Steel 88862222-69Recalls & reported problems
Covidien LP · Monofilament Stainless Steel · Class 2 · Implant
All models of Steel (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10884521102330), received between Nov 1, 2017 and Sep 6, 2023 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break4
- Product Quality Problem2
- Detachment of Device or Device Component1
- Difficult To Position1
- Unraveled Material1
Patient problems
- No Known Impact Or Consequence To Patient3
- Insufficient Information2
- Tissue Damage1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 6, 2023Break; Product Quality Problem; Break; Product Quality Problem
ACCORDING TO THE REPORTER, DURING AN ANTERIOR INTERVENTRICULAR CORONARY ARTERY BYPASS GRAFTING, THE SURGEON NOTICED THAT THE STEEL WIRES WERE BREAKING WHILE THE WIRE WAS BEING TIGHTENED. IT WAS ALSO NOTED THAT THE WIRE APPEARED TO BE MORE FRAGILE IN THE CRIMPING AREA. A NEW DEVICE WAS USED TO RESOLVE THE ISSUE.
Patient problems: Insufficient Information
MAUDE report 17691262 ↗MalfunctionSep 6, 2023Break; Product Quality Problem; Break; Product Quality Problem
ACCORDING TO THE REPORTER, THE SURGEON NOTICED THAT THE STEEL WIRES WERE BREAKING WHILE THE WIRE WAS BEING TIGHTENED. IT WAS ALSO NOTED THAT THE WIRE APPEARED TO BE MORE FRAGILE IN THE CRIMPING AREA.
Patient problems: Insufficient Information
MAUDE report 17691108 ↗InjuryJan 15, 2020Unraveled Material; Unraveled Material
ACCORDING TO THE REPORTER, WHEN THE STERNUM WAS CLOSED DURING A MYOCARDIAL REVASCULARIZATION, THREE SUTURES DISLODGED. THERE WAS TISSUE DAMAGE AS A RESULT OF THIS PROBLEM.
Patient problems: Tissue Damage
MAUDE report 9590661 ↗MalfunctionJun 12, 2018Break; Break
ACCORDING TO THE REPORTER, WHILE TRYING TO CLOSE A STERNUM USING A CLASSIC SUTURING TECHNIQUE DURING AN OPEN CARDIAC SURGERY, THE NEEDLE BROKE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7592063 ↗MalfunctionApr 5, 2018Difficult To Position; Difficult To Position
ACCORDING TO THE REPORTER, DURING A CARDIAC SURGERY, WHEN THEY WERE USING STEEL WIRES TO CLOSE THE STERNUM OF PATIENT, THE NEEDLE LOOSENED. THEY USED ANOTHER PRODUCT TO COMPLETE THE CASE. PATIENT STATUS WAS ASKED BUT UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7399894 ↗MalfunctionNov 1, 2017Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
ACCORDING TO THE REPORTER: OCCURRED DURING AN OPEN VALVE REPLACEMENT PROCEDURE. THE SUTURE WAS NOT TREATED OR HYDRATED PRIOR TO USE. THE NEEDLE DETACHED FROM THE THREAD WHEN EMPLOYING A NORMAL SUTURING TECHNIQUE DURING STERNUM CLOSURE. IN ORDER TO RESOLVE THE ISSUE AND COMPLETE THE CASE, AN ADDITIONAL SUTURE WAS USED WITHOUT ISSUE. THERE WAS NO PATIENT HARM. THE PATIENT OUTCOME IS ALIVE, NO INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6994406 ↗
Other SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE devices
- ETHICON M660GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE102 reports
- ETHICON M650GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE68 reports
- ETHIBOND EXCEL X425HEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE1 recall12 reports
- Steel 88862224-49Covidien LP · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE1 recall3 reports
- ETHICON M655GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE46 reports
- ETHIBOND EXCEL MB66GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE42 reports
- Steel 88862424-89Covidien LP · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE39 reports
- ETHICON M649GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE36 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page