Puritan Bennett 980X1ITDIECRecalls & reported problems

Covidien LP · 980 Ventilator · Class 2

All models of Puritan Bennett (119)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884521171602), received between Dec 13, 2017 and Mar 4, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Defective Alarm” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 25, 2022Class IOpen, Classified

    The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially resulting in hypoxia, dyspnea, cardiac arrest, or death.

    Z-0966-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Defective Alarm1
  • No Audible Alarm1
  • No Flow1

Patient problems

  • No Patient Involvement2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 4, 2025Defective Alarm; Defective Alarm

    IT WAS REPORTED THAT THIS 980 VENTILATOR DOES NOT MAKE SOUND AND THE ALARM CANNOT BE HEARD. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: Insufficient Information

    MAUDE report 21520793 ↗
  • MalfunctionAug 31, 2018No Flow; No Flow

    IT WAS REPORTED THAT A 980 VENTILATOR HAD A "SYSTEM FAILURE" ERROR CODE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 7836839 ↗
  • MalfunctionDec 13, 2017No Audible Alarm; No Audible Alarm

    IT WAS REPORTED THAT A 980 VENTILATOR FAILED THE BREATH DELIVERY (BD) AUDIO TEST ON AN EXTENDED SELF TEST (EST) AND HAD A "POWER FAIL CAP NOT DISCHARGING" MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 7113227 ↗