Puritan Bennett 4-840220AIQC-ENRecalls & reported problems

Covidien LP · 800 Series Ventilator System,Model 840 · Class 2

All models of Puritan Bennett (119)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10884521076976), received between Jun 14, 2017 and Sep 26, 2018 — classified as 5 malfunctions.

The most frequently reported specific device problem is “No Flow” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • No Flow4
  • Appropriate Device Problem Term/Code Not Available1
  • No Display/Image1

Patient problems

  • No Consequences Or Impact To Patient4
  • No Code Available3
  • No Patient Involvement1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 26, 2018No Flow; Appropriate Device Problem Term/Code Not Available; No Flow; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE 840 VENTILATOR STOPPED VENTILATION AND THE SYSTEMS LOCKED OUT. THERE WAS NO INJURY TO THE PATIENT AS A RESULT OF THE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7912672 ↗
  • MalfunctionSep 26, 2018No Flow; No Flow

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE 840 VENTILATOR STOPPED VENTILATION AND THE SYSTEMS WERE LOCKED OUT. THERE WAS NO INJURY TO THE PATIENT AS A RESULT OF THE EVENT.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 7912648 ↗
  • MalfunctionSep 26, 2018No Flow; No Flow

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE 840 VENTILATOR STOPPED VENTILATION AND THE SYSTEMS WERE LOCKED OUT. THERE WAS NO INJURY TO THE PATIENT AS A RESULT OF THE EVENT.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 7910550 ↗
  • MalfunctionSep 26, 2018No Flow; No Flow

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE 840 VENTILATOR STOPPED VENTILATION AND THE SYSTEMS WERE LOCKED OUT. THERE WAS NO INJURY TO THE PATIENT AS A RESULT OF THE EVENT.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 7910518 ↗
  • MalfunctionJun 14, 2017No Display/Image; No Display/Image

    IT WAS REPORTED THAT AN 840 VENTILATOR HAD A FAULTY DISPLAY. THE VENTILATOR WAS NOT ON A PATIENT AT THE TIME OF THE EVENT. THE SERVICE ENGINEER REPLACED THE BACKLIGHT INVERTER BOARD, THE GUI (GRAPHICAL USER INTERFACE) BOARD AND DOWNLOADED SOFTWARE TO CORRECT THE ISSUE. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS.

    Patient problems: No Patient Involvement

    MAUDE report 6642042 ↗