or Device Component

    THE CUSTOMER REPORTED THE NEEDLE WAS NOT FULLY TIGHTENED ON THE SYRINGE AND THE MEDICATION SQUIRTED OUT AROUND THE CONNECTION. THE NEEDLE SHOT OFF THE SYRINGE CAUSING A NEEDLE STICK INJURY.

    Patient problems: No Information

    MAUDE report 10534240 ↗
  • MalfunctionMar 4, 2020Crack; Crack

    THE CUSTOMER REPORTED THAT THERE WAS A CRACK ON THE SYRINGE BARREL.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9789282 ↗
  • MalfunctionFeb 27, 2020Crack; Crack

    THE CUSTOMER REPORTED THAT THE SYRINGE BARREL WAS CRACKED.

    Patient problems: No Patient Involvement

    MAUDE report 9763592 ↗
  • MalfunctionFeb 21, 2020Break; Break

    THE CUSTOMER REPORTED THAT THE LUER TIP WHERE THE NEEDLE AND SYRINGE CONNECT IS DAMAGED. THERE WAS NO PATIENT HARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9738796 ↗
  • MalfunctionNov 21, 2019Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component

    THE CUSTOMER REPORTED TO THEIR DISTRIBUTOR THAT TWO SYRINGE/NEEDLE COMBOS HUBS WERE CRACKED AND BROKEN OFF FROM THE SYRINGE.

    Patient problems: No Patient Involvement

    MAUDE report 9356803 ↗
  • MalfunctionNov 19, 2019Crack; Crack

    THE CUSTOMER REPORTED TO THEIR DISTRIBUTOR THAT THE SYRINGE BARREL IS CRACKED.

    Patient problems: No Patient Involvement

    MAUDE report 9342113 ↗
  • MalfunctionNov 19, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED TO THE DISTRIBUTOR THAT THE NEEDLE WAS BROKEN. ACCORDING TO THE PHOTOGRAPH PROVIDED, THE HUB WAS BROKEN/DETACHED FROM THE SYRINGE.

    Patient problems: No Patient Involvement

    MAUDE report 9341727 ↗
  • MalfunctionNov 6, 2019Disconnection; Disconnection

    THE CUSTOMER REPORTED THAT THE SYRINGES WERE FOUND DETACHED FROM THE LUER INSIDE OF THE PACKAGE BEFORE THE PRODUCT WAS OPENED.

    Patient problems: No Patient Involvement

    MAUDE report 9285627 ↗
  • MalfunctionNov 6, 2019Crack; Crack

    THE CUSTOMER REPORTED THAT THE HUB OF THE NEEDLE WAS CRACKED. ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER STATED THAT THE ISSUE WAS DISCOVERED BEFORE USE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Patient Involvement

    MAUDE report 9284103 ↗
  • MalfunctionNov 4, 2019Break; Break

    THE CUSTOMER REPORTED THAT THEY DISCOVERED A BROKEN NEEDLE HUB.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9271950 ↗
  • MalfunctionOct 29, 2019Break; Break

    THE CUSTOMER REPORTED THAT THE NEEDLE WAS DEFECTIVE. BASED ON THE PHOTOS PROVIDED BY THE CUSTOMER, THE HUB OF THE NEEDLE IS COMPLETELY CRACKED INTO TWO PIECES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9253882 ↗
  • MalfunctionAug 16, 2019Break; Material Deformation; Break; Material Deformation

    THE CUSTOMER REPORTED THE TIP OF THE SYRINGE WAS BROKEN AND MELTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8901272 ↗
  • MalfunctionNov 1, 2017Leak/Splash; Leak/Splash

    THE CUSTOMER STATES THAT DURING VACCINE ADMINISTRATION, THE NEEDLE SEPARATED FROM SYRINGE AND VACCINE RAN DOWN CLIENT'S ARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6993123 ↗

Other Needle, hypodermic, single lumen devices