Puritan Bennett 4-840220DIJI-ENRecalls & reported problems

Covidien LP · 800 Series Ventilator System,Model 840 · Class 2

All models of Puritan Bennett (119)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10884521077065), received between May 24, 2017 and Aug 10, 2017 — classified as 7 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • No Display/Image3
  • Battery Problem2
  • Device Inoperable1
  • Insufficient Device Problem Information1

Patient problems

  • No Patient Involvement7
  • No Information1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 10, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED AN 840 VENTILATOR HAD A FAULTY OXYGEN SENSOR. THERE WAS NO PATIENT INVOLVEMENT AT THE TIME OF THE REPORTED INCIDENT. THE SERVICE ENGINEER CLEANED AND CALIBRATED THE EXPIRATORY VALVE.

    Patient problems: No Patient Involvement

    MAUDE report 6784567 ↗
  • MalfunctionAug 7, 2017Battery Problem; Battery Problem

    IT WAS REPORTED THAT AN 840 VENTILATOR HAD A BACKUP BATTERY FAILURE. THE VENTILATOR WAS NOT ON A PATIENT AT THE TIME OF THE EVENT. THE SERVICE ENGINEER REPLACED THE BACKUP BATTERY TO CORRECT THE ISSUE. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS.

    Patient problems: No Patient Involvement

    MAUDE report 6771388 ↗
  • MalfunctionJun 27, 2017No Display/Image; No Display/Image

    IT WAS REPORTED AN 840 VENTILATOR HAD A FAULTY DISPLAY. AT THIS TIME, PATIENT INVOLVEMENT IS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Information; No Patient Involvement

    MAUDE report 6669699 ↗
  • MalfunctionJun 12, 2017Battery Problem; Battery Problem

    IT WAS REPORTED THAT AN 840 VENTILATOR HAD A BATTERY THAT DISCHARGED QUICKLY. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT. A SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND REPLACED THE BATTERY. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS.

    Patient problems: No Patient Involvement

    MAUDE report 6633637 ↗
  • MalfunctionJun 6, 2017No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE 840 VENTILATOR HAD A FAULTY DISPLAY. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 6613893 ↗
  • MalfunctionJun 1, 2017Device Inoperable; Device Inoperable

    THE EVALUATION AND REPAIR OF THE DEVICE HAS BEEN COMPLETED.

    Patient problems: No Patient Involvement

    MAUDE report 6606254 ↗
  • MalfunctionMay 24, 2017No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE 840 VENTILATOR HAD A GUI PCB (GRAPHICAL USER INTERFACE PRINTER CIRCUIT BOARD) ERROR. THE VENTILATOR WAS NOT ON A PATIENT AT THE TIME OF THE EVENT. THE EVALUATION AND REPAIR OF THE DEVICE HAS NOT BEEN COMPLETED.

    Patient problems: No Patient Involvement

    MAUDE report 6587845 ↗

Other VENTILATOR, CONTINUOUS, FACILITY USE devices