Puritan Bennett 980X1ESDIECRecalls & reported problems
Covidien LP · 980 Ventilator · Class 2
All models of Puritan Bennett (119)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 10884521171633), received between Mar 30, 2017 and May 2, 2025 — classified as 19 malfunctions.
The most frequently reported specific device problem is “Improper Flow or Infusion” (10 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 25, 2022Class IOpen, Classified
The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially resulting in hypoxia, dyspnea, cardiac arrest, or death.
Z-0966-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Oct 16, 2020Class IIOpen, Classified
In-house lab observations and customer reports of overheating and/or thermal damage to the battery pack, create the risk of exposure to smoke or hazardous materials, which may result in burns and/or bronchospasm,
Z-0425-2021 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Improper Flow or Infusion10
- Output Problem9
- No Flow3
- Battery Problem1
- Device Displays Incorrect Message1
- Device Emits Odor1
- Device Inoperable1
- Device Operates Differently Than Expected1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- No Patient Involvement4
- No Code Available3
- No Information2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 2, 2025Erratic or Intermittent Display; Device Emits Odor; No Flow; Erratic or Intermittent Display; Device Emits Odor; No Flow
IT WAS REPORTED THAT DURING THE SHORT SELF TEST (SST) ON THIS 980 VENTILATOR, THE VENTILATOR HAD A BLACK SCREEN AND WAS NOTED THERE WAS A "BURNED SMELL". THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT. REPORTABILITY WAS REASSESSED BASED ON FIELD SERVICE ENGINEER FINDINGS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21950581 ↗MalfunctionJan 24, 2025Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATORS FAILED THE EXTENDED SELF TEST (EST). THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21230907 ↗MalfunctionAug 23, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THESE 980 VENTILATORS FAILED THE EXTENDED SELF TEST (EST) WITH AN INSPIRATORY AUTO ZERO SOLENOID NOT OPERATIONAL MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20052881 ↗MalfunctionMar 13, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR FAILED THE SHORT SELF TEST (SST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18895088 ↗MalfunctionNov 9, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR'S SHORT SELF TEST (SST) FAILED. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18103501 ↗MalfunctionNov 9, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THE 980 VENTILATOR FAILED THE EXTENDED SELF-TEST (EST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18103256 ↗MalfunctionOct 19, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR OXYGEN (O2) PROPORTIONAL SOLENOID VALVE (PSOL) HAD A LEAK. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17971208 ↗MalfunctionOct 3, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR DID NOT PASS THE SHORT SELF TEST (SST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17862466 ↗MalfunctionAug 22, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR DISPLAYED A "INSPIRATORY AUTOZERO SOLENOID NOT OPERATIONAL" MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17596958 ↗MalfunctionJun 19, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR DID NOT PASS THE INSPIRATORY AUTO ZERO TEST PORTION OF THE SHORT SELF TEST (SST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17159614 ↗MalfunctionMay 9, 2023Obstruction of Flow; Improper Flow or Infusion; Obstruction of Flow; Improper Flow or Infusion
IT WAS REPORTED THAT DURING USE ON A PATIENT, THIS 980 VENTILATOR ALARMED A SEVERE OCCLUSION. PATIENT OUTCOME WAS NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16899576 ↗MalfunctionDec 31, 2021No Flow; No Flow
IT WAS REPORTED THAT, WHEN CONNECTED TO A PATIENT, A 980 VENTILATOR "DEVICE GETS STUCK." THE PATIENT WAS REMOVED FROM THE VENTILATOR AND MANUALLY VENTILATED VIA AN AMBU BAG BEFORE BEING PLACED ON AN ALTERNATE VENTILATOR WITH NO INJURY. A REVIEW OF THE DIAGNOSTIC LOGS INDICATES THERE WAS EVIDENCE OF A LOSS OF VENTILATION AT THE TIME OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13128752 ↗MalfunctionJul 19, 2018No Audible Alarm; Loss of Power; No Flow; No Flow
IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, A 980 VENTILATOR SHUT DOWN AND GENERATED AN AUDIBLE ALARM. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE REPORTED EVENT.
Patient problems: No Code Available; No Known Impact Or Consequence To Patient
MAUDE report 7704940 ↗MalfunctionJun 5, 2018Device Inoperable; Device Inoperable
IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, A 980 VENTILATOR DISPLAY TURNED OFF, ALARMED CONSTANTLY AND STOPPED CYCLING WHILE IN PRESSURE SUPPORT MODE. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE REPORTED EVENT.
Patient problems: No Code Available
MAUDE report 7571772 ↗MalfunctionJan 31, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected
IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, A 980 VENTILATOR ALARMED WITH ERROR CODES, WAS INOPERABLE AND BEGAN BACKUP VENTILATION. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE REPORTED EVENT.
Patient problems: No Code Available
MAUDE report 7230814 ↗MalfunctionAug 23, 2017Failure to Deliver; Failure to Deliver
THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Information; No Patient Involvement
MAUDE report 6814254 ↗MalfunctionJul 27, 2017Battery Problem; Battery Problem
IT WAS REPORTED THAT A 980 VENTILATOR HAD A BATTERY FAILURE. ALTHOUGH REQUESTED, INFORMATION PERTAINING TO THE USE OF THE DEVICE AT THE TIME OF THE REPORTED EVENT AND PATIENT OUTCOME (IF APPLICABLE) HAS NOT BEEN PROVIDED.
Patient problems: No Information; No Patient Involvement
MAUDE report 6747840 ↗MalfunctionJul 3, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT A 980 VENTILATOR HAD AN OXYGEN (O2) SENSOR FAILURE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT. THE SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND VERIFIED THE REPORTED ISSUE. THE SE CALIBRATED THE O2 SENSOR AND PERFORMED EXTENDED SELF TESTING. ALL TESTS PASSED.
Patient problems: No Patient Involvement
MAUDE report 6684807 ↗MalfunctionMar 30, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT, A 980 VENTILATOR FAILED THE POWER ON SELF TEST (POST) AND GENERATED AN INOPERABLE ERROR MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE EVENT OCCURRED. THE SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND FOUND A DIAGNOSTIC CODE RELEVANT TO THE REPORTED INCIDENT RECORDED IN THE VENTILATORS MEMORY LOGS. THE SE REPLACED THE POWER SUPPLY. THE SE UPDATED THE SOFTWARE TO THE CURRENT REVISION, PERFORMED CALIBRATIONS AND EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED.
Patient problems: No Patient Involvement
MAUDE report 6448594 ↗
Other VENTILATOR, CONTINUOUS, FACILITY USE devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page