Steel 88862226-69Recalls & reported problems
Covidien LP · Monofilament Stainless Steel · Class 2 · Implant
All models of Steel (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10884521102385), received between May 23, 2017 and Aug 22, 2023 — classified as 2 injuries, 8 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break6
- Detachment of Device or Device Component5
- Product Quality Problem2
- Material Twisted/Bent1
- Unraveled Material1
Patient problems
- No Code Available4
- Insufficient Information2
- No Known Impact Or Consequence To Patient2
- Hemostasis1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- No Information1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 22, 2023Break; Product Quality Problem; Break; Product Quality Problem
ACCORDING TO THE REPORTER, THE SURGEON NOTICED THAT THE STEEL WIRES WERE BREAKING WHILE THE WIRE WAS BEING TIGHTENED. IT WAS ALSO NOTED THAT THE WIRE APPEARED TO BE MORE FRAGILE IN THE CRIMPING AREA.
Patient problems: Insufficient Information
MAUDE report 17600351 ↗MalfunctionAug 21, 2023Break; Product Quality Problem; Break; Product Quality Problem
ACCORDING TO THE REPORTER, DURING AN ANTERIOR INTERVENTRICULAR CORONARY ARTERY BYPASS GRAFTING, THE SURGEON NOTICED THAT THE STEEL WIRES WERE BREAKING WHILE THE WIRE WAS BEING TIGHTENED. IT WAS ALSO NOTED THAT THE WIRE APPEARED TO BE MORE FRAGILE IN THE CRIMPING AREA. A NEW DEVICE WAS USED TO RESOLVE THE ISSUE.
Patient problems: Insufficient Information
MAUDE report 17587271 ↗MalfunctionNov 1, 2021Break; Break
ACCORDING TO THE REPORTER, ON A LAPAROSCOPIC AORTIC VALVE REPLACEMENT, WHEN CLOSING THE STERNUM, THE PASSAGE WAS DIFFICULT, AND THE NEEDLE BROKE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12729708 ↗MalfunctionJul 2, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING MITRAL VALVE REPAIR, AT STERNOTOMY CLOSURE, THE NEEDLE LOOSENED FROM THE THREAD. THE SURGICAL TIME EXTENDED 30 MINUTES OR MORE DUE TO THE REPORTED CONDITION. THERE WAS NO PATIENT INJURY.
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 8753345 ↗InjuryMay 14, 2019Detachment of Device or Device Component; Material Twisted/Bent; Detachment of Device or Device Component; Material Twisted/Bent
ACCORDING TO THE REPORTER, AT THE TIME OF CLOSURE DURING THE EXPLORATORY STERNOTOMY, THE NEEDLE USED APPEARED FRAGILE FROM THE FIRST TRANSOSSEOUS PASSAGE. THE SUTURED DULLED AND TWISTED UNTIL IT BECAME STRAIGHT AND DISENGAGED FROM THE STEEL WIRE. IN ADDITION, STEEL WIRE WAS DIFFICULT TO TIGHTEN AT THE BONE EDGES BECAUSE IT BROKE. A FOREIGN BODY WAS IN THE SURGICAL SITE. THE SURGICAL TIME WAS EXTENDED TO 30 MINUTES OR MORE AND INCISION WAS EXTENDED TO MORE THAN 1 INCH. THE EXTENSION OF THE SURGICAL TIME WEAKENED THE STERNAL BONE.
Patient problems: No Code Available
MAUDE report 8609478 ↗MalfunctionMay 6, 2019Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING STERNUM CLOSURE, AT THE END OF THE PROCEDURE, WHEN THE STERNUM WAS CLOSED AND THE STEEL WIRES WERE LAID, 3 OUT OF 6 NEEDLES BROKE AND HAD TO BE REMOVED TO INSTALL NEW ONES. THE NEEDLE AND THREAD DISENGAGED WHEN PASSING THROUGH THE STERNUM. THE DEVICE REPLACED TO COMPLETE THE CASE. THE SURGICAL TIME WAS EXTENDED 30 MINUTES OR MORE DUE TO THE PRODUCT PROBLEM.
Patient problems: No Code Available
MAUDE report 8584911 ↗MalfunctionApr 22, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, WHEN TRYING TO CLOSE THE STERNUM ON AN OPEN PROCEDURE, THE STEEL WIRES WERE DISLODGED. THEY USED ANOTHER DEVICE TO COMPLETE THE CASE.
Patient problems: No Information
MAUDE report 8537194 ↗InjuryMar 21, 2019Unraveled Material; Unraveled Material
ACCORDING TO THE REPORTER, DURING AN OPEN PROCEDURE, WHEN CLOSING THE SURGICAL SITE, THE WIRE OF THE STEEL LOOSENED AS PASSING THE STERNUM. THE PROBLEM COMPLICATES THE CLOSURE AND REQUIRES ADDITIONAL HEMOSTASIS OF THE STERNUM. THE CLOSURE TIME WAS LENGTHENED.
Patient problems: Hemostasis; No Code Available
MAUDE report 8439913 ↗MalfunctionFeb 27, 2018Break; Break
ACCORDING TO THE REPORTER, DURING THE CLOSURE OF THE "STERNOMOTY", ON THE INTRODUCTION OF THE NEEDLE INTO THE TISSUES, THE NEEDLE BROKE. THERE WERE MORE THAN ONE DEFECTIVE DEVICE INVOLVED. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE IN ORDER TO COMPLETE THE CASE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7299242 ↗MalfunctionMay 23, 2017Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
THE NEEDLE BREAKS OFF DURING USAGE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6584766 ↗
Other SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE devices
- ETHICON M660GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE102 reports
- ETHICON M650GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE68 reports
- ETHIBOND EXCEL X425HEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE1 recall12 reports
- Steel 88862224-49Covidien LP · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE1 recall3 reports
- ETHICON M655GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE46 reports
- ETHIBOND EXCEL MB66GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE42 reports
- Steel 88862424-89Covidien LP · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE39 reports
- ETHICON M649GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE36 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page