Skip to content

Puritan Bennett 4-840220DRAC-ENRecalls & reported problems

Covidien LP · 800 Series Ventilator System,Model 840 · Class 2

All models of Puritan Bennett (119)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10884521077171), received between Mar 30, 2017 and Jun 23, 2017 — classified as 5 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • No Display/Image4
  • Erratic or Intermittent Display1

Patient problems

  • No Patient Involvement3
  • No Information2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 23, 2017Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS REPORTED THAT AN 840 VENTILATOR HAD AN ERRATIC LOWER DISPLAY. THE VENTILATOR WAS NOT ON A PATIENT AT THE TIME OF THE EVENT. THE SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB) AND UPDATED THE BACKLIGHT INVERTER PCBS. THE UNIT PASSED ALL TESTING AND OPERATED WITHIN THE MANUFACTURING SPECIFICATIONS.

    Patient problems: No Patient Involvement

    MAUDE report 6663945 ↗
  • MalfunctionMay 8, 2017No Display/Image; No Display/Image

    IT WAS REPORTED THAT AN 840 VENTILATOR HAD AN ERROR CODE, IT WAS REQUIRED TO CHANGE THE GUI CPU PCB. THE VENTILATOR WAS NOT ON A PATIENT AT THE TIME OF THE EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 6551058 ↗
  • MalfunctionApr 25, 2017No Display/Image; No Display/Image

    IT WAS REPORTED THAT, AN 840 VENTILATOR HAD A SOFTWARE UPDATE DUE TO A DEFECTIVE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED. THE SERVICE ENGINEER (SE) VERIFIED THE MALFUNCTION AND REPLACED THE GUI PCB. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS.

    Patient problems: No Patient Involvement

    MAUDE report 6518235 ↗
  • MalfunctionMar 31, 2017No Display/Image; No Display/Image

    IT WAS REPORTED THAT, AN 840 VENTILATOR HAD A FAULTY DISPLAY. PATIENT INVOLVEMENT INFORMATION WAS NOT PROVIDED.

    Patient problems: No Information

    MAUDE report 6451062 ↗
  • MalfunctionMar 30, 2017No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE PB840 VENTILATOR HAS A DISPLAY ISSUE. PATIENT INVOLVEMENT WAS NOT PROVIDED. THE MEDTRONIC SERVICE ENGINEER (SE) EVALUATED THE VENTILATOR AND REPLACED THE GRAPHIC USER INTERFACE (GUI) CENTRAL PROCESSING UNIT (CPU) PRINTED CIRCUIT BOARD (PCB) AND THE BACKLIGHT INVERTER PCBS. THE VENTILATOR PASSED ALL TESTING AND WAS PLACED BACK IN USE AT THE CUSTOMER SITE.

    Patient problems: No Information

    MAUDE report 6447592 ↗