Steel 88862393-89Recalls & reported problems
Covidien LP · Monofilament Stainless Steel · Class 2 · Implant
All models of Steel (15)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10884521102484), received between Jul 1, 2019 and Apr 29, 2024 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Break” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break8
- Detachment of Device or Device Component3
- Material Twisted/Bent3
- Manufacturing, Packaging or Shipping Problem1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient4
- Insufficient Information1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 29, 2024Break; Break
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC PROCEDURE, FOUR SUTURE THREADS BROKE. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19205513 ↗MalfunctionApr 29, 2024Break; Break
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC PROCEDURE, FOUR SUTURE THREADS BROKE. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19205393 ↗MalfunctionApr 29, 2024Break; Break
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC PROCEDURE, FOUR SUTURE THREADS BROKE. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19205308 ↗MalfunctionMar 28, 2024Break; Break
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC PROCEDURE, THE THREAD BROKE. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18992916 ↗MalfunctionFeb 26, 2024Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
ACCORDING TO THE REPORTER, PRE-OPERATIVELY, THERE WAS ONE PLUS SUTURE ON THE PACKAGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18778380 ↗MalfunctionOct 13, 2022Break; Use of Device Problem; Break; Use of Device Problem
ACCORDING TO THE REPORTER, DURING OPEN THORACOTOMY, AT STERNUM CLOSURE, THE NEEDLE WAS BENDING, MAKING THE SUTURE DIFFICULT TO PASS AND DID NOT HAVE ADEQUATE RESISTANCE. THE NEEDLE THEN BROKE. ANOTHER SUTURE WAS USED TO RESOLVE THE ISSUE IN ORDER TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15597106 ↗MalfunctionOct 13, 2022Break; Material Twisted/Bent; Break; Material Twisted/Bent
ACCORDING TO THE REPORTER, THE NEEDLE BROKE.
Patient problems: Insufficient Information
MAUDE report 15596697 ↗MalfunctionOct 19, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, WHEN STERNUM CLOSURE ON AN OPEN CARDIOVASCULAR PROCEDURE, THE NEEDLE BROKE AND FELL INTO THE PATIENT¿S CAVITY. ANOTHER PRODUCT WAS OPENED. THE EVENT RESULTED TO AN EXTENDED SURGICAL TIME TO 30 MINUTES OR MORE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12653467 ↗MalfunctionJan 8, 2020Break; Material Twisted/Bent; Break; Material Twisted/Bent
ACCORDING TO THE REPORTER, DURING A MYOCARDIAL REVASCULARIZATION PROCEDURE, THE SUTURE NEEDLE WAS BROKEN. IT WAS REPORTED THAT WHEN THE WIRE PASSES THROUGH THE STERNUM, THE NEEDLE IS BENT COMPLETELY, THE NEEDLE IS FULLY TWISTED WHILE TRYING TO PASS THROUGH THE BONE. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9567231 ↗MalfunctionJan 6, 2020Break; Material Twisted/Bent; Break; Material Twisted/Bent
ACCORDING TO THE REPORTER, DURING OPEN CARDIOVASCULAR SURGERY, IN THE FIRST ATTEMPT, THE NEEDLE WAS BENT. FOR THE SECOND ATTEMPT, IT WAS ON PART OF THE WIRE IN THE AREA OF THE NEEDLE JUNCTION. WHEN MAKING THE THIRD ATTEMPT, THE NEEDLE BODY LOST ITS SHAPE. THE NEEDLE BROKE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9554864 ↗MalfunctionJul 2, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING THE RE-APPROXIMATION OF THE STERNOTOMY ON AN OPEN CORONARY ARTERY BYPASS GRAFTING, THE SUTURE BROKE FROM THE SWAGED END. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8751607 ↗MalfunctionJul 1, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
FDA clearance and classification
Cleared through 510(k) K955723 on Feb 27, 1996 (decision: Substantially Equivalent for Some Indications). Submitted by Davis & Geck, Inc. as “Stainless Steel Sutures, Flexon Steel Sutures (Stainless)”. The FDA review took 71 days. FDA record ↗
The FDA regulates product code GAQ (suture, nonabsorbable, steel, monofilament and multifilament, sterile) under 21 CFR 878.4495 as a Class II device. 21 CFR 878.4495 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Steel 88862393-89
Has the Steel 88862393-89 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10884521102484), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Steel 88862393-89?
The FDA's MAUDE database lists 12 adverse event reports for this device, received between Jul 1, 2019 and Apr 29, 2024. In those reports, the most common specific device problems are “Break” (8), “Detachment of Device or Device Component” (3) and “Material Twisted/Bent” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Steel 88862393-89?
None of the reports are classified as deaths, 0 as injuries and 12 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Steel 88862393-89 FDA approved?
It was cleared through 510(k) K955723 on Feb 27, 1996. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Steel 88862393-89 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the Steel 88862393-89, and is it still sold?
The FDA's device database lists Covidien LP as the company responsible for it (brand Steel, model 88862393-89). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE devices
- ETHICON M660GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE100 reports
- ETHICON M650GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE67 reports
- ETHIBOND EXCEL X425HEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE1 recall12 reports
- Steel 88862224-49Covidien LP · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE1 recall3 reports
- ETHICON M655GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE46 reports
- ETHIBOND EXCEL MB66GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE40 reports
- Steel 88862424-89Covidien LP · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE39 reports
- ETHICON M649GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE35 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.