Puritan Bennett 980X1PLDIPCRecalls & reported problems
Covidien LP · 980 Ventilator · Class 2
All models of Puritan Bennett (119)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10884521171619), received between Jun 22, 2017 and Sep 1, 2023 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (3 reports).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Mar 25, 2022Class IOpen, Classified
The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially resulting in hypoxia, dyspnea, cardiac arrest, or death.
Z-0966-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Nov 4, 2021Class ITerminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
Z-0350-2022 · Root cause: Reprocessing Controls · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Displays Incorrect Message3
- Improper Flow or Infusion2
- No Flow2
- Appropriate Device Problem Term/Code Not Available1
- Calibration Problem1
- Defective Alarm1
- Device Operates Differently Than Expected1
- Inappropriate or Unexpected Reset1
Patient problems
- No Patient Involvement6
- No Information5
- No Code Available2
- Insufficient Information1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 1, 2023Improper Flow or Infusion; Improper Flow or Infusion
UPDATED: IT WAS REPORTED THAT THIS 980 VENTILATOR GENERATED AN EXHALATION BACKUP VENTILATION MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: Insufficient Information
MAUDE report 17674054 ↗MalfunctionAug 26, 2019No Flow; No Flow
IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, A 980 VENTILATOR GENERATED A ¿BACKGROUND CRASH¿ MESSAGE, THE SCREEN WAS UNRESPONSIVE AND THEN THE VENTILATOR STOPPED WORKING. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR. ALTHOUGH REQUESTED, INFORMATION PERTAINING TO PATIENT OUTCOME HAS NOT BEEN PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8932315 ↗MalfunctionJun 19, 2019Improper Flow or Infusion; Improper Flow or Infusion
IT WAS REPORTED THAT A 980 VENTILATOR ENTERED BACK UP VENTILATION WITH THE INSPIRATORY PROPORTIONAL SOLENOID STUCK OPEN AND AN ERROR CODE. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN PROVIDED.
Patient problems: No Information
MAUDE report 8712021 ↗MalfunctionSep 5, 2018Defective Alarm; Defective Alarm
IT WAS REPORTED THAT A 980 VENTILATOR BUTTONS DID NOT WORK AND THE ALARM DID NOT SOUND. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Patient Involvement
MAUDE report 7848694 ↗MalfunctionAug 17, 2018No Flow; No Flow
IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, A 980 VENTILATOR HAD A LOSS OF VENTILATION AS IT STOPPED CYCLING AND GENERATED A DEVICE ALERT WITH AN ERROR CODE. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND MANUALLY VENTILATED VIA AN AMBU BAG BEFORE BEING PLACED ON AN ALTERNATE VENTILATOR. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE REPORTED EVENT.
Patient problems: No Code Available
MAUDE report 7794965 ↗MalfunctionNov 27, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected
IT WAS REPORTED THAT A 980 VENTILATOR WAS SWITCHING INTO BACKUP VENTILATION MODE RANDOMLY. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE REPORTED EVENT.
Patient problems: No Code Available
MAUDE report 7062152 ↗MalfunctionAug 24, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE OXYGEN (O2) SENSOR ON A 980 VENTILATOR EXPERIENCED A FAILURE. THERE WAS NO PATIENT INVOLVEMENT AT THE TIME OF THE REPORTED EVENT. THE SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND VERIFIED THE REPORTED ISSUE. THE SE REPLACED TO O2 SENSOR. THE SE PERFORMED CALIBRATIONS AND EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED.
Patient problems: No Patient Involvement
MAUDE report 6818777 ↗MalfunctionAug 10, 2017Calibration Problem; Calibration Problem
IT WAS REPORTED THAT THE OXYGEN (O2) SENSOR ON A 980 VENTILATOR FAILED CALIBRATION. ALTHOUGH REQUESTED, IT IS UNKNOWN IF THE PATIENT WAS CONNECTED TO THE VENTILATOR AT THE TIME OF THE REPORTED EVENT. THE SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND VERIFIED THE REPORTED ISSUE. THE SE REPLACED THE O2 SENSOR. THE SE PERFORMED CALIBRATIONS AND EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED.
Patient problems: No Information; No Patient Involvement
MAUDE report 6785871 ↗MalfunctionJul 31, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Information; No Patient Involvement
MAUDE report 6754898 ↗MalfunctionJul 21, 2017Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset
THERE WAS NO PATIENT INVOLVEMENT AT THE TIME OF THE EVENT.
Patient problems: No Information; No Patient Involvement
MAUDE report 6732283 ↗MalfunctionJul 7, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT, DURING TESTING, A 980 VENTILATOR GENERATED A POWER ON SELF-TEST (POST) ERROR CODE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT. THE SERVICE ENGINEER (SE) EVALUATED THE VENTILATOR AND VERIFIED THE REPORTED ISSUE. THE FLOW VALVE NEEDED TO BE REPLACED. THE EVALUATION AND REPAIR OF THE DEVICE HAS NOT BEEN COMPLETED.
Patient problems: No Patient Involvement
MAUDE report 6695119 ↗MalfunctionJun 22, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT, A 980 VENTILATOR GENERATED A COMMUNICATION ERROR MESSAGE. ALTHOUGH REQUESTED, IT IS UNKNOWN IF THE PATIENT WAS CONNECTED TO THE VENTILATOR AT THE TIME OF THE EVENT. THE SERVICE ENGINEER (SE) INSPECTED THE DEVICE. THE SE REPLACED THE MEMORY CARD AND PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED.
Patient problems: No Information
MAUDE report 6661296 ↗
Other VENTILATOR, CONTINUOUS, FACILITY USE devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page