Puritan Bennett 4-840220DIPC-PRRecalls & reported problems
Covidien LP · 800 Series Ventilator System,Model 840 · Class 2
All models of Puritan Bennett (119)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10884521077102), received between May 10, 2017 and Sep 27, 2018 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Device Sensing Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Sensing Problem2
- Appropriate Device Problem Term/Code Not Available1
- Communication or Transmission Problem1
- Contamination1
- Device Inoperable1
- Mechanical Problem1
- No Display/Image1
- No Flow1
Patient problems
- No Patient Involvement6
- No Information2
- No Code Available1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 27, 2018No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE 840 VENTILATOR STOPPED VENTILATING. THE PATIENT WAS PLACED ON AN ALTERNATE VENTILATOR AND THERE WERE NO SIGNIFICANT CONSEQUENCES TO THE PATIENT AS A RESULT OF THE EVENT.
Patient problems: No Code Available
MAUDE report 7915914 ↗MalfunctionAug 23, 2017Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT AN 840 VENTILATOR HAD A TOUCH SCREEN THAT WAS BLOCKED. ALTHOUGH REQUESTED, IT IS UNKNOWN IF THE PATIENT WAS CONNECTED TO THE VENTILATOR AT THE TIME OF THE EVENT. THE SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND CLEANED THE DISPLAY. THE SE PERFORMED TESTING AND CALIBRATIONS AND THE VENTILATOR OPERATED WITHIN THE MANUFACTURING SPECIFICATIONS.
Patient problems: No Information
MAUDE report 6814702 ↗MalfunctionJul 17, 2017Contamination; Contamination
IT WAS REPORTED THAT, AN 840 VENTILATOR HAD A TOUCH SCREEN CONTAMINATED WITH LIQUID. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED. THE SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) TOUCH FRAME PRINTED CIRCUIT BOARD (PCB). THE UNIT PASSED ALL TESTING AND OPERATED WITHIN THE MANUFACTURING SPECIFICATIONS.
Patient problems: No Patient Involvement
MAUDE report 6719852 ↗MalfunctionJul 11, 2017Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT AN 840 VENTILATOR HAD A COMMUNICATION ERROR WITH THE BREATH DELIVERY UNIT (BDU) PRINTED CIRCUIT BOARD (PCB). THE VENTILATOR WAS NOT ON A PATIENT AT THE TIME OF THE EVENT. THE SERVICE ENGINEER INSPECTED THE DEVICE AND REPLACED THE BDU PCB CONNECTION CABLE, TO CORRECT THE ISSUE. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS.
Patient problems: No Patient Involvement
MAUDE report 6702373 ↗MalfunctionJun 6, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE, THE VENTILATOR HAD A KEYBOARD MALFUNCTION. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE EVENT. THE SERVICE ENGINEER INSPECTED THE DEVICE AND FOUND FLUID INGRESS ON THE KEYBOARD. THE SE CLEANED THE TOUCH FRAME PRINTED CIRCUIT BOARD (PCB). THE VENTILATOR PASSED ALL TESTING AND OPERATES WITHIN MANUFACTURING SPECIFICATIONS.
Patient problems: No Patient Involvement
MAUDE report 6616589 ↗MalfunctionMay 12, 2017Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT AN 840 VENTILATOR HAD AN INTERMITTENTLY BLOCKED TOUCHSCREEN. AT THIS TIME, IT IS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT. A SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND CONFIRMED THE REPORTED CONDITION. THEY FOUND LIQUID INFILTRATION ON THE FRONT HOUSING OF THE TOUCHSCREEN. TO RESOLVE THE ISSUE, SE REPLACED THE GRAPHICAL USER INTERFACE (GUI) TOUCHSCREEN, GUI KEYBOARD, AND GUI FRONT HOUSING. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS.
Patient problems: No Information
MAUDE report 6563186 ↗MalfunctionMay 10, 2017No Display/Image; No Display/Image
IT WAS REPORTED THAT, AN 840 VENTILATOR HAD A BLANK DISPLAY. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED. THE SERVICE ENGINEER (SE) VERIFIED THE MALFUNCTION AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB) AND THE BACKLIGHT INVERTER PCB¿S WERE UPGRADED. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS.
Patient problems: No Patient Involvement
MAUDE report 6559024 ↗MalfunctionMay 10, 2017Device Inoperable; Device Inoperable
IT WAS REPORTED THAT AN, 840 VENTILATOR GENERATED ERROR CODE WHICH RENDERED THE VENTILATOR INOPERABLE WITH SAFETY VALVE OPEN CONDITION. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED. THE SERVICE ENGINEER (SE) VERIFIED THE MALFUNCTION AND REPLACED THE BREATH DELIVERY UNIT (BDU) PRINTED CIRCUIT BOARD (PCB). THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS.
Patient problems: No Patient Involvement
MAUDE report 6557393 ↗MalfunctionMay 10, 2017Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT AN 840 VENTILATOR HAD A HARD KEYBOARD. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED CONDITION. TO RESOLVE THE ISSUE, A SERVICE ENGINEER (SE) REPLACED THE KEYBOARD. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS.
Patient problems: No Patient Involvement
MAUDE report 6556564 ↗
Other VENTILATOR, CONTINUOUS, FACILITY USE devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page