Puritan Bennett 4-840220NISC-ENRecalls & reported problems

Covidien LP · 800 Series Ventilator System,Model 840 · Class 2

All models of Puritan Bennett (119)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884521077218), received between Mar 29, 2017 and Jul 13, 2020 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Insufficient Device Problem Information1
  • No Display/Image1
  • No Flow1

Patient problems

  • No Patient Involvement2
  • High Blood Pressure/ Hypertension1
  • Low Blood Pressure/ Hypotension1
  • No Information1
  • Tachycardia1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 13, 2020No Flow; No Flow

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, AN 840 VENTILATOR GENERATED A HIGH VOLUME ALARM, RED SCREEN AND THE UNIT STOPPED VENTILATING THE PATIENT. AFTER THE VENTILATOR WAS REBOOTED, IT WAS STILL NOT WORKING. THE PATIENT WAS REMOVED FROM THE VENTILATOR, MANUALLY VENTILATED AND PLACED ON AN ALTERNATE VENTILATOR. THE PATIENT EXPERIENCED TACHYCARDIA (90 BEATS PER MINUTE), HYPERTENSION FOLLOWED BY HYPOTENSION WHICH WAS TREATED WITH NORADRENALINE. THE PATIENT RETURNED TO BASELINE WITHIN 30 MINUTES.

    Patient problems: High Blood Pressure/ Hypertension; Low Blood Pressure/ Hypotension; Tachycardia

    MAUDE report 10268179 ↗
  • MalfunctionAug 1, 2017No Display/Image; No Display/Image

    IT WAS REPORTED THAT, AN 840 VENTILATOR HAD A BLANK THE UPPER DISPLAY. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE EVENT OCCURRED. THE SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND REPLACED THE POWER SUPPLY TO CORRECT THE ISSUE. THE UNIT PASSED ALL TESTING AND OPERATED WITHIN THE MANUFACTURER SPECIFICATIONS.

    Patient problems: No Patient Involvement

    MAUDE report 6758930 ↗
  • MalfunctionMar 29, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT, AN 840 VENTILATOR HAD A BATTERY ISSUE. INFORMATION ABOUT PATIENT INVOLVEMENT WAS NOT PROVIDED. THE SERVICE ENGINEER (SE) EVALUATED THE VENTILATOR AND VERIFIED THE CUSTOMER REPORTED ISSUE. THE SE REPLACED THE BATTERY AND UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS.

    Patient problems: No Information; No Patient Involvement

    MAUDE report 6444875 ↗

Other VENTILATOR, CONTINUOUS, FACILITY USE devices

All VENTILATOR, CONTINUOUS, FACILITY USE devices →