Puritan Bennett 980S1ENDICURecalls & reported problems
Covidien LP · 980 Pediatric-Adult Ventilator · Class 2
All models of Puritan Bennett (119)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10884521171114), received between Sep 2, 2020 and Jan 10, 2023 — classified as 6 malfunctions.
The most frequently reported specific device problem is “No Flow” (4 reports).
3 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Mar 25, 2022Class IOpen, Classified
The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially resulting in hypoxia, dyspnea, cardiac arrest, or death.
Z-0966-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Nov 4, 2021Class ITerminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
Z-0340-2022 · Root cause: Reprocessing Controls · Recall details · FDA record ↗
- Oct 16, 2020Class IIOpen, Classified
In-house lab observations and customer reports of overheating and/or thermal damage to the battery pack, create the risk of exposure to smoke or hazardous materials, which may result in burns and/or bronchospasm,
Z-0425-2021 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- No Flow4
- Improper Flow or Infusion2
- Infusion or Flow Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 10, 2023Improper Flow or Infusion; Improper Flow or Infusion
IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR GENERATED A ¿VENTILATOR INOPERATIVE¿ MESSAGE AND DIAGNOSTIC CODE INDICATING BACKUP VENTILATION. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITH NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16132221 ↗MalfunctionJul 26, 2022Improper Flow or Infusion; Improper Flow or Infusion
IT WAS REPORTED THAT, WHILE INITIATING THE PATIENT ON BIPAP (BILEVEL POSITIVE AIRWAY PRESSURE) APPROXIMATELY 3 MINUTES ELAPSED AND THIS 980 VENTILATOR ENTERED INTO BACKUP VENTILATION WITH A CONSTANT HIGH EMERGENCY ALARM THAT COULD NOT BE SILENCED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITH NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15104795 ↗MalfunctionMar 31, 2021No Flow; No Flow
IT WAS REPORTED THAT DURING USE ON A POST CARDIAC SURGICAL PATIENT, THIS 980 VENTILATOR SPONTANEOUS SHUT OFF WITHOUT WARNING. AFTER ABOUT ONE MINUTE THE VENTILATOR REBOOTED ITSELF. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITH NO HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11596009 ↗MalfunctionJan 26, 2021No Flow; No Flow
IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR ¿TURNED OFF¿. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITH NO HARM OR INJURY. BASED ON THE REVIEW OF THE VENTILATOR LOGS, A LOSS OF VENTILATION OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11230483 ↗MalfunctionDec 15, 2020No Flow; No Flow
IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, A 980 VENTILATOR GENERATED A HIGH PRIORITY ALARM THAT COULD NOT BE SILENCED OR RESET AND A SAFETY VALVE OPEN ALERT CONDITION. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITH NO HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11012905 ↗MalfunctionSep 2, 2020Infusion or Flow Problem; No Flow; Infusion or Flow Problem; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, IN A NON-INVASIVE MODE, THIS 980 VENTILATOR WAS NOT VENTILATING AND GENERATED THE FOLLOWING MESSAGES, "VENT INOP (VENTILATOR INOPERATIVE), NO O2 (OXYGEN) SUPPLY, NO AIR SUPPLY". THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITH NO HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10485606 ↗
Other VENTILATOR, CONTINUOUS, FACILITY USE devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page