Puritan Bennett G-061792-00Recalls & reported problems
Covidien LP · 700 Series Ventilator,Model 740 · Class 2
All models of Puritan Bennett (119)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10884521078369), received between Nov 20, 2017 and Aug 7, 2020 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Device Inoperable” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Inoperable3
- No Flow3
- Defective Alarm1
- Electrical /Electronic Property Problem1
- Mechanical Problem1
- Smoking1
Patient problems
- No Consequences Or Impact To Patient3
- No Patient Involvement3
- No Code Available2
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 7, 2020No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE 740 VENTILATOR ALARMED AND TURNED OFF. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITHOUT HARM AS THE RESULT OF THE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10382500 ↗MalfunctionMay 23, 2019Defective Alarm; No Flow; Defective Alarm; No Flow
IT WAS REPORTED THAT DURING USE ON PATIENT THE 740 VENTILATOR LOST COMPLETE POWER SUPPLY AND STOPPED VENTILATION. ADDITIONALLY, IT WAS NOTED THAT THE DEVICE DID NOT GENERATE A SHUTDOWN ALARM. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITH NO HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8638374 ↗MalfunctionApr 25, 2019Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING PREVENTATIVE MAINTENANCE, THE 740 VENTILATOR EXPERIENCED OXYGEN (O2) SOLENOID PRESSURE ERRORS AND DID NOT PASS THE EXTENDED SELF TEST (EST) DUE TO A LEAK. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Patient Involvement
MAUDE report 8552769 ↗MalfunctionApr 17, 2019No Flow; No Flow
IT WAS REPORTED THAT DURING USE ON PATIENT THE 740 VENTILATOR SHUTDOWN. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITHOUT HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8525400 ↗MalfunctionAug 17, 2018Smoking; Smoking
IT WAS REPORTED THAT THE 740 VENTILATOR AFTER TURNING ON THE DEVICE "POWER SUPPLY SMOKED". THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Patient Involvement
MAUDE report 7794147 ↗MalfunctionJul 3, 2018Device Inoperable; Device Inoperable
IT WAS REPORTED THAT DURING USE ON A PATIENT, THE VENTILATOR BECAME INOPERATIVE AND DISPLAYED A 'SAFETY VALVE OPEN' MESSAGE. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR. THERE WAS NO ADVERSE EVENT OR CONSEQUENCES AS A RESULT.
Patient problems: No Code Available
MAUDE report 7660481 ↗MalfunctionApr 19, 2018Device Inoperable; Device Inoperable
IT WAS REPORTED THAT THE VENTILATOR BECAME INOPERABLE WHILE ON A PATIENT. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR. THERE WAS NO ADVERSE EVENT OR CONSEQUENCES AS A RESULT OF THE REPORTED EVENT.
Patient problems: No Code Available
MAUDE report 7444240 ↗MalfunctionNov 20, 2017Electrical /Electronic Property Problem; Device Inoperable; Electrical /Electronic Property Problem; Device Inoperable
IT WAS REPORTED THAT A 740 VENTILATOR FAILED THE POWER ON SELF-TEST (POST) AND ALARMED A VENT INOPERABLE. THERE WAS NO PATIENT INVOLVEMENT AT THE TIME OF THE EVENT.
Patient problems: No Patient Involvement
MAUDE report 7046089 ↗
Other VENTILATOR, CONTINUOUS, FACILITY USE devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.