Puritan Bennett 4-840220DIEC-SPRecalls & reported problems
Covidien LP · 800 Series Ventilator System,Model 840 · Class 2
All models of Puritan Bennett (119)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Puritan Bennett hospital ventilator provides continuous breathing support in facilities. Reports mostly describe display and startup failures, error codes and unexpected shutdowns, including smoke or a burning smell, often when no patient was connected. There are 28 reports naming its identifier, with no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 28
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 10884521077058), received between Mar 29, 2017 and Dec 6, 2022 — classified as 28 malfunctions.
The most frequently reported specific device problem is “Communication or Transmission Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Communication or Transmission Problem4
- Device Inoperable4
- Erratic or Intermittent Display4
- No Display/Image4
- No Flow4
- Device Displays Incorrect Message2
- Smoking2
- Battery Problem1
Patient problems
- No Patient Involvement22
- No Information5
- No Code Available3
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 6, 2022Smoking; Smoking
IT WAS REPORTED THAT THIS 840 VENTILATOR EMITTED SMOKE FROM THE EXPIRATORY BLOCK. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15924166 ↗MalfunctionNov 24, 2021No Flow; Output Problem; No Flow; Output Problem
IT WAS REPORTED THAT, WHEN THE 840 VENTILATOR WAS TURNED ON, THE UNIT GENERATED A BEEPING SOUND AND THE SCREEN WOULD NOT TURN ON. IT WAS ALSO REPORTED THAT UNIT WOULD NOT STOP BEEPING. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE EVENT. REPORTABILITY REASSESSED BASE ON THE PRODUCT ANALYSIS FINDINGS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12871058 ↗MalfunctionFeb 11, 2019No Flow; No Flow
IT WAS REPORTED THAT DURING SERVICE THE 840 VENTILATOR GENERATED A BACKGROUND FAULT ERROR CODE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Patient Involvement
MAUDE report 8327859 ↗MalfunctionFeb 8, 2019No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE 840 VENTILATOR ALARMED LOSS OF COMMUNICATION AND THE SCREEN WOULD NOT TURN ON; THE UNIT WAS REPORTED TO HAVE GENERATED AN INOPERABLE ERROR CODE. THE PATIENT WAS PLACED ON AN ALTERNATE VENTILATOR AND THERE WAS NO HARM TO THE PATIENT AS A RESULT OF THE EVENT.
Patient problems: No Code Available
MAUDE report 8323882 ↗MalfunctionJun 6, 2018Thermal Decomposition of Device; Smoking; Thermal Decomposition of Device; Smoking
IT WAS REPORTED THAT THE 840 VENTILATOR FAILED THE SHORT SELF TEST (SST). THE VENTILATOR EXPERIENCED SMOKE AND A BURNING SMELL. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Patient Involvement
MAUDE report 7575356 ↗MalfunctionNov 3, 2017Device Inoperable; Inappropriate or Unexpected Reset; Device Inoperable; Inappropriate or Unexpected Reset
IT WAS REPORTED THAT THE 840 VENTILATOR HAD SHUT DOWN WITHOUT ANY ALARM. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE REPORTED EVENT.
Patient problems: No Code Available
MAUDE report 7000720 ↗MalfunctionAug 24, 2017No Display/Image; No Display/Image
IT WAS REPORTED AN 840 VENTILATOR HAD A FROZEN/LOCKED SCREEN. ALTHOUGH REQUESTED IT IS UNKNOWN IF THE PATIENT WAS CONNECTED TO THE VENTILATOR AT THE TIME OF THE REPORTED EVENT. THE SERVICE ENGINEER REPLACED THE TOUCH FRAME/DISPLAY. THE SERVICE ENGINEER PERFORMED TESTING AND CALIBRATIONS AND THE UNIT OPERATED WITHIN THE MANUFACTURER'S SPECIFICATIONS.
Patient problems: No Information; No Patient Involvement
MAUDE report 6818126 ↗MalfunctionAug 21, 2017Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT DURING PREVENTATIVE MAINTENANCE, THE 840 VENTILATOR DID NOT PASS THE EXTENDED SELF-TEST (EST), THE LOWER DISPLAY DID NOT VIEW PROPERLY AND THE KEYBOARD DID NOT WORK PROPERLY (BLOCKED ERROR MESSAGE).
Patient problems: No Patient Involvement
MAUDE report 6808558 ↗MalfunctionAug 15, 2017Device Inoperable; Device Inoperable
IT WAS REPORTED THAT AN 840 VENTILATOR GENERATED A SAFETY VALVE OUT OF RANGE ERROR MESSAGE WHICH RENDERED THE DEVICE INOPERABLE. THE VENTILATOR WAS NOT ON A PATIENT AT THE TIME OF THE EVENT. A SERVICE ENGINEER INSPECTED THE DEVICE AND REPLACED THE BREATH DELIVERY (BD) PRINTED CIRCUIT BOARD (PCB). THE UNIT PASSED ALL TESTING AND OPERATED WITHIN THE MANUFACTURING SPECIFICATIONS.
