Skip to content

Puritan Bennett 980X1ENDIPCRecalls & reported problems

Covidien LP · 980 Ventilator · Class 2

All models of Puritan Bennett (119)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

3 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Mar 25, 2022Class IOpen, Classified

    The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially resulting in hypoxia, dyspnea, cardiac arrest, or death.

    Z-0966-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Nov 4, 2021Class ITerminated

    Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.

    Z-0346-2022 · Root cause: Reprocessing Controls · Recall details · FDA record ↗

  3. Oct 16, 2020Class IIOpen, Classified

    In-house lab observations and customer reports of overheating and/or thermal damage to the battery pack, create the risk of exposure to smoke or hazardous materials, which may result in burns and/or bronchospasm,

    Z-0425-2021 · Root cause: Device Design · Recall details · FDA record ↗

Other VENTILATOR, CONTINUOUS, FACILITY USE devices

All VENTILATOR, CONTINUOUS, FACILITY USE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page