Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884521171602), received between Dec 13, 2017 and Mar 4, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Defective Alarm” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 25, 2022Class IOpen, Classified
The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially resulting in hypoxia, dyspnea, cardiac arrest, or death.
Z-0966-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Defective Alarm1
- No Audible Alarm1
- No Flow1
Patient problems
- No Patient Involvement2
- Insufficient Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 4, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THAT THIS 980 VENTILATOR DOES NOT MAKE SOUND AND THE ALARM CANNOT BE HEARD. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: Insufficient Information
MAUDE report 21520793 ↗MalfunctionAug 31, 2018No Flow; No Flow
IT WAS REPORTED THAT A 980 VENTILATOR HAD A "SYSTEM FAILURE" ERROR CODE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Patient Involvement
MAUDE report 7836839 ↗MalfunctionDec 13, 2017No Audible Alarm; No Audible Alarm
IT WAS REPORTED THAT A 980 VENTILATOR FAILED THE BREATH DELIVERY (BD) AUDIO TEST ON AN EXTENDED SELF TEST (EST) AND HAD A "POWER FAIL CAP NOT DISCHARGING" MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Patient Involvement
MAUDE report 7113227 ↗
Other VENTILATOR, CONTINUOUS, FACILITY USE devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page