Patient problems: No Patient Involvement
MAUDE report 6794985 ↗MalfunctionAug 8, 2017Device Inoperable; Device Inoperable
IT WAS REPORTED THAT AN 840 VENTILATOR BECAME INOPERATIVE. AT THIS TIME, IT IS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT. A SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND REPLACED THE GUI (GRAPHICAL USER INTERFACE) PRINTED CIRCUIT BOARD (PCB) AND UPDATED THE BACKLIGHT INVERTER PCBS TO ADDRESS THE ISSUE. THE UNIT PASSED ALL TESTING AND OPERATED WITHIN THE MANUFACTURING SPECIFICATIONS.
Patient problems: No Information; No Patient Involvement
MAUDE report 6775593 ↗MalfunctionAug 4, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT AN 840 VENTILATOR HAD A SCREEN LOCK ERROR. THERE WAS NO PATIENT INVOLVEMENT AT THE TIME OF THE EVENT.
Patient problems: No Patient Involvement
MAUDE report 6768920 ↗MalfunctionAug 1, 2017No Audible Alarm; No Flow; No Audible Alarm; No Flow
IT WAS REPORTED THAT THE 840 VENTILATOR HAD A VENT INOP ALARM FAILURE. INFORMATION ABOUT PATIENT INVOLVEMENT WAS NOT PROVIDED. THE SERVICE ENGINEER (SE) EVALUATED THE VENTILATOR AND REPLACED THE BREATH DELIVERY UNIT (BDU) PRINTED CIRCUIT BOARD (PCB) THE ANALOG INTERFACE (AI) PCB AND THE EXHALATION VALVE. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURER SPECIFICATIONS.
Patient problems: No Information; No Patient Involvement
MAUDE report 6758201 ↗MalfunctionJul 31, 2017Battery Problem; Battery Problem
ADDITIONAL INFORMATION WAS RECEIVED, THE CUSTOMER STATED THAT A NON-MEDTRONIC BATTERY WAS USED AT THE TIME OF THE REPORTED EVENT. THERE WAS NO PATIENT INVOLVEMENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Information; No Patient Involvement
MAUDE report 6755589 ↗MalfunctionJul 31, 2017No Display/Image; No Display/Image
IT WAS REPORTED THAT AN 840 VENTILATOR HAD A SCREEN BLOCKED DUE TO BEING DROPPED AND DAMAGED. THE VENTILATOR WAS NOT ON A PATIENT AT THE TIME OF THE EVENT. THE SERVICE ENGINEER REPLACED THE GUI CPU (GRAPHICAL USER INTERFACE CENTRAL PROCESS UNIT) AND BACKLIGHT INVERTER TO CORRECT THE ISSUE. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS.
Patient problems: No Patient Involvement
MAUDE report 6753946 ↗MalfunctionJul 28, 2017Failure to Power Up; Failure to Power Up
IT WAS REPORTED THAT, WHEN AN 840 VENTILATOR WAS TURNED THE DEVICE FAILED THE POST (POWER ON SELF-TEST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE EVENT OCCURRED. THE SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND REPLACED THE GRAPHIC USER INTERFACE (GUI) CENTRAL PROCESSING UNIT (CPU) PRINTED CIRCUIT BOARD (PCB) AND BACKLIGHT INVERTER PCBS. THE SE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED.
Patient problems: No Patient Involvement
MAUDE report 6752586 ↗MalfunctionJul 28, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT, AN 840 VENTILATOR GENERATED AN OXYGEN (O2) SENSOR ERROR MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE EVENT OCCURRED. THE SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND WAS NOT ABLE TO DUPLICATE THE REPORTED CONDITION. THE SE CALIBRATED THE O2 SENSOR AND PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED.
Patient problems: No Patient Involvement
MAUDE report 6752555 ↗MalfunctionJul 27, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT, AN 840 VENTILATOR HAD AN OXYGEN SENSOR MALFUNCTION DURING SST. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE EVENT OCCURRED. THE COVIDIEN SERVICE ENGINEER (SE) INSPECTED AND VERIFIED THE ISSUE REPORTED. THE SE FOUND A LEAK IN THE EXHALATION VALVE. THE SE REPAIRED THE LEAK BY RE-INSTALLING THE EXHALATION VALVE. THE UNIT PASSED ALL TESTING AND OPERATED WITHIN THE MANUFACTURES SPECIFICATIONS.
Patient problems: No Patient Involvement
MAUDE report 6747631 ↗MalfunctionJul 17, 2017Erratic or Intermittent Display; Erratic or Intermittent Display
IT WAS REPORTED THAT AN 840 VENTILATOR HAD A FAULTY DISPLAY. THERE WAS NO PATIENT INVOLVEMENT AT THE TIME OF THE EVENT. THE SERVICE ENGINEER INSPECTED THE DEVICE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB), UPGRADED THE BACKLIGHT INVERTER PCBS AND DOWNLOADED THE RECENT VERSION OF THE SOFTWARE. THE VENTILATOR PASSED ALL TESTING AND OPERATED WITHIN THE MANUFACTURER'S SPECIFICATIONS.
Patient problems: No Information
MAUDE report 6719422 ↗MalfunctionJul 11, 2017Erratic or Intermittent Display; Erratic or Intermittent Display
IT WAS REPORTED THAT, AN 840 VENTILATOR HAD AN ERRATIC DISPLAY. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE EVENT OCCURRED. THE SERVICE ENGINEER (SE) VERIFIED THE ISSUE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB). THE UNIT PASSED ALL TESTING AND OPERATED WITHIN THE MANUFACTURING SPECIFICATIONS.
Patient problems: No Patient Involvement
MAUDE report 6702876 ↗MalfunctionJun 14, 2017Device Inoperable; Device Inoperable
IT WAS REPORTED THAT AN 840 VENTILATOR WAS BLOCKED AND COULD NOT VENTILATE. THE VENTILATOR WAS NOT ON A PATIENT AT THE TIME OF THE EVENT. THE EVALUATION AND REPAIR OF THE DEVICE HAS NOT BEEN COMPLETED.
Patient problems: No Patient Involvement
MAUDE report 6641392 ↗MalfunctionJun 13, 2017No Display/Image; No Display/Image
IT WAS REPORTED THAT AN 840 VENTILATOR HAD A FAULTY DISPLAY. THE VENTILATOR WAS NOT ON A PATIENT AT THE TIME OF THE EVENT. THE SERVICE ENGINEER REPLACED THE GUI CPU (GRAPHICAL USER INTERFACE) BOARD, BACKLIGHT INVERTER AND UPLOADED SOFTWARE TO CORRECT THE ISSUE. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS.
Patient problems: No Patient Involvement
MAUDE report 6638273 ↗MalfunctionJun 5, 2017Communication or Transmission Problem; Communication or Transmission Problem
A GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB) AND BACKLIGHT INVERTERS PCB WERE RETURNED TO COVIDIEN/MEDTRONIC¿S FAILURE ANALYSIS. A VISUAL INSPECTION FOUND NO NOTABLE CONDITIONS. AN INVESTIGATION WAS PERFORMED AND THE IDENTIFIED ROOT CAUSE OF THE RE PORTED ISSUE ON THE GUI PCB WAS ISOLATED TO A COMPONENT (COAXIAL TRANSCEIVER INTERFACE U35). NO FAILURE WAS FOUND ON THE BACKLIGHT INVERTERS PCB.
Patient problems: No Patient Involvement
MAUDE report 6613367 ↗MalfunctionJun 1, 2017No Display/Image; No Display/Image
IT WAS REPORTED THAT, AN 840 VENTILATOR HAD A BLANK DISPLAY WHEN DEVICE WAS TURNED ON. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE EVENT OCCURRED. THE SERVICE ENGINEER (SE) VERIFIED THE ISSUE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB), THE BACKLIGHT INVERTER PCB¿S WERE UPDATED AND INSTALLED THE LATEST VERSION OF THE OPERATIONAL SOFTWARE. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS.
Patient problems: No Patient Involvement
MAUDE report 6604222 ↗MalfunctionMay 22, 2017Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT, AN 840 VENTILATOR HAD A LOSS OF COMMUNICATION BETWEEN THE GRAPHIC USER INTERFACE (GUI) DISPLAY AND THE BREATH DELIVERY UNIT (BDU). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED. THE SERVICE ENGINEER (SE)VERIFIED THE MALFUNCTION AND REPLACED THE BREATH DELIVERY UNIT (BDU) PRINTED CIRCUIT BOARD (PCB), THE GRAPHICAL USER INTERFACE (GUI) PCB, THE BACKLIGHT INVERTER PCBS, AND SOFTWARE WAS UPDATED TO THE CURRENT REVISION. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS.
Patient problems: No Patient Involvement
MAUDE report 6583539 ↗MalfunctionMay 15, 2017Erratic or Intermittent Display; Erratic or Intermittent Display
IT WAS REPORTED THAT, AN 840 VENTILATOR HAD AN ERRATIC (RED) DISPLAY. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED. THE SERVICE ENGINEER (SE) VERIFIED THE MALFUNCTION AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB) AND THE BACKLIGHT INVERTER PCB¿S WERE UPGRADED. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS.
Patient problems: No Patient Involvement
MAUDE report 6568061 ↗
Other VENTILATOR, CONTINUOUS, FACILITY USE devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